Material |
Velocity ( m/s) |
Air |
330 |
Water |
1497 |
Fat | 1440 |
Average soft tissue |
1540 |
Blood | 1570 |
Muscle |
1500 -
1630 |
Bone |
2700 -
4100 |
Metal | 3000 -
6000 |
Material interface |
Reflection coefficient |
Liver/Kidney |
0.006 |
Kidney/spleen |
0.003 |
Blood/Kidney |
0.009 |
Liver/fat |
0.11 |
Liver/Bone |
0.59 |
Liver /Air |
0.9995 |
a |
b |
c |
Second harmonic (1.7/3.5 MHz)
left and fundmental (3.5 MHz) right
images of LV septum, showing how the echo from the
blood/septum interface (arrows) is thicker in harmonic
imaging, due to the reduction in frequency. Observe,
however, how cavity noise is much reduced in harmonic
imaging, resulting in a far more favorable
signal-to-noise ratio. |
The thickness of the surtface
echoes is dependent n the pulse length, and thus also on
the frequency. This picture of the septum
illustrates how the leading-to-leading ASE convention
shown in red, eliminates the pulse length in measurement
(as the echo blooms in both directions), while the Penn
convention will result in increasing overestimation of
the thickness by increasing pulse length as it
incorporates the interface on both sides. |
The return pulse is a
waveform. To process this, multiple samples along the
waveform is needed. However, if the full waveform is
stored, both amplitude and frequency can be post
processed. |
The pulse has a certain
amplitude, depending on how much energy that is
reflected |
And the amplitude information is only a single value that is represented without furthwer information of the waveform. This is the only data that are used in B-mode and M-mode imaging | In addition to the
amplitude, the wave frequency can be stored as
additional information of the waveform, but still a
fairly simple dataset. This additional information that
can be used in Doppler. |
The absorption is dependent on many factors (117):
Signal
intensity is attenuated exponentially by depth as shown
by this curve. |
Attenuation
can be measured in decibels (logarithmic scale) as a
function of depth, the decibel scale will be linear |
Tissue |
Attenuation (Db / cm /MHz) |
Water |
0.02 |
Blood |
0.15 |
Liver |
0.40 |
Brain |
0.44 |
Muscle |
0.57 |
Bone |
22 |
Basically, discrete objects with high reflexivity wil cause
attenuation shadows, as shown below. However, the shadows are both
due to the absorbtion in dense tissue, and of the reflection at
the border zones, if they have a high reflexivity coefficient.
Behind organs with low density (reflexivity) on the other hand,
the tissue appears brighter (colouring). This is simply lack of
attenuation - acoustic enhancement.
Liver with a gallbladder in front, containing gallstones. The gallstones are dense, with rapidly decreasing echo by depth. The attenuation causes a shadow behind. The liver, is less dense, showing a more homogenous reflection. The rest of the gallbladder, however, is fluid filled, the fluid attenuates much less, so the liver tissue behind the gall bladder appears denser than the neighbouring tissue due the ultrasound being less attenuated (t"colouring") | A more subtle degree of
attenuation/colouring is seen in this short axis image
of the heart. The cavities of the RV and LV are both low
attenuating, so the septum and inferior wall are
brighter, due to colouring. The lateral walls, on the
other hands, have ultrasound beams passing along the
walls, so there is more attenuation and they appear
darker. |
In echocardiography, the shadows may even be useful, as a clue to
a high reflexivity of structures, meaning they are usually
calcified (thick, dense stuctures may seem very bright in the
usual scanner setting but will usually not cast a shadow if not
calcified.:
Calcification
in the posterior mitral ring seen both in parasternal
and apical views from the same patient. The shadow is
clearly visible in both views. |
|
Uncompensated image, showing
decreasing signal intensity (and, hence, visibility)
with depth, due to attenuation. |
Increasing over all gain, will
increase the amplitude of the
signal, and the structures at the
bottom of the sector becomes more visible. But the gain
in the top of the sector are also increased, including
the cavity noise, thus decreasing contast in this part
of the image. |
TGC controls.
Basically, each slider controls gain selectively at a
certain depth: |
In older models, the
TGC should be set manually to achieve a balanced image: |
Present models, however, have automatic TGC. Thus the default control setting should be neutral to achieve a balanced picture: | Using manual
setting by old habit will result in a double
compensation, with too much gain in the bottom, too
little in the top: |
Image with default gain, reject and compress settings | Principle of gain, reject and compress. All curves display brightness of the display in relation to the amplitude of the rejected signal. An ordinary gain curve is shown in black, using a linear brightness scale, displays the full range of amplitudes. Increasing gain (red curve), will increase all signals, including the weakest, as in the noise. The disadvantage, in addition to increasing noise, is that the strongest signals will be saturated, so details may disappear. Compress is shown as the blue curve. This results in a steeper brightness curve, resulting in less brightness of the weakest echoes, and more brightness of the strongest. Thus, weak echoes may disappear together with background noise, while strong echoes will be saturated, resulting in loss of detail. Finally reject is shown by the light grey zone, siply displaying all signals below a certain amplitude as black. (The black brightnes curve drops abruptly to zero at the reject limit (dark grey line). A combination of high gain and reject will give an effect fairly similar to the compress function. |
Same image with high gain
(top) showing increased density of the endocardium, but
loss of detail due to brightness saturation and a
corresponding increase in cavity noise and low gain
(bottom), showing reduction in cavity noise, but loss of
detail (see endocardium in lateral wall). |
Same image with increased
reject (top) showing reduction in cavity noise, but also
with slight loss of detail (endocardium in lateral wall)
and compress function (bottom) with less detail in the
myocardium due to increased brightness. |
The angle dependency of Doppler measurements, is well known.
However, M-mode measurements are just as angle dependent:
Effect of angulation in thickness
measurement. The true thickness is L0,
the measurement, however may be done along an M-mode
beam parallel to the length L. As the cosine to the
angle between them is defined as cos
( ) = L0 / L, the true
length L0 is overestimated by the cosine to
the angle, the measured line is the true line divided by
the cosine to the angle. |
Example from reconstructed M-mode,
with vertical scales aligned at zero and 6 cm depth for
comparison. To the left, the line crosses the septum
transversely resulting in a diastolic thickness measure
of 7mm, to the right the M-mode line is skewed, and the
measurement across the septum is longer (10 mm). |
Thus, a skewed cross sectional M.mode will overestimate both wall
thicknesses and chamber diameters.
Reconstructed M-mode with
a fairly straight cross angle between the M-mode line
and the LV long axis. |
Reconstructed M-mode from
the same loop, but with the M-mode line crossing the LV
long axis at a skewed angle, showing thicker walls and
wider cavity, due to the angulation. |
The angle distortion is eliminated, however, by using ratios.
Fractional shortening and wall thickening are ratios of diastolic
values and systolic changes, where the overestimation is present
in both diastole and systole (unless the angle changes during
systole, of course), and the ratios will remain unchanged.Thus
wall thickening and fractional shortening can be estimated even if
the line is skewed.
However measuring absolute dimensions or motion by M-mode, will
lead to an over estimation by M-mode just as in measuring
distances:
Angle dependency of motion measurement by M-mode. As a reflector moves from a to b in the direction 1, the true motion (displacement) is L1. If the ultrasound beam deviates from the direction of the motion by the angle , the apparent length along the ultrasound beam will be L2, which is the hypothenuse of the triangle, and thus L2 = L1 / cos (). Thus angle deviation of M-mode measures will always over estimate the real motion (as opposed to Dopller measurements). | The angle error in displacement
measurement demonstrated in a reconstructed M-mode. As
the skewed M-mode line is shorter, scales have been
lignes at 0 and 6 cm (green lines). But the caliper
measures are showing how increasing angle between M-mode
line and direction of motion increases the
overestimation of the MAPSE. |
Again measuring displacement relative to end diastolic wall
length, will give correct values, as both wall length and
displacement will have the same ratio despite the angulation
error, if measured along the same straight line. Thus, global
strain in not affected to the same degree.
This means, that in motion tracking by B-mode or M-mode, the
Normal long axis image. The
motion of the base of the ventricle towards the apex
is evident in the long axis view. |
Looking at the short axis view
from the base, this is not evident, but comparing
with the image on the left, this mus mean that
during systole, an entirely new part of the
ventricle moves into the imaging plane. |
As can be seen, the base of the
heart moves through the M-mode line during the heart
cycle. |
This means that measurements in
fact are taken from different part of the ventricle
in end diastolie and end systole. It seems to
indicate that systolic measurements are done in a
part of the ventricle with narrower lumen and
thicker wall, thus may over estimating both fractional
shortening and wall
thickening. |
Interestingly, the M-mode values of HUNT3 showed a
substantial higher wall thickening in the PW than in the
septum, while the 2D measurements in HUNT 4 did not reproduce
this finding. This effect is probably due to the specific
vulnerability of M-mode to the effects of the long axis
shortening, making the M-mode cress different parts of the LV
in systole and diastole. The configuration of the posterior
wall in then base may thus induce a statistical bias towards
over estimation of wall thickness as shown below.
Comparing with longitudinal
deformation of the two walls, we found in HUNT3 that
MAPSE was about 14% higher in the posterior wall than the
anteroseptum, but the posterior wall was also around 10%
longer than the anteroseptum (156).
Thus, the relative shortening (longitudinal wall strain) in
HUNT 3 was 16.6% in the anteroseptum, vs 16.5% by segmental
strain, and 14.7% vs 15.5% (relative difference 5%)
by normalised MAPSE.
Again, modern technology now allows a much more complex
processing technology allows using input data in a way that
also improves the beamforming characteristics in processing,
as they are used for the generation of a picture. Thus the
simple principles of beamforming outlined here are an over
simplification compared to the most advanced high end
scanners.
It is important to realise that the last
couple of years has seen tremendous improvements in both
hardware (allowing a much higher data input to the
scanner as well as processing technology), and software
(allowing more data processing at higher speed).
It even allow using input data in a way that also
improves the beamforming characteristics in processing,
as they are used for the generation of a picture. Thus
the simple principles of beamforming and focussing
outlined here are an over simplification compared to the
most advanced high end scanners.
However, they will still serve to give an idea. And
simpler equipment still conform more closely to
the basic principles described here.
A. Mechanical
transducer. The sector is formed by rotating a single
transducer or array of transducers mechanically,
firing one pulse in each direction and then
waiting for the return pulse before rotating the
transducer one step. In this beam there is electronic
focusing as well, by an annular array. |
B. Electronic
transducer in a phased array. By stimulating the
transducers in a rapid sequence , the ultrasound will
be sent out in an interference pattern. According to
Huygens principle, the wavefront will behave as a
single beam, thus the beam is formed by all
transducers in the array, and the direction is
determined by the time sequence of the pulses sent to
the array. Thus, the beam can be electronically
steeredand will then sweep stepwise over the sector in
the same way as the mechanical transducer in A,
sending a beam in one direction at a time. |
Dynamic focusing.
The same principle of phase steering can be applied to
make a concave wavefront, resulting in focusing
of the beam with its narrowest part a distance
from the probe. Combining the steering in B and
C will result in a focussed beam that sweeps across
the sector, as in the moving image above. |
Resulting Ultrasound beam as shown by a computer simulation, focusing due to the concave wavefront created by the dynamic focusing. The wavelength is exaggerated for illustration purposes. Image Courtesy of Hans Torp. |
Annular focusing
in all directions both in plane and transverse to
the plane.
|
Linear focusing
in the imaging plane only.
|
The lateral
resolution of a beam is dependent on the focal depth,
the wavelength and probe diameter (aperture) of
the ultrasound probe. A near shadow will reduce the
effective aperture, and thus the lateral resilution as
illustrated here. (Reproduced from Hans Torp by permission) |
Two points in a sector that is to be scanned. | The ultrasound scan will smear the points out according to the lateral resolution in each beam. |
Thus a small scatterer
will appear to be "smeared out", and the apparent
size in the image is determined by the beam width
and pulse length. As the pulse length is
less than the beam width, the object will be "smeared out" most in the lateral direction. |
Two scatterers at the
same depth, separated laterally by less than the
beam width, will appear as one. |
Two scatterers at different
depths will appear separate if separated by more
than the pulse length. |
But, if separated both
laterally and in depth, they will appear as being in
the same line, if lateral separation is within the
beam. |
Top, a common reverberation in
the lateral wall, seen as a stationary echo
(arrows). Below, the principle shown
diagrammatically, a reflector causing the ultrasound
pulse to bounce, for each bounce back, the echo is
interpreted as a structure at a depth corresponding
to multiples of the original depth. |
This is even more evident in
this image, showing multiple, stationary
reverberations from the apex. All the reverberations
have the same distance. In the blow up below, the
reverberation space can be seen to be a echolucent
space in front of the apical pericardium, and the
distance between the reverberations equals the
original distance between the probe and the
pericardium. |
The red jet shown in the
atrium, is a reverberation originating from the
aortic regurgitation jet. |
From the B-mode acquisition,
there can be seen a slight, possibly clutter line as
well, buty in this case the reverberation signal is
predominantly in the Doppler signal. |
To
document that this is not a pathological jet, the
apical long axis and four-chamber views do not show
such a jet in the same location. |
The simultaneous duration of
the two jets shown on the reconstructed M-mode also
confirms that this is a reflection, and not
something else (f.i. a venous signal or fistula) |
The distance between the jets
is compatible with the reflecting layer being the
immovable structure outside the pericardium. |
-and quantitative analysis
shows the reversal of the phase in the reflected
signal. |
The comet-tail artefact is used to describe the
ring down phenomenon doing utlrasound of the lungs, with a
cardiac probe (281).
This has been seen to be a marker of interstitial fluid in
the lungs (282),
i.e. edematous interstitial septa (equivalent to the
Kerley B-lines on X-ray), and has been seen to be quick
and reliable. The reverberations should then be within
the edematous interstitial septa, as air filled alveolar
clusters in front and behind would be strong reflectors,
causing the reverberation within a very short distance (283).
As penetration through the lung is poor, they have to
originate close to the lung surface:
Ring
down echoes from the pericardium. They can be seen
as bright bands radiating
down, and the source seem to be real, as the
ring down beams are visible both in long and short
axis views from the same patient.
The reverberating space is probably the
pericardial space itself. The uneven
distribution of the ring down beams may be due to
the varying reflectivity due to different directions
of the surfaces relative to the transmitted beams. |
|
Ring down beam
seen to originate from the apicolateral pericardium.
As with sidelobes, in this case the shadow is not
constant, probably due to the source moving
in and out of the plane. |
Parasternal image from a
patient with a mechanic aortic valve, combining
shadows and ring down shadows. The thick
metal ring itself gives rise to an ordinary shadow
from the anterior part,, while the thin part of the
carbon fibre ring protruding out into the sinus
valsalvae, gives rise to a ring down beam. The
reverberating space may be the sinus in front of the
protruding carbon ring. |
Heavy reverberation band across this long axis image. The shadow is not ditinct, and thus far less coherent than the examples above. | Shadowy
reverberations covering the naterior wall in
this 2-chamber image. It is differentiated from
the drop out shown above,
as we can se a "fog" of structures covering the
anterior wall. The structures are stationary. On
the other hand, this is not distinct
reverberations shadows, but incoherent clutter. |
Simulated beam with
focusing, showing interference pattern dispersing
some of the beam to the sides. (image courtesy of
Hans Torp). |
Side lobes from a single focussed ultrasound beam. These side lobes will also generate echoes from a scatterer hit by the ultrasound energy in the side lobes, i.e. outside the main beam. |
As echoes from a scatterer in the side lobe pathway is perceived coming from the main beam, this will result in a false echo, apparently from the main beam.. | AS the beam with side lobes sweeps back and forth a cross the sector, each echo from the scatterer in both the main beam and the side lobes will generate the false echo in the position of the main beam. | This again will result in the echo being smeared out across the sector, resulting in a smeared out echo across a large part of the sector. |
Side lobes
originating from the fusion line of the aortic
cusps, seen to extend into both the LV cavity
and the aortic root cavity (arrows). |
As opposed to reverberations,
the side lobes moves with the structure, and
may change with time (in this case the echo
intensity of the fusion line decreases as the
valve opens, and thus the intensity of the
side lobes too) . |
A pulse is sent out, ultrasound is reflected, and the B-mode line is built up from the reflected signals. | Linear array. |
Curvilinear array |
By making the ultrasound beam
sweep over a sector, the image can be made to build up
an image, consisting of multiple B-mode lines. |
c. In principle, the image is built up line by line, by emitting the pulse, waiting for the reflected echoes before tilting the beam and emitting the next pulse. Resulting in an image being built up with a whole frame taking the time for emitting the total number of pulses corresponding to the total number of lines in the image. |
Two different lateral
resolutions, the speckles can be seen to be "smeared".
In this case the loss of resolution in the right image
is due to lower line density . By rights the image
should appear as split in different lines as indicated
in the middle, as each beam is separated, line density
being less than optimal relative to the beam width.
Instead the image is interpolated beween lines. This
reduction in line density is done to achieve a higher
frame rate, as illustrated below. |
|
||||
A:
Beam
width. Speckles (true speckles: black) are
smeared out across the whole beam width ( Apparent
speckles dark grey, top). This means that with this
beam width the speckles from to different layers
cannot be differentiated, and layer specific motion
cannot be tracked. |
B:
Line
density. Only the lines in the ultrasound beams
(black) are detected, and can be tracked, beams
between lines are not detected or tracked. The spaces
between lines cannot be seen in the final image due to
image lateral
smoothing. |
C:
Divergence
of lines in the depth due to the sector image will
both increase beam width and decrease line density in
the far field. this may result in the line density and
width being adequate (in this example for two layer
tracking) in the near field, but inadequate in the far
field, situation there being analoguous to A. |
D:
Focussing.
The beams being focussed at a certain depth mau mean
that line density may be inadequate at the focus
depth. Thus speckles in some layers may be missed. IN
general, the default setting will usually give the
best line density at the focus depth, so unless frame
rate is increased, this problem may be minor.
Howewever, line density will decrease ifalso if sector
width is increased, there is a given number of lines
for a given frame rate and depth. In any case, in the
far field, the beams will be broader, and the beam
width will be more like A and C. |
E:
Focussing may even result in beams overlapping int the
far field. A speckle in the overlap zone may be
smeared out across two
beams. |
As the depth
of the sector determines the time before next pulse
can be sent out, higher depth results in longer time
for building each line, and thus longer time for
building the sector from a given number of lines, i.e.
lower frame rate. |
Thus reducing the
desired depth of the sector results in shorter time
between pulses, and thus shorter time for building
each line, shorter time for building the same number
of lines, i.e. higher frame rate. In this case, the
depth has been halved, and the time for building a
line is also halved. |
In this case, in the image to the left, the depth has been halved, reducing the time for building each line to half, thus also reducing the time for building the full sector, increasing the frame rate. |
A sector with a given depth, sector width and line density determines the frame rate. | Reducing sector width, but maintaining the line density, gives unchanged lateral resolution but higher frame rate, at the cost of field of view. | Reducing the line density instead and maintaining sector width, results in lower number of lines, i.e. lateral resolution, and gives the same increase in frame rate. |
MLA angle discrepancy. The width of one transmit beam is exaggerated for visualisation. One wide beam is transmitted, and four narrow recieve beams. The transmit beam has has a main direction shown by the red arrow. The receive beams has directions (blue arrows) with an angle to the transmit beam, and this angle increases with increasing distance of the receive beam from the middle of the transmit, i.e. with the MLA factor. | MLA angle artefacts in B-mode. Left: single line acquistion, where frame rate is acquired by a fairly low line density, and the image is then smoothed with interpolation between scanlines as described above, right, 4MLA acquisition. This should in principle result in a quadrupling of the number of lines, and an image with better lateral resolution. However, the increasing angle deviation between the Tx beam and the RX beams in the lateral parts of the Tx, will result in the lines being visible as blocks, the improvement in image quality being negligible or none. Image courtesy of Tore Bjaastad. |
3D ultrasound increases complexity a lot, resulting in a new
set of additional challenges.
The number of crystals need to be increased, typically from
between 64 and 128 to between 2000 and 3000. However, the
probe footprint still needs to be no bigger than being able to
fit between the ribs. And the aperture
size must still be adequate for image resolution.
The number of data channels increases also by the square,
from 64 to 642 = 4096. This means that the
transmission capacity of the probe connector needs to be
substatially increased, and some processing has to take place
in the probe itself to reduce number of transmission channels.
.
The number of lines also increase by the square of the number
for 2D, given the same line density, meaning that each
plane shall have the same number of lines, and a full volume
then shall be n=built by the same number of planes. This means
that given 64 lines per plane, the number of planes should be
64, which means a total of 64 x 64 = 4096 lines. This means
that the frame rate (usually termed the "volume rate" in 3D
imaging), will be 0.19 ms x 4096 = 778 ms, or about 0.8 secs.
Meaning about 1 volume per heartbeat for a heart rate of 75.
This is illustrated below.
Building a 2D sector with
lines. (Even though each line (and the sector) has a
definite thickness, this is usually not considered
in 2D imaging, except in beamforming for image
quality. |
Building a 3D volume. Each
plane has the same number of lines as in the 2D
sector to the left, and takes as long to build. The
number of planes equals the number of lines in each
plane. Here is shown only the building of the first
plane (compare with left), but the time spent on
each of the following planes are in proportoion. The
time for a full volume is then equal to the square
of the number of lines in each plane. |
Surface rendering of a 3D
volume. The image shows a cut through the LV between
base and apex, looking down toward the base, the
papillary muscles and mitral valve can be seen.
The illustration also shows that the temporal
resolution is to low to actually show the opening of
the mitral valve during trial systole, only a slight
flicker can be seen at end diastole. |
The same volume, now displayed
as a series of short axis slices
from apex (top left) to base (bottom right).
A slight stiching artefact (spatial
discontinuity) can be seen in the anterior wall (top
of each slice). |
3D acquisition of a ventricle
with inferior infarct. The display is shown as the
apical planes to the left, and nine cross sectional
planes to the right, going from the apex (top left)
to the base (bottom right - reading order). The
infarct can be seen as inferoseptal a - to
dyskinesia in the basal sections. The image also
illustrates that the software can be enabled to
track the planes, thus eliminating out of plane
artefacts when evaluating wall motion. Note
that there is drop outs that cannot be eliminated by
moving the imaging plane, in the anterior wall.
Image courtesy of Dr. A. Thorstensen . |
Styitching artefacts. In this
volume, reconstructed from four heartbeats, i.e.
four sub volumes, there are stitching artefacts
between each of the sub volumes. This is due
to motion of either the heart (f.i.) because of
respiration, or of the probe. In the inferior wall
(bottom of each slice), the spatial discontinuity
is very evident, less so at the other
stiches,, but in the anterior wall there is a
discontinuity that illudes a dyssynergy. |
Fourier analysis of the resulting signal in native frequency (left) and second harmonic mode (left) shows that the native signal contains much more energy at all depth, while the harmonic signal contains most of the energy at a certain depth, in this case at the level of the septum, showing a much better signal-to-noise ratio.(image courtesy of Hans Torp). | Energy distribution of the signal from cavity (lower curve) and septum (upper curve), showing the same phenomenon as the middle picture. The difference between cavity signal (being mostly clutter) and tissue is small in the native frequency domain (1.7 MHz), but there is little clutter at the harmonic frequency (3.4 MHz). Thus, filtering the native signal will reduce clutter, as shown below. (image courtesy of Hans Torp). |
The same image
in harmonic (left) and fundamental (right) mode, showing
the improved signal-to-noise ratio in harmonic
imaging, especially in rducing noise from the
cavity. (Thanks to Eirik Nestaas for
correcting my left-right confusion in this image
text) |
Stationary
reverberation in harmonic (left) and fundamental
(right) imaging, showing the effect of harmonic
imaging on clutter.
|
In the present ultrasound the methods available for multiple
sites motion measurement, are speckle tracking and colour
tissue Doppler.
Interference
pattern. Here is simulated two wave sources or scatterers
at the far field (white points). The emitted or reflected
waves are seen to generate a speckle pattern (oval dots)
as the amplitude is increased where wave crests cross each
other, while the waves are neutralised where a wave crest
crosses a though. This can be seen by throwing two stones
simultaneously in still water . The speckle pattern can be
seen in front of the scatterers, towards the probe. |
Irregular interference pattern. This is generated by more scatterers somewhat randomly distributed. The speckle pattern is thus random too. Again there may be a considerable distance between the speckles and the scatterers generating the pattern. |
1. The randomness of the speckle pattern ensures that each region of the myocardium has its own unique speckle pattern: that can differentiate a region from other region | 2. The speckle
pattern remains reasonably stable, and the speckles
follow the myocardial motion. This can be demonstrated
by M-mode, showing how the speckle pattern follows the
myocardial motion. |
Defining a kernel
in the myocardium will define a speckle pattern within
(red). Within a defined search area ( blue), the new
position of the kernel in the next frame (green) can be
recognised by finding the matchin speckle pattern in a
new position. The movement of the kernel (thick
blue arrow) can then be measured. |
Speckle tracking
search algorithm. The kernel is defined in the original
frame at t=0 (red square). In the next frame, at t=t,
the algorithm defines a search area (white square), and
the search is conducted in all directions for the
matching kernel. |
This is dependent on both line width (being dependent again on
frequency and focus depth), and line density (being dependent on frame rate
and sector width). In addition, there has to be adequate alignment
of the ultrasound beam, as angle deviation will reduce the number of
lines within the wall in the far field. This is true for speckle
tracking as well as tissue Doppler. Finally, focusing
of the beams will result in different lateral resolution at
different depths.
Speckle tracking has the advantage of a higher line
density of B-mode, at the cost of a lower temporal
resolution. The very low lateral
resolution used in tissue Doppler in order to achieve a high
frame rate, results in a low line
density, and in practice limits the measurement in the beams
in the longitudinal (and tangential - for circimferential
measures) direction to the entire wall thickness, for a standard
set up.
|
||||
A:
Beam
width. Speckles (true speckles: black) are smeared
out across the whole beam width ( Apparent speckles dark
grey, top). This means that with this beam width the
speckles from to different layers cannot be
differentiated, and layer specific motion cannot be
tracked. |
B:
Line
density. Only the lines in the ultrasound beams
(black) are detected, and can be tracked, beams between
lines are not detected or tracked.and differential mtion
of the two speckles cannot be tracked. The spaces between
lines cannot be seen doe to image lateral
smoothing. |
C:
Divergence
of lines in the depth due to the sector image will both
increase beam width and decrease line density in the far
field. this may result in the line density and width being
adequate (in this example for two layer tracking) in the
near field, but inadequate in the far field, situation
there being analoguous to A. |
D:
Focussing.
The beams being focussed at a certain depth mau mean that
line density may be inadequate at the focus depth. Thus
speckles in some layers may be missed. IN general, the
default setting will usually give the best line density at
the focus depth, so unless frame rate is increased, this
problem may be minor. Howewever, line density will
decrease ifalso if sector width is increased, there is a
given number of lines for a given frame rate and depth. In
any case, in the far field, the beams will be broader, and
the beam width will be more like A and C. |
E:
Focussing may even result in beams overlapping int the far
field. A speckle in the overlap zone may be smeared out
across two beams. |
Longitudinal | Transverse |
|
Strain rate |
||
Strain |
LOngitudinal speckle tracking in apical 4 chamber view. The resulting tracking of the kernels shown in motion. As can be seen, with a drop out apicolateral, this ROI tracks less than perfect, giving too low strain both in LA and MA segments. | Speckle tracking can be applied crosswise. In this parasternal long axis view, the myocardial motion is tracked both in axial and transverse (longitudinal) direction. It is evident that the tracking is far poorer in the inferior wall, due to the poor lateral resolution at greater depth. |
Reverberation in the lateral
wall affecting speckle tracking. As is visually evident, the
application does not track across the reverberation, thus the
two segments apical to the reverberations are seen as akinetic,
the basal as hyperkinetic. All shortening is seen in the basal
segment. In this case, the smoothing
is seen to spread the effect of the reverberation out across two segments apical
to the reverberation.
Effect of a reverberation on
the border between the apical lateral and the midwall
lateral segment. A kernel in this area will not
track, as illustrated by the arrow. The next
border between the basal and midwall segment moves
normally, leading to an exaggerated shortening of the
midwall segment, while the basal segment shortens
normally. (The segmental strain in the apex is the
difference between the apical motion (zero) and
the apparent motion ( near zero) in the reverberation,
the midwall strain is the difference between the apparent motion ( near
zero) in the reverberation and the (normal) motion of
the border below.) This is evident by the curves
(compare to the average curve: The apical curve shows
little strain, the midwall curve shows far more than the
wall average, and the basal shows average strain).
Compare with the image above.
Image courtesy of H Dahlen. |
Christian Andreas Doppler | Christophorus Henricus Diedericus Buys-Ballot |
Über das farbige Licht der Doppelsterne
und einiger anderer Gestirne des Himmels. Abhandlungen der
königl. Böhm. Gesellschaft der Wissenschaften. 1843; 2:
465-82 Versuch einer das Bradely'sche Aberrations-theorem als integrirenden theil in sich schliessenden allgemeineren Theorie |
Akustische Versuche auf der Niederländischen Eisenbahn nebst gelegentlichen Bemerkungen zur Theorie des Hrn. Prof. Doppler. Poggendorfs Annalen der Phüsik und Chemie 1845; 66: 321-351 |
The Doppler effect
for a moving source and a stationary observer.
In the time the original wave has moved a
wavelength (which is 1 / f0),
the source has moved the distance closer to the
observer, determined by the velocity V
of the source. The next wave will then meet
the observer after the distance - ,
corresponding to the perceived wavelength . |
The Doppler effect for a stationary wave source and a moving observer. In the time the wave has moved the distance , the observer has moved the distance closer to the source, determined by the velocity v of the observer, and will meet the wavefront earlier , corresponding to the perceived wavelength , which is - . |
If the
source moves toward a stationary observer In the time the wave moves one
wavelength, the source moves the distance:
The motion of the wave and the motion of the source happen during the same time interval (the time interval of one wave): The distance from the next wave emitted from the new position of the source (small dotted red circle) to the observer (blue) is shortened by in the direction of the motion, so the new wavelength representing the distance between the first and second waves is and thus:
and:
the change in frequency, the
Doppler shift by the velocity of a moving source
is:
As we see, there is a difference between a moving source and a moving observer, but if v<< c, Then for a moving source: (The approximation is small, the velocity of ultrasound in tissue is 1540 m/s, while the velocity of blood is between 0,2 and 8 m/s, and tissue between 0.05 and 0.2 m/s, giving a v/c of maximum 0.005, i.e the approximation is maximum 0.5% and in reflected ultrasound 0.2%). |
An observer (blue) moving towards a
stationary wavesource with the velocity: will meet the wave as the wave
have moved a distance , which is the
perceived wavelength. The observer has moved the
distance:
The motion of the wave and the motion of the observer happen during the same time interval (the time interval of the perceived wavelength ), which is: The
motion of the observer thus shortens the
original wavelength by , , so the new wavelength
representing the distance between the first
and second waves is
and thus:
And the change in frequency,
the Doppler shift by the velocity of a
moving observer is:
|
Mitral flow, showing a fairly narrow
spectrum band, indicating a relatively
homogeneous velocity distribution within the
sample volume, which is placed between the
tip of the mitral cusps during filling,
where the inflow jet is most narrow.. |
Pulmonary venous flow in the same
subject, showing a wide distribution of
velocities, within the sample volume placed
in the right upper pulmonary vein. The
sample volume is the same size.Venous
velocities are much lower, but also varies
from 0 to 0.5 m/s simultaneously. |
Phase analysis. If the waveform is treated as a sine curve, every point on the curve corresponds to an angle, and the phase of the point in the curve can be described by this angle; the phase angle . From the diagram, it's also evident that a full wavelength, , is equivalent to 2, and for every point the corresponding fraction of a wavelength is equivalent to an angle which is the fraction of 2. | However, from the diagram at the top, it is evident that by sampling the waveform only once, the phase is ambiguous, it is not possible to separate the phase of point a from point b. The two points are separated by a quarter of a wavelength, or 90° (). In order to determine the phase of the points unambiguously, the pulse has to be sampled at to points separated by less than a quarter wavelength. Then it can be seen that point a is in increasing phase from a1 to a2, corresponding to a phase angle of 0 - /2 while b is in a decreasing phase corresponding to an angle of /2 - . |
A series of pulses shot
successively. It is also evident the even
without motion, there is a phase shift
between pulses, but this is equal for
transmitted and reflected ultrasound |
Measuring the velocity
of an object by phase analysis. The velocity
of the scatterer is shown by the dotted red
line, showing the phase in each pulse, and
then the phase shift through the pulse
package is illustrated by the full sinusoid
red line shown to the left, where the
troughs and peaks of the red line represents
the scatterer's position at the peaks and
troughs of each pulse, i.e. the phase in
trelation to the twop pulses. In principle,
the phase shift can be sampled between each
pulse pair. The sinusoid curve is the phase
shift curve, and the frequency is equal to
the Doppler frequency. |
Cw Doppler signal from
LVOT. The velocities can be seen to be
present in all ranges, although the peak
velocities are close to the curve shown by
Pw Doppler to the left. The reason for the
wider distribution, is that all velocities
from the whole LVOT acceleration zone are
displayed. In addition, a part of the mitral
flow can be seen as well, showing that the
overlap area of the two cw sectors is less
well focused. |
Pw Doppler signal from
the same LVOT. Most of the velocities can be
seen to be collected in a narrow band,
roughly corresponding to the peak velocities
of the cw Doppler. Also, there is less
contamination from the mitral flow, as the
sample volume (range gate) is in the focused
part of the beam. |
In pw Doppler
flow, modal velocity is the middle of the
band, and represents the mean velocity
within the sample volume. |
Pw Doppler recording
from aorta descendens. Gain is set so high
that the thermal noise is very visible. The
main (modal) velocities is shown to be in a
saturated band. |
Same recording at low
gain. The modal velocities is still visible,
while the less intense velocities above that
are not visible, and the band is narrower,
thus the peak velocities will be slightly
lower. |
Mitral flow in high
gain. Around the main spectrum is seen some
noise spikes. |
Same recording in low
gain, removes noise spikes. |
As a reflector moves
from a to b in the direction 1, the true
motion (displacement) is L1. If the ultrasound beam
deviates from the direction of the motion
by the angle , the apparent length along the ultrasound beam will be L2, which is the hypothenuse of the triangle, and thus . Thus angle deviation of M-mode measures will always over estimate the real motion (as opposed to Doppler measurements). |
The angle error in displacement measurement demonstrated in a reconstructed M-mode. Increasing angle between M-mode line and direction of motion increases the overestimation of the MAPSE. |
Basically, the measured velocities decrease by the cosine function, being 0 at 90° insonation: | Angle distortion in
Doppler. The image on the left has applied
angle correction, and then adjusted to
scale.This is only partly correct, the ange
distortion will also broaden, not narrow the
spectrum as shown here. |
Depth |
SI |
PRF |
1 cm |
0.000013s |
77000 Hz |
5 cm |
0.00006s |
15400 Hz |
10 cm |
0.00013s |
7700 Hz |
15 cm |
0.00019s |
5133 Hz |
20 cm |
0.00026 s |
3850 Hz |
Phase shift analysis.
The distance between the pulses represent
the pulse interval, or 1/pulse repetition
frequency (1/PRF) |
This means that the
phase shift curve can be sampled only with a
frequency equal to the PRF.
Halving the pulse repetition frequency
doubles the sampling interval. This results
in the Doppler shift curve being sampled at
half the number of pulses. This may result
in velocity ambiguity as described below. |
A more rapidly moving scatterer will then result in a higher frequency of the phase shift, i.e. a higher Doppler frequency. Thus the frequency is proportional to the velocity. It also means that the phase shift curve is sampled fewer times per oscillation, giving an equivalent effect as reducing the PRF.. |
The Nykvist phenomenon (231) is an effect of the relation between the sampling frequency and the observed velocity. If you sample at a certain frequency, the direction of the motion becomes ambiguous, more frequent sampling will give the correct direction, less frequent sampling results in an apparent motion in the opposite direction. This can be observed with a stroboscopic light, for instance illuminating the flow of water
Cw Doppler, sampling
the phase shift curve (Dopplwer frequency)
once per pulse. The curve is very well
reproduced. |
Pw Doppler samples the
curve with much lower sampling frequency
(PRF), but still sufficient so the curve can
be reproduced, both the value and direction
of the velocity can be measured. |
Pw Doppler where
sampling frequency (PRF) is 4 × the Doppler
frequency. The curve will still reproduce
the troughs and peaks of the curve, and the
information of the direction, and doesn't
fit the alternate curve (same frequency, but
out of phase (corresponding the the same
velocity in the opposite direction. |
Sampling at 2 × the
Doppler frequency (i.e.) twice per
oscillation, PRF = 2
× fd, the curve cannot be reproduced, i.e.
the Doppler frequency cannot be measured,
the samples fit both curves equally well,
and the velocity direction is ambiguous. |
- and the samples fit
equally well the double Doppler frequency,
i.e. twice the velocity. |
Finally when PRF is
< 2 × Fd, the sampling will fit other
velocities as well, in this case 1.5 ×
velocity. |
Constant rotation velocity, decreasing sampling frequency:The easiest is to
show how reducing the sampling frequency
affects the apparent motion. All circles
rotate with the same rotation velocity
clockwise. The sampling frequency is reduced
from left to right. It can be seen that the
red dots is at the same positions when they
are seen to move.
|
|||
a:
8:1 8 samples per rotation, the red point is seen in eight positions during the rotation. |
b:
4:1 4 samples per rotation, the red point is seen to rotate just as fast, but is only seen in four positions |
c: 2:1 2 samples per rotation, i.e. the sampling frequency is exactly half the rotation frequency. Here, the red dot is only seen in two positions, (but it is evident that it is in the same positions at the same time as in a and b). However, it is impossible to decide which way it is rotating. This is the Nykvist limit; sampling rate = 1/2 rotation rate. |
d:
1.5:1 1.5 samples per rotation,or one sample per three quarter rotation, making it seem that the red dot is rotating counter clockwise. Again, the dot is in the same position at the same time as in a and b. |
|
|||
a:
1:8 One rotation per 8 samples. The sampling catches the red dot in 8 positions during one rotation. |
b:
1:4 Rotation velocity twice that i a; one rotation per four samples, the sampling catches the red dot only in four positions during one rotation. |
c: 1:2 Rotation velocity four times a; one rotation per two samples, this catches the red dot in only two positions, giving directional ambiguity as above. |
d: 1:1,5 Rotation velocity six times a; one rotation per 1,5 samples, or 3/4 rotation per sample, giving an apparent counter clockwise rotation. |
Sampling from increasing depth
will increase the time for the pulse returning,
thus increasing the sampling interval and decrease the
sampling frequency. The Nykvist limit thus
decreases with depth. This means that pulsed Doppler
has depth resolution, but this leads to a limit to the
velocities that can be measured.
Frequency aliasing occurs at a Doppler shift
that is equal to half of the PRF. fD = ½ ×
PRF, i.e. two samples per wavelength, as described
above.
fDmax =
½ × PRF
vmax = c × PRF / 4 f0 cos()
Depth |
Maximum (Nykvist)
velocity |
||
Transmit frequency (f0) |
2 MHz |
5 MHz |
10 MHz |
1 cm |
1480 cm/s |
590 cm/s | 295 cm/s |
5 cm |
295 cm/s | 120 cm/s | 60 cm/s |
10 cm |
150 cm/s | 60 cm/s | 30 cm/s |
15 cm |
100 cm/s | 40 cm/s | 15 cm/s |
20 cm |
75 cm/s | 30 cm/s | 15 cm/s |
Pulsed wave LVOT flow
velocity curve,
sampled at adequate PRF, corresponding to a
Nykvist limit of ca 1.2 m/s. |
The same LVOT flow
velocity curve sampled at too low PRF,
corresponding to a Nykvist limit of ca 0.6
m/s. Aliasing is evident. both positive and
negative velocities are present. |
Aorta flow velocity
curve sampled at same PRF. By baseline
adjustment, the limit for aliasing can be
adjusted to 2× Nykvist, (but at the cost of
total aliasing in the other direction. |
If the velocities are much higher
than the Nykvist, aliasing will occur at many
multiples of the Doppler freqency:
Aortic insufficiency shown by cw Doppler. It van be seen that there are a fair distribution of velocities in the whole spectrum. However, There are far more velocities below 2 m/s. In this case, the low pass filter is only set to suppress tissue velocities. If the point is to get a clear visualisation of the maximal velocities in the jet, at 4 - 6 m/s, the filter should be set higher. | The same patient by pulsed Doppler of the LVOT. The outflow can be seen as a narrow band, within the velocity range, while the regurgitant jet has velocities far outside the Nykvist range, and there is total velocity ambiguity. |
The principle of HPRF. Pulses are transmitted with three times the frequency that is necessary to allow the echo from the furthest depth to return. Thus, the echo of pulse 1 will return from level 3 at the same time as the echo of pulse 2 from level 2 and and of pulse 3 from level 1, and there is no way to determine whether a signal is from level 1, 2 or 3. | HPRF pulsed
Doppler recording (right). with one
sample volume in mid ventricle and one
in the mitral ostium. The recording
shows a systolic dynamic gradient (due
to inotropic stimulation with
dobutamine), as well as an ordinary
mitral inflow curve. There is no
way in the pulsed recording to determine
which velocities that originate from
which sample volume (except from á
priori knowledge, of course, a dynamic
gradient like this is usually mid
ventricular, and the mitral inflow in
the annulus is easily recognised).
|
Amplitude imaging (B-mode). Tissue echoes have a high amplitude, blood a low amplitude (due to low reflexivity). Thus tissue is visualised , while the blood is not visible in the present gain setting. | As seen from
the B-mode image to the left, the
tissue is not stationary, but the
velocities are low, compared to
blood. |
Too see blood
flow velocities, it is possible to
increase gain, but that will also
increase the low velocity signals
from the tissue to saturation. They
are generally considered clutter
nois, when dealing with blood flow,
although it is not true
clutter in the reverberation sense.
|
It is possible to filter the low velocities by a "high pass filter", that allows high velocities to pass, while removing low velocity signals independent of amplitude. This will remove both tissue signals and reverberation noise. The low velocities around the baseline have been removed, the width of the filter is indicated by the green band to the left. This is also called "clutter filter" or "low velocity reject". |
As
described above, a pulse has a certein
bandwidth,
describing the frequency content of
the pulse. In spectral analysis, this
will give a spectrum of a certain
width, corresponding to the velocity
distribution of flow velocities. In
phase analysis, this will correspont
to a certain distribution of phase
angles as illustrated.
Autocorrelation, however, will only
result in the average phase angle.
|
In the case
of stationary noise (clutter) as f.i.
reverberations, the autocorrelation will
result in an average phase angle that is
in between the signal and the noise. The
clutter noise will have to be removed by
a low velocity filter in order to avoid
severe underestimation of flow
velocities. |
In
colour Doppler one pulse package
is sent out as in Pw Doppler, but
the return signal is sampled
multiple times as in B-mode. Since
there is only one transmit pulse
(package) at a time, there is no
range ambiguity, each return
sample corresponds to one specific
deph, as in B-mode.. |
Relation
between PRF and frame rate. The
diagram illustrates a scatterer
moving in a Doppler field. In
order to do phase analysis, at
least two pulses (a pulse package)
need to be sent out along one
line, the time between them
corresponding to the PRF, which
again is limited by the maximum
depth of the colour sector. When
the Doppler shifts have been
sampled along one line by a pulse
package, a new pulse package is
sent out along the neighboring
line, building a sector image
analogous to B-mode. Thus, the
position of the scatterer can be
seen to be sampled only with the
frame rate, which is lower than
the PRF, depending on the depth,
width and resolution of the colour
sector. |
CFM sector superposed on a B-mode sector. By reducing sector size, line density and sampling frequency, the CFM image can achieve an acceptable frame rate. This is feasible because the region of interest for the flow is usually only a part of the ROI for The B-mode, flow being intracavitary as shown below. |
Power
Doppler image of the renal
circulation. The amplitude is a
function of the number of
scatterers, i.e. the number of
blood cells with a Doppler shift.
This is shown as the brightness
(hue) of the signal. In addition,
direction of
flow can be imaged by different
colours (red - positive flow -
towards probe, blue - negative
colours - away from probe), and
still the brightness may show the
amplitude. |
Colour
flow showing a large mitral
regurgitation. Velocities away
from the probe is shown in blue
(converting to red where there is
aliasing), towards the probe is
red. In this image, the green
colour is used to show the spread
(variance) of velocities. This
will also reflect areas of high
velocities (high variance due to
aliasing). |
Recording from a patient with apical hypertrophic cardiomyopathy. Ejection can be seen in blue, and there is a delayed, separate ejection from the apex due to delayed relaxation. There is an ordinary mitral inflow (red), but no filling of the apex in the early phase (E-wave), while the late phase (A-wave) can be seen to fill the apex. Left, a combined image in HPRF and colour M-mode. The PRF is adjusted to place two samples at thr mitral annulus and in the mid ventricle just at the outlet of the apex. The mitral filling is shown by the green arrows, and the late filling of the apex is marked by the blue arrow. In addition, theere is a dynamic mid ventricular gradient shown by the red arrow, with aliasing in the ejection signal in colur Doppler. The delayed ejection from the apex is marked by the yellow arrow (the case is described in (87). The utility of the different methods is evident: HPRF (or cw Doppler) for timing and velocity measurement, but with depth ambiguity, colour M-mode for timing and location of the different jets, direction being displayed by the colour. |
The phase analysis is often done by the
process known as autocorrelation.
This will result in a values that does not
reflect the spectrum, but only mean values in
the spectrum. But if there is clutter in the
region (stationary echoes), this will be
incorporated in the mean, resulting ion lower
values. In Doppler flow, this can be filered
by the high pass filter, and thus will
represent a small problem. In tissue
Doppler, this may be a more significant
problem, as the velocities are only about 1/10
of the flow values, and thus clutter may be
more difficult to separate from true
velocities. Thus, a substantial amunt of
clutter may reduce autocorrelation values for
tissue Doppler more than pulsed Doppler as
discussed below.
In addition, it is customary to analyse the
tissue Doppler values in native, rather than
harmonic imaging, due to the Nykvist
limitation. Thus, there is a greater
amount of clutter than if harmonic imaging had
been used, as
shown
in B-mode images.
For
optimal colour flow, it is important to
realise that there may, in some scanners, be
an inverse relation between the gain of colour
Doppler and B-mode. (In some scanners it is
possible to adjust the priority, or to adjust
the gain settings separately). This, however,
is an acquisition finction, and not image
adjustment, and thus cannot be compensated
afterwards. This is illustrated below:
Effect on B-mode gain on colour Doppler
imaging. Left pulmonary venous flow by
pwDoppler, showing a systolic flow
component, although low velocities.
Middle, colour M-mode of the same patient.
Only the diastolic flow component can be
seen. Right, reducing B-mode gain
increases the gain of colour flow, and the
systolic pulmonary venous flow can be
seen.
This, however, holds only for constant flow
velocity. Blood flow is pulsatile, but the
fundamental equations of motion still hold:
As the area A1 is larger than A2, in order to push the same amount of blood through A2, the velocity v2 must be higher than v1. As the flow is the same, and given by A×v for continuous, and A×VTI for pulsatile flow, the ratio of velocities / velocity time integrals is the inverse of the ratio of areas. This is the continuity equation. | Using the continuity equation, as the LVOT diameter (and area) is known, tracing the VTI of the LVOT flow (pw Doppler to do it in the correct level) as well as the VTI through the valve (cw Doppler). The VTI equals the stroke length, and the stroke length times the atra, equals the stroke volume. As the stroke volume is constant, the two cylinders have equal volume, and thus, the valve stenosis area (AVA) can be calculated by AVA = LVOT area × VTILVOT / VTIAO |
Fundamentally, both velocity and pressure
represents energy. The potential energy in a
fluid under pressure, is given by E = P × V,
while the kinetic energy is E = ½ m v2.
But this means that when velocity increases,
this kinetic energy has to be recruited from
somewhere, which is the pressure energy.
Thus, as velocity increases, pressure has to
drop:
Left is
perfectly laminar flow through
the stenosis. In this case, the
post stenotic velocity
decelerates without energy loss,
and the kinetic energy is
converted back into pressure
again. Here, there is no net
pressure drop through the
stenosis. Driving pressure at P1
does not have to be increased to
maintain pressure at P3,
and the pressure drop at P2
is temporary. It must be
remarked, however, that pressure
recovery cannot be more than the
initial pressure. |
In the
middle is partial pressure
recovery. Some of the pressure
energy converted to kinetic
energy through the stenosis is
lost when the flow velocity
decelerates after the stenosis,
in the form of turbulence
resulting in friction. But some
of the energy is recovered to
pressure energy again. Thus,
there is a net gradient over the
stenosis, but this is less than
the maximum gradient. The
maximum gradient by Doppler will
over estimate the net gradient.
The red line represents the situation if the driving pressure is constant, thus the post stenotic pressure drops. The blue line represent the situation if P3 is regulated (as in the aorta). Then P1 has to be increased corresponding to Pnet in order to maintain P3. This represents the extrta work or load induced by the stenosis. |
Right,
there is total energy loss
through the stenosis, all
kinetic energy due to the
velocity increase through the
stenosis is lost in turbulence
and friction. Thus, Pmax
= Pnet
and the maximum gradient is a
measure of increased work (load)
in order to maintain P3.
|
As shown above, the tissue
velocities are present in the Doppler
spectrum, and with far higher amplitude
than the blood flow signals. In blood
flow Doppler, the tissue signals are
usually removed by the clutter
filter, in order to get cleaner
blood flow signals.
In Tissue Doppler , only the Pw mode is
useful, as velocities are so low, that
cw Doppler doesn't add anything useful,
while the location of the sampåle volume
is always important.
The
diagram to the left shows the
placement of flow and tissue
signals on this intensity
(amplitude) / velocity diagram.
Velocity given as the height ogf
the bars, intensity shown both
by the placement on the x axis,
as well as the darkness of the
bars, black being the highest
intensity. The flow signals are
low intensity but mostly high
velocity, while the tissue is
exclusively low velocity, high
intensity. The heart valves,
however, are solid structures
which moves with the velocity of
the passing blood, resulting in
high velocity signals. Intensity
may be seen to be higher.
A typical flow curve from the
LVOT ventricular outflow tract
is shown to the left, with the
valve click. |
|
Application
of a high pass filter (low
velocity reject) shown
schematically to the left and in
practice applied to the LVOT
flow curve to the right.
Velocities lower than the limits
of the green bar (showing the
range of the filter) are removed
seen in the dark zone in the
middle of the spectrum. The
setting rejects velocities below
15 cm/s. Wall velocities
are generally lower, and is
filtered. |
|
The
filter is adjustable, and is
here reduced below 10 cm/s. This
results in high intensity
signals becoming visible,
especially in early diastole.
This are tissue signals from the
LVOT. |
|
Further
reduction results in high
intensity tissue signals around
the baseline. The signal is
difficult to analyse, as it has
so high amplitude that the
display is saturated. |
|
Fully
decreasing the filter, and
decreasing the gain, (shown as
all signals being illustrated in
lighter colour, but with the
same relative placement on the x
axis), discloses the tissue
velocity curve, while the flow
signal, having a much lower
amplitude, is much less visible. |
|
Reducing the
scale, increases the resolution
of the tissue velocities, that
are still taken with ordinary
Doppler. |
|
All
modern ultrasound machines today
has separate applications for
tissue Doppler which optimises
the signal for this purpose,
among other things by applying a
low pass filter that removes
most of the flow velocities.
This results in a cleaner
signal. |
Recordings from basal septal mitral ring in a subject without substantial clutter. Spectral Doppler shows the dispersion of velocities, although this is probably an effect of bandwidth. The colour Doppler recording is superposed and aligned with both vertical and horizontal scale. In this instance can be seen to give values close to the middle of the spectrum (modal velocity). | Spectral Doppler
reconstructed from IQ data.
Candidates for measuring peak
systolic velocity by the PW
tissue Doppler spectrum. RED
circle: peak of the spectrum
at normal gain, GREEN circle:
upper edge of the strongest
part (the part visualised at
minimal gain), BLUE circle:
middle of the strongest part.
MAGENTA circle and line:
autocorrelation. As seen, imn
this example the
autocorrelation corresponds to
the middle of the spectrum.
(Figure courtesy of Svein Arne
Aase |
Reference method. A: pw Doppler from the mitral ring (reconstructed from RF data). Peak velocity of the ring displacement can be identified. B: This corresponds to the maximal slope of the M-mode line at the same time point. C: The M-mode in the same time window from the RF data. This gives a far better resolution in space and slope. D: In the RF M-mode the steepest sloe was identified automatically. This will be a reference for the maximal velocity. (Figure courtesy of Svein Arne Aase | As this figure shows, the peak spectrum results in a substantial overestimation. reducing the gain improves the over estimation, while the modal velocity is closest to the reference. Autocorrelation on the other hand results in significant under estimation, due to the presence of clutter. Only four subjects showed almost total correspondence between autocorrelation and modal velocity from spectral doppler. (Figure courtesy of Svein Arne Aase |
The relation between motion (velocity
and displacement) and deformation
(strain and strain rate) is treated here.
The concept
of velocity gradient was introduced by
Fleming et al (20),
defined as the slope
of the linear regression of the
myocardial velocities along the M-mode
line across the myocardial wall. If
velocities are linearly distributed
through the wall, this is equal to the
difference in endocardial and epicardial
velocities divided by the instantaneous
wall thickness (W):
As the apex is stationary, while the base moves toward the apex in systole, away from the apex in diastole, the ventricle has to show differential motion, between zero at the apex and maximum at the base. | As motion decreases from apex to base, velocities has to as well. This is seen very well in this plot of pwTissue Doppler recordings showing decreasing velocities toward apex. Thus, there is a velocity gradient from apex to base |
The strain rate can be described by the instantaneous velocity gradient, in this case between two material points, but divided by the instantaneous distance between them. In this description, it is the relation to the instantaneous length, that is the clue to the Eulerian reference. | Strain rate is calculated as the velocity gradient between two spatial points. As there is deformation, new material points will move into the two spatial points at each point in time. Thus, the strain that results from integrating the velocity gradient, is the Eulerian strain. |
Strain rate and strain can be visually
assessed by the offset between the
curves, when the velocity curves are
obtained from points with a known (and
equal) distance.
Velocity
measures with some
amount of noise. |
Unsmoothed
strain rate curves
from the same loop.
The increase in noise
is evident. |
Velocity and strain rate imaging of the same (normal) left ventricle. The colour sector can bee seen to be equal to the B-mode sector.Velocity is red in systole when all parts of the heart muscle moves toward the probe (apex) and blue in diastole. The changes are too quick to observe entirely, to make full use of the information the image has to be stopped and scrolled. | Curved anatomical M-mode (CAMM). A line is drawn from apex to base, and velocity data over time are sampled along the line and displayed in colour along a straight line. The numbers on the curve and the M-mode are included for reference and corresponds to the numbers on the B-mode image. This example shows the septum from the apex to base along one axis, and one heart cycle along the other, in a two - dimensional space - time plot. S: systole, E: early relaxation, A: atrail contraction. |
Colour
M-mode (CAMM) of tissue
velocities in fundamental
(above) and harmonic (below)
imaging. Slight aliasing can
be seen in native imaging in the e' wave at the base. In harmonic imaging, there is aliasing both in the S' wave, and the e' wave (double). |
Colour
tissue Doppler curved M-mode
in harmonic imaging,
velocity plot (above),
strain rate (below). As can
be seen there is heavy
aliasing in the velocity plot, but no aliasing in strain rate imaging. |
Shadowy reverberations covering the anterior wall in this 2-chamber image. It is differentiated from drop out, as we can se a "fog" of structures covering the anterior wall. The structures are stationary. This is not distinct reverberations shadows, but incoherent clutter. | Recordings from the basal anterior ring of the subject to the left. The true signal is clearly visible as a normal curve, and can be seen separately from the clutter band, which is the horizontal spectral band along the baseline. The colour Doppler recording is superposed and aligned with both vertical and horizontal scale. The colour Doppler, using the autocorrelation algorithm, results in mean velocities that incorporate both signal and clutter, giving a severe underestimation of velocities. |
Basal
spectral tissue Doppler curve
in the anterior wall. Peak
systolic velocity ca 8.5 cm/s. |
Midwall
spectral tissue Doppler curve
in the anterior wall. Peak
systolic velocity ca 6.5 cm/s. |
Apcal
spectral tissue Doppler curve
in the anterior wall. Peak
systolic velocity ca 5 cm/s. |
IColour
M-mode from the image shown
above. The curved M-mode shows
a fairly homogenous and normal
signal in the inferior wall
(top), but more or less random
noise in the anterior wall
(bottom), where the noise is
seen as vertical stripes of
alternating colours. |
Velocity
curves from the anterior wall,
showing noise, and not much
more, but at low level (within
± 0.3 cm/s). |
The principle of the effect of clutter. V-plot with clutter showing how the mean velocities are reduced, compared to the mormal expected values (red line). But in additions the variation of the velocity estimates from pixel to pixel is much higher, resulting in increased noise, but with reduced mean values. | Combined
pulsed Doppler (yellow bands)
and colour Doppler green
Aligned horizontally and
vertically. The noise level
can be seen to be b´very low,
compared to the peak
velocities shown in the pulsed
Doppler recording. The clutter
is the horizontal band around
the baseline, and the width of
the spectrum in this case is
the noise. |
Image from another subject in the study shown above (240). In this subjech there is some clutter from reverberations, as seen by the band in systole close to the zero line. In this case the peak velocity by autocorrelation is lower than the modal velocity of the main spectral band, which still was the one closest to the RF M-mode reference. (Figure courtesy of Svein Arne Aase) | Clutter filtering may reduce the problem, as seen here. There is a band of clutter close to zero velocities, but as seen here, the spectral modality makes it very easy to separate the true and clutter velocities. However, the clutter affects the autocorrelation velocity (red line), giving lower velocities, but with clutter filter this effect is removed (red line) , and the peak value is substantially higher. Image modified from (243). |
Ultra high frame rate tissue Doppler is done by combining more principles:
By this method, using two broad, unfocussed
(planar) beams, each covering one
wall, as well as 16 MLA and sparse
interleaved B-mode imaging, it has
proved possible to increase the TDI
frame rate substatially (172,
268).
it has been possible to increase frame
rate to 1200 FPS in 2D imaging.
Few
beams give high frame rate. Image
courtesy of Svein Arne Aase,
modified from (172).
Already this has shown new information
about both the pre
ejection and post
ejection dynamics.
With this method, it is possible to
acquire IQ (RF) data with FR > 1000.
This makes it possible to process
restrospective tissue Doppler from the
whole field (i.e. that covered by the
two transmit beams), simultaneously from
one heart cycle. as in colour Doppler.
The V-plot, derived from
autocorrelation, is vulnerable to clutter
and drop outs, as decribed in the pitfalls
section. this means that the
V-plots are difficult to see differences
between artefacts and pathology.
However, with Ultra high frame rate
tissue Doppler (UFR-TDI), it is possible
to process restrospective tissue Doppler
from the whole field simultaneously,
which means that not only will tissue
Doppler be available that are relatively
unaffected by clutter noise, but
also that spectral V-plots can be
processed, which shares the reduced
vulnerability to clutter as shown in the
pitfalls
section.
Tissue Doppler is still limited to one
velocity direction only. This means that
the term "3-dimensional" refers to a
three dimensional distribution
of tissue velocities only, not
velocity vectors in a three
dimensional coordinate system. However,
data from the whole ventricle can be put
together on a surface model of the left
ventricle.
3D tissue
Doppler is basically a grid of
numerical values on a
ventrricular surface. |
Triplane
tissue Doppler, showing three
standard planes, with the
assumption that the angle
between them is 60°, the rest
of the data between the planes
are than interpolated. This
gives a circumferential
resolution of 60°. |
As tissue data as about acquiring (and
displaying, f.i. by colour) numerical
data, the method do not have the same
limitation as 3D B-mode. One method is
to combine information from three
standard planes, and then interpolating
the data between the planes by for
instance spline. The method has been
explained elsewhere.
This has been done both by combining
sequentially acquired standard planes.
It could also be done as a simultaneous
triplane acquisition, but at the cost of
a substantially reduced frame rate.
Thus, freehand scanning has been
preferred.
This version of three dimensional
tissue Doppler may be used for display,
but also for area measurement, as the
data are distributed over a
representation of the (approximate) real
ventricular area.
With the Ultra
high frame rate tissue Doppler
method, it is also possible to acquire
three dimensional tissue Doppler in real
time.
Using a 3D matrix probe, sending an
array of 3x3 broad unfocussed or planar
beams, and using a 4x4 matrix of receive
beams for each transmit beam, giving a
16 (or 4x4) MLA, we have been able to
achieve a volume rate of about 500 VPS (280),
i.e. Ultra high frame rate 3D tissue
Doppler.
Principle
of beam
formation, showing a matrix of
3x3 wide transmit beams
(brown circles) and for
each beam an array of 4x4
receive beams, i.e. a 16 MLA.
(After 280). |
Distribution
of the transmit beams in
relation to a cross section of
the ventricle, endocardial
and epicardial surfaces marked
with black lines and arrows.
the energu distribution of the
beams is shown by the colour
hue. The
transverse plane shown to the
right is marked by the thick
line. (After 280).
|
Distribution of the transmit beams in an apical plane, the level of the cross section to the left is marked by the thick line. As evident from the illustration, the transmit beams do not cover the whole sector, but will cover most of the walls. (After 280). |
By this method, t is possible to
achieve a high circumferential
resolution through the MLA technique at
the same time as a high temporal
resolution. The result can be displaued
as a 3D figure, as with reconstructed
3D, and both curved M-modes and
tivelocity curves can be extracted from
this matrix:
3D surface with tissue velocity display. The ring represents a line for extraction of the curved M-mode shown top, left. (After 280). | Data display from the 3D velocity figure to the right. Top: curved M-mode, showing the time variation of apically directed velocities in a ring around the mid ventricle. Bottom, velocity curves from the basolateral part, red from UFR 3D TVI, extracted from the 3D data to the left, blue velocity from the same point in the same subject, but acquired bt conventional colur TDI (i.e. a different heartbeat). |