| Material |
Velocity
( m/s) |
| Air |
330 |
| Water |
1497 |
| Metal |
3000 - 6000 |
| Fat |
1440 |
| Blood |
1570 |
| Soft tissue |
1540 |
| Basically, a reflected ultrasound pulse is
a waveform. However, storing the full waveform, called RF data, is
demanding in terms of storage, as each point on the curve would have to
be represented in some way or other. However, if the full RF data are
stored, the amplitude and frequency data could both be calculated in
post processing. |
The pulse has a certain amplitude. Just
storing the amplitude is much les demanding (corresponding more or less
to one number per pulse). This is the only data that are used in grey
scale imaging, where the amplitude is displayed as brightness of the
point correspåonding to the scatterer as in B-mode
and M-mode. |
However, the reflected ultrasound pulse has
a frequency (or a spectrum of frequencies), and this can be represented
as a numerical value per image pixel as well, as described in Doppler imaging. Still, the amount of data is far
less than the RF data. |
![]() |
![]() |
||
| Effect of size and direction of the reflecting surface. The two images on the left shows a perfect reflecting surface. Most of the energy (but not all, as the wavefront is not flat), will reflect back to the transducer resulting in a high amplitude echo, when the surface is perpendicular to the ultrasound beam. On the other hand, if this surface is tilted 45º, almost all energy will be reflected away from the surface, resulting in a very low amplitude echo. The two images to the right shows a scatterer with a more curved surface, resulting in more energy being spread out in different directions, this will give a lower amplitude signal when oriented toward the beam, but reflects more energy back towards the probe if it is tilted. | The effect of the direction of the
reflecting surface in a long axix image of the left
ventricle. The echo resulting from the septum-blood
interface (arrows) is far stronger in the regions where the
surfaces are perpendicular to the ultrasound beamns (blue arrows),
compared to the region between where the surface is slanted compared to
the ultrasound beams. |
Cyclic variations in the amplitude in
reflected ultrasound (integrated backscatter) with heart cycle. This
reflect the variations in reflexivity, but not myocardial density, as
the myocardium is incompressible. Thus, most of the amplitude
variations must be due to changes in fibre directions. |
![]() |
![]() |
| Fig. 3a. The ultrasound image is built up as a line of echoes based on the time lag and amplitude of the reflected signals. |
3b. The reflected signals can be displayed in three
different modes. A-mode (Amplitude) shows the depth and the reflected
energy from each scatterer. B-mode (Brightness) shows the energy
or signal amplitude as the brightness (in this case the higher energy
is shown darker,
against a light background) of the point. The bottom scatterer is
moving. If the depth is shown in a time plot, the motion is seen as a
curve, (and horizontal lines for the non moving scatterers) in a M-mode
plot (Motion).
|
| Attenuation. Imaging of a
homogeneous tissue, f.i. liver will change the apparent density behind
structures with different attenuation. Behind a structure with high reflexivity (e.g. a calcification), there will be high attenuation, (white; left). Hence, the sector behind receives less energy, and appears less dense (darker), the area behind may even be a full shadow. Behind a strcture with low reflexivity (e.g. a fluid) there is little attenuation (black; right), the tissue receives more energy and appears denser (brighter - "colouring") than the surrounding tissue. |
Liver with a gallbladder in front,
containing gallstones. The gallstones are dense, with a shadow behind.
The
rest of the gallbladdeer is fluid filled, thus the sector behind the
fluid appears denser than the
neighbouring tissue due to "colouring". |
| |
![]() |
![]() |
| Principle of gain, reject and compress. All curves display brightness of the display in relation to the amplitude of the rejected signal. An ordinary gain curve is shown in black, using a linear brightness scale, displays the full range of amplitudes. Increasing gain (red curve), will increase all signals, including the weakest, as in the noise. The disadvantage, in addition to increasing noise, is that the strongest signals will be saturated, so details may disappear. Compress is shown as the blue curve. This results in a steeper brightness curve, resulting in less brightness of the weakest echoes, and more brightness of the strongest. Thus, weak echoes may disappear together with background noise, while strong echoes will be saturated, resulting in loss of detail. Finally reject is shown by the light grey zone, siply displaying all signals below a certain amplitude as black. (The black brightnes curve drops abruptly to zero at the reject limit (dark grey line). A combination of high gain and reject will give an effect fairly similar to the compress function. | ||
![]() |
![]() |
![]() |
| Image with default gain, reject and
compress settings |
Same image with high gain (top) showing
increased denity of the endocardium, but loss of detail due to
birghtness saturation and a corresponding increase in cavity noise and
low gain (bottom), showing reduction in cavity noise, but loss of
detail (see endocardium in lateral wall). |
Same image with increased reject (top)
showing reduction in cavity noise, but also with slight loss of detail
(endocardium in lateral wall) and compress function with less detail in
the myocardium due to increased brightness. |
| a |
b |
c |
| Halving the frequency results in half the
number of oscillations per time unit, or longer time (= pulse length)
for the same number of oscillations. Thus halving the frequency, as in second harmonic analysis, will result in longer
pulse length. However, the bandwidth is far
less affected. |
Second harmonic (1.7/3.5 MHz) left and fundmental (3.5 MHz) right images of LV
septum, showing how the echo from the blood/septum interface (arrows)
is thicker in harmonic imaging, due to the reduction in frequency.
Observe, however, how cavity noise is much reduced in harmonic imaging,
resulting in a far more favorable signal-to-noise ratio. |
The thickness of the surtface echoes is
dependent n the pulse length, and thus also on the frequency.
This picture of the septum illustrates how the leading-to-leading ASE
convention shown in red, eliminates the pulse length in measurement (as
the echo blooms in both directions), while the Penn convention will
result in increasing overestimation of the thickness by increasing
pulse length as it incorporates the interface on both sides. |
![]() |
![]() |
![]() |
![]() |
| A pulse is sent out, ultrasound is reflected, and the B-mode line is built up from the reflected signals. | Linear array. |
Curvilinear array |
![]() |
![]() |
| By
making the ultrasound beam sweep over a sector, the image can be made
to build up an image, consisting of multiple B-mode lines. |
c. In principle, the image is built up line by line, by emitting the pulse, waiting for the reflected echoes before tilting the beam and emitting the next pulse. Resulting in an image being built up with a whole frame taking the time for emitting the total number of pulses corresponding to the total number of lines in the image. |
![]() |
![]() |
![]() |
|||
| Fig. 7A. Mechanical transducer. The
sector is formed by rotating a single transducer or array of
transducers mechanically, firing one pulse in each direction and
then waiting for the return pulse before rotating the transducer
one step. In this beam there is electronic focusing as well, by an
annular array. |
B. Electronic transducer in a phased
array. By stimulating the transducers in a rapid sequence , the
ultrasound will be
sent out in an interference pattern. According to Huygens principle,
the wavefront will behave as a single beam, thus the beam is formed by
all transducers in the array, and the direction is determined by the
time sequence of the pulses sent to the array. Thus, the beam can be
electronically steeredand will then sweep stepwise over the sector in
the
same way as the mechanical transducer in A, sending a beam in one
direction at a time. |
|
| Dynamic focusing. The
same
principle of
phase steering can be applied to make a concave wavefront, resulting in
focusing of the beam with its narrowest part a distance
from the probe. Combining the steering in B and C will result in
a focussed beam that sweeps across the sector, as in the moving image
above. |
Resulting Ultrasound beam as shown by a computer simulation, focusing due to the concave wavefront created by the dynamic focusing. The wavelength is exaggerated for illustration purposes. Image Courtesy of Hans Torp. |
![]() Annular focusing in all
directions both in plane and transverse to the plane.
|
![]() Linear focusing in the imaging
plane only.
|
| The
lateral resolution of
a beam is dependent on the focal depth, the wavelength and probe
diameter (aperture) of the ultrasound probe. (Reproduced from Hans Torp by permission) |
Thus a small scatterer will appear to be
"smeared
out", and the apparent size in the image is determined by the beam
width and pulse length.
As the pulse length is less than the beam width, the object will be "smeared out" most in the lateral direction. |
Two scatterers at the same depth, separated
laterally by less than the beam width, will appear as one. |
Two
scatterers at different depths will appear separate if separated
by more than the pulse length. |
But, if separated both laterally and in
depth, they will appear as being in the same line, if lateral
separation is within the beam. |
| Two points in a sector
that is to be scanned. |
The ultrasound scan will
smear the points out according to the lateral resolution in each beam. |
Two
different lateral resolutions, the speckles can be seen to be
"smeared". In this case the loss of resolution in the right image is
due to lower line density . By rights the image should appear as split
in different lines as indicated in the middle, as each beam is
separated, line density being less than optimal relative to the beam
width. Instead the image is interpolated beween lines. This reduction
in line density is done to achieve a higher frame rate, as illustrated below. |
![]() |
![]() |
| As the depth of
the sector determines the time before next pulse can be sent out,
higher depth results in longer time for building each line, and thus
longer time for building the sector from a given number of lines, i.e.
lower frame rate. |
Thus reducing the desired depth
of the sector results in shorter time between pulses, and thus shorter
time for building each line, shorter time for building the same number
of lines, i.e. higher frame rate. In this case, the depth has been
halved, and the time for building a line is also halved. |
![]() |
![]() |
| In
this case, in the image to the left, the depth has been halved,
reducing the time for building each line to half, thus also halving the
time for building the full sector, doubling the frame rate. |
|
![]() |
![]() |
![]() |
| Fig.7a. A sector with a given depth,
sector width and line density determines the frame rate. |
b. Reducing sector
width, but maintaining the line density, gives
unchanged lateral resolution but higher frame rate,
at the cost of field of view. |
c. Reducing the line density instead and maintaining sector width, results in lower number of lines, i.e. lateral resolution, and gives the same increase in frame rate. |
![]() |
![]() |
| Illustration
of effects of shadows on an ultrasound beam. Left: no shadow. Middle, a
shadow distant from the beam (e.g. a calcification or the lung seen at
a distance), resulting in a shadow with no image below it. Left a
shadow close to the transducer surface (e.g. lung edge or rib) will
result in a narrow beam (reduced apparent aperture) which will not be
seen as a shadow in the picture, but rather a reduced lateral
resolution. (Simulation image courtesy of Hans Torp.) The effect
of the depth of the origin of the shadows in the images is shown below,
indicated by the green arrows. |
If the near shadow is in the centre of the
probe, the result may be that the beam is split in two, resulting in
two apparent apertures. The effect on the image is shown below. |
|
![]() |
![]() |
![]() |
| Left, shadow originating at a depth of ca 3 cm, as can seen by the visible structures of the chest wall closer to the probe. The shadow is probably due to the edge of the lung. Right; a small repositioning of the probe solves the problem. | Left shadow originating close to the chest wall (< 1 cm), probably the edge of a costa. It can be seen as a shadow, but the main effect is loss of lateral resolution in the shadow, and again a small repositioning of the probe solves the problem as seen to the right. | Split image due to two virtual appertures,
caused by a near shadow in the middle of the probe footprint. |
![]() |
![]() |
|
| Drop out of the anterior wall in this
2-chamber view due to a lung shadow distant from the probe. However,
the lateral resolution may be seen to be reduced at the basal part of
the border
between the picture and the shadow. |
Reduced lateral resolution due to costal
shadow. The effects of both costae and shadows will vary, according to
the distance from the probe. In this case the patient was extremely
thin, thus there was virtually no distance between the probe and the
costa. In this case, no localised shadow can be seen, the costa was to
the left in the image, where resolution is poorest. |
![]() |
| Simulated beam with focusing, showing
interference pattern dispersing some of the beam to the sides. (image
courtesy of Hans Torp). |
Side lobes from a single focussed ultrasound beam. These side lobes will also generate echoes from a scatterer hit by the ultrasound energy in the side lobes, i.e. outside the main beam. |
| This
results in a false echo in the main beam |
AS the beam with side lobes sweeps across
the sector, each echo from the scatterer in both the main beam and the
side lobes will generate the false echo in the position of the main
beam |
Resulting in a smeared out echo across a
large part of the sector. |
![]() |
![]() |
![]() |
| Side lobes originating from the fusion line
of the aortic cusps, seen to extend into both the LV cavity and the
aortic root cavity (arrows). |
As opposed to reverberations,
the side lobes
moves with the structure, and may change with time (in this case the
echo intensity of the fusion line decreases as the valve opens, and
thus the intensity of the side lobes too) . |
An image riddled with side lobes. Again
they can be seen to move in the axial direction with the structures
they are originating from. (The echo in the middle of the right
ventricle cavity is an ICD lead, and not an artifact.) |
| Apical position of the probe. The two
orthogonal planes can be seen to bisect the apex. |
Foreshortening by erroneous placement of
the probe. The two orthogonal planes can be seen to bisect the wall,
not the apex. In one plane this will not be evident, as the intersecting wall still shows an ellipsoid shape. |
![]() |
![]() |
![]() |
| A: LV in correct 4-chamber plane. Probe in
the 5th intercostal space. Image is less than optimal in quality. |
B: Foreshortened image in 4 chamber plane,
by placing the probe in the 4th intercostal space. Wall definition as
well as detail become clearer. |
C: Rotating the probe in position B to the
two chamber plane, shows the apex to be out of sector, thus the four
chamber image is foreshortened. |
| Non-linear propagation. The upper panel shows the waveform of a pulse as originally transmitted, and after 6 mm transmission through tissue. The lower panels shows how the energy distribution is shifted to a more evenly distribution between more frequencies. (image courtesy of Hans Torp). | Fourier analysis of the resulting signal in native frequency (left) and second harmonic mode (left) shows that the native signal contains much more energy at all depth, while the harmonic signal contains most of the energy at a certain depth, in this case at the level of the septum, showing a much better signal-to-noise ratio.(image courtesy of Hans Torp). | Energy distribution of the signal from cavity (lower curve) and septum (upper curve), showing the same phenomenon as the middle picture. The difference between cavity signal (being mostly clutter) and tissue is small in the native frequency domain (1.7 MHz), but there is little clutter at the harmonic frequency (3.4 MHz). Thus, filtering the native signal will reduce clutter, as shown below. (image courtesy of Hans Torp). |
![]() |
![]() |
|
| The same image in harmonic (left) and fundamental (right) mode, showing the improved
signal-to-noise ratio in
harmonic imaging, especially in rducing noise from the cavity.
(Thanks to Eirik Nestaas for correcting my left-right confusion in this
iamge text) |
Stationary
reverberation in harmonic (left) and fundamental (right) imaging,
showing the effect of harmonic imaging on clutter.
|
Fundamental (left) and harmonic (right)
images of the left ventricle at the level of the chorda tendineae. The
echo of the chorda (blue arrow) can be seen to be thicker in harmonic
imaging, due to the longer pulse
length. The echo generates a side lobe that can be seen to the
right of the chorda. The side lobe is more prominent in fundamental
than harmonic imaging. Note also the reduction in cavity noise from the
right and left ventricle in the harmonic image. |
![]() |
![]() |
| Interference pattern.
Here is simulated two wave sources or scatterers at the far field
(white points). The emitted or reflected waves are seen to generate a
speckle pattern (oval dots) as the amplitude is increased where wave
crests cross each other, while the waves are neutralised where a wave
crest crosses a though. This can be seen by throwing two stones
simultaneously in still water . The speckle pattern can be seen in
front of the scatterers, towards the probe. |
Irregular interference
pattern. This is generated by more scatterers somewhat randomly
distributed. The speckle pattern is thus random too. Again there
may be a considerable distance between the speckles and the scatterers
generating the pattern. |
| Humpback whale diving. Each humpback has an unique (speckle) pattern on the underside of the tail (and flukes). Thus each individual can be identified by its speckle pattern. Photographs at different times and places can thus track the wandering of each individual all over the area it wanders, without recourse to anything else than the pattern. - Speckle tracking! This is thus a method with low frame rate, giving mainly the extent of wandering over a long time period (the sampling intwerval). To measure swimming velocity, a Doppler sonar would have been useful. |

![]() |
![]() |
| An M-mode along the septum demonstrates how the speckles is shown as motion curves. It is evident that many speckles are only visible during part of the heart cycle, but if the speckle pattern is compared from frame to frame, the changes will be small. The grainy texture of the lines is due to the limited frame rate as the M-mode on the right is reconstructed from the 2D image at the left. When the speckle pattern is followed by an M-mode in the wall, the alternating bright and dark points are seen as alternating bright and dark lines. The lines remaining to a large degree unbroken, shows the pattern to be relatively stable, the speckles moving along with the true myocardial motion, and thus myocardial motion can be tracked by the speckles. | |
![]() |
![]() |
|
| Speckle tracking. Real
time M-mode
demonstrates how the speckle pattern follows the myocardial motion. |
Defining a kernel in the
myocardium will
define a speckle pattern within (red). Within a defined search area (
blue), the new
position of
the kernel in the next frame (green) can be recognised by finding the
same speckle pattern
in a new position. The movement of
the kernel (thick blue arrow) can then be
measured. |
Speckle tracking search
algorithm. The
kernel is defined in the original frame at t=0 (red square). In the
next frame, at t=t, the algorithm defines a search area (white square),
and the search is conducted in all directions for the matching kernel. |
![]() |
![]() |
| LOngitudinal speckle tracking in apical 4 chamber view. The resulting tracking of the kernels shown in motion. As can be seen, with a drop out apicolateral, this ROI tracks less than perfect, giving too low strain both in LA and MA segments. | Speckle tracking can be applied crosswise. In this parasternal long axis view, the myocardial motion is tracked both in axial and transverse (longitudinal) direction. It is evident that the tracking is far poorer in the inferior wall, due to the poor lateral resolution at greater depth. |
![]() |
The Doppler effect |
![]() |
| Christian Andreas Doppler (1803 - 1853) |
My cat Doppler (2004 - |
The Doppler effect was discovered by Christian Andreas Doppler (1803
- 1853), and shows how the frequency of an emitted wave changes
with the velocity of the emitter or observer. The theory was presented
in the royal Bohemian society of Science in 25th of May1842 (5
listeners at the occasion!), and published in 1843
(119). The premises for his
theoretical work was faulty, as he built his theory on the work of
James Bradley who erroneously attributed the apparent motion of stars
against the background (parallax effect) to the velocity of the earth
in its orbit (instead of the effect of Earth's position in orbit on the
angle of observation). Further, Doppler attributed the differences in
colour
of different stars to be due to the Doppler effect, assuming all stars
to be white. Finally, he theoretised over the effect of the
motion of double stars that rotate around each other, assuming
a Doppler effect from the motion. The changes in
wavelength from the Doppler
effect, however, is too small to be observed.
However, Doppler did a theoretical derivation of the effect of the motion of the source or observer on the perceived wavelength from the premises of a constant propagation velocity of the waves in the medium, and this is entirely correct, valid both for sound waves and electromagnetic radiation of all kinds. The basis for the Doppler effect is that the propagation velocity of the waves in a medium is constant, so the waves propagates with the same velocity in all directions, and thus there is no addition of the velocity of the waves and the velocity of the source. Thus, as the source moves in the direction of the propagation of the waves, this does not increase the propagation velocity of the waves, but instead increases the frequency.The original derivation of the Doppler principle as well as the extension to reflected waves is explained in more detail here. As a work of theoretical physics, it is thus extremely important. In addition, it has become of practical importance, as the basis for the astronomical measurement of the velocity of galaxies by the red shift of the spectral lines, in Doppler radar, Doppler laser and Doppler ultrasound.
The theory was experimentally validated by the Dutchman Christoph
Hendrik Diderik Buys Ballot (120),
with the Doppler effect on sound waves, who placed musicians along a
railway line and on a flatbed truck, all blowing the same note, and
observed by subjects with absolute pitch, who observed the tones being
a half note higher when the train was approaching as compared to the
stationary musicians, and a half note lower as the train receded.
(This can be observed in everyday phenomena such as the sound of f.i. an ambulance siren, the pitch (frequency) is higher when the ambulance is coming towards the observer, hanging as it passes, and lower as it goes away.
This is illustrated below:
The Doppler
effect. As the velocity of sound in air (or any other medium ) is
constant, the
sound wave will propagate outwards in all directions with the same
velocity, with
the center at the point where it was emitted. As the engine moves, the
next
sound wave is emitted from a point further forward, i.e. with the
center a
little further forward. Thus the distance between the wave crests is
decreased
in the direction of the motion, and increased in the opposite
direction. As the distance between the wave crests is equal to the
wavelength, wavelength decreases (i.e. sound frequency increases) in
front
of the
engine, and increases (sound frequency decreases) behind it. This
effect can be
heard, as the pitch of the train whistle is higher coming
towards a listener than moving away, changing as it passes. The effect
on the pitch of the train whistle was published directly, but later
than
Doppler and Buys Ballot.

The Nykvist phenomenon (121) is an effect of the
relation between the sampling frequency and the observed velocity. If
you sample at a certain frequency, the direction of the motion becomes
ambiguous, more frequent sampling will give the correct direction, less
frequent sampling results in an apparent motion in the opposite
direction. This can be observed with a stroboscopic light, for instance
illuminating the flow of water, or with old fashioned wagon wheels in
old moves which often seem to revolve slowly backwards when the wagon
moves forwards.
This is illustrated below.
Constant rotation velocity, decreasing sampling frequency:The easiest is to show how
reducing the sampling frequency affects the apparent motion. All
circles rotate with the same rotation velocity clockwise. The sampling
frequency is reduced from left to right. It can be seen that the red
dots is at the same positions when they are seen to move.
|
|||
![]() |
![]() |
![]() |
![]() |
| a: 8:1
8 samples per rotation, the red point is seen in eight positions during the rotation. |
b: 4:1 4 samples per rotation, the red point is seen to rotate just as fast, but is only seen in four positions |
c:
2:1 2 samples per rotation, i.e. the sampling frequency is exactly half the rotation frequency. Here, the red dot is only seen in two positions, (but it is evident that it is in the same positions at the same time as in a and b). However, it is impossible to decide which way it is rotating. This is the Nykvist limit; sampling rate = 1/2 rotation rate. |
d: 1.5:1 1.5 samples per rotation,or one sample per three quarter rotation, making it seem that the red dot is rotating counter clockwise. Again, the dot is in the same position at the same time as in a and b. |
|
|
|||
![]() |
![]() |
![]() |
![]() |
| a: 1:8 One rotation per 8 samples. The sampling catches the red dot in 8 positions during one rotation. |
b: 1:4 Rotation velocity twice that i a; one rotation per four samples, the sampling catches the red dot only in four positions during one rotation. |
c: 1:2 Rotation velocity four times a; one rotation per two samples, this catches the red dot in only two positions, giving directional ambiguity as above. |
d: 1:1,5 Rotation velocity six times a; one rotation per 1,5 samples, or 3/4 rotation per sample, giving an apparent counter clockwise rotation. |
Sampling from increasing depth will increase
the time for the pulse returning, thus increasing the sampling interval
and decrease the sampling frequency. The Nykvist limit thus
decreases with depth. This means that pulsed Doppler has depth
resolution, but this leads to a limit to the velocities that can be
measured.
The Nyquist limit is dependent on the depth (D) of the sampling volume. The larger distance from the probe, the longer time for the pulse to pass to the desired depth and back, and the lower the Nyquist limit. The time for the return of the pulse is:
T = 2D/c and the
maximum PRF as PRF = c/2D
Aliasing is no
problem in pulsed tissue Doppler, as
tissue velocities are far below the
Nykvist limit. However, in colour tissue Doppler, harmonic imaging is
halving the effective frequency, and leads to aliasing as shown below.
on the other hand, will measure
all velocities along the ultrasound beam: The beam is transmitted
continuously, and the received echoes are sampled continuously with no
range gating. Thus,
there is no information about the time interval from the signal to the
reflection, and, hence, no information about the depth of the received
signal; the signal may come from any depth. The continuous Doppler has
no Nykvist limit, and can measure maximal velocities. It is used for
measuring high velocities.
This means that both methods has limitations:
pulsed Doppler has velocity ambiguity at high velocities, and
continuous wave Doppler has depth or range ambiguity. Thus, for
continuous Doppler the pulse length can be long, as there is no depth
resolution, while in pulsed Doppler it has to be shorter in order to
achieve a sufficient depth resolution.
Examples
of continuous wave versus pulsed wave, and the Nykvist effect. Left:
Aortic insufficiency shown by cw Doppler. It van be seen that there are
a fair distribution of velocities in the whole spectrum. However, There
are far more velocities blow 2 m/s. In this case, the low pass filter
is only set to suppress tissue velocities. If the point is to get a
clear
visualisation of the maximal velocities in the jet, at 4 - 6 m/s, the
filter should be set higher. Left, the same patient by pulsed Doppler
of the LVOT. The outflow can be seen as a narrow band, within the
velocity range, while the regurgitant jet has velocities outside the
Nykvist range, and there is total velocity ambiguity.
| The principle of HPRF. Pulses are transmitted with three times the frequency that is necessary to allow the echo from the furthest depth to return. Thus, the echo of pulse 1 will return from level 3 at the same time as the echo of pulse 2 from level 2 and and of pulse 3 from level 1, and there is no way to determine whether a signal is from level 1, 2 or 3. | HPRF pulsed Doppler recording (right). with
one sample volume in mid ventricle and one in the mitral ostium. The
recording shows a systolic dynamic gradient (due to inotropic
stimulation with dobutamine), as well as an
ordinary mitral inflow curve. There is no way in the pulsed
recording to determine which velocities that originate from which
sample
volume (except from á priori knowledge, of course). |
| Two pulses sent toward a scatterer with a time delay t2 - t1 = 1/PRF. Given that the scatterer has a velocity, it will have moved a distance, d, that is a function of the velocity and the time (d = v x t). Thus, pulse 2 travels a longer (or shorter) distance equal to d with the speed of sound, c, before it is reflected. | During the time pulse2 has travelled the distance d to the new position of the scatterer and back to the point of the reflection of pulse 1, i.e. a distance 2d pulse 1 has travelled the same distance away from the reflection point. (The scatterer will have travelled further, but this is not relevant). Thus the diasplacement of the waveform of pulse 2 relative to pulse 1, is 2d. | By sampling the two pulses simultaneously at two timepoints, as shown in the previous illustration, the phase of each pulse can be determined. The phase analysis of the relative positions of all four points is done by autocorrelation, a quick (and dirty?) method that allows online computation. |
| Principle of package acquisition. Time
depth diagram of
the position of a moving scatterer. Each dot represent one pulse.
Packages of two pulses are sent to
the scatterer with intervals. The time between the pulses in one
package is given by 1/PRF, and decides the Nykvist limit. The time
between packages is the time it takes to build a full sector of lines
in colour flow mode (CFM) and is given by the frame rate (FR), the time
interval being 1/FR. This decides the temporal resolution of the CFM. |
CFM
sector superposed on a B-mode sector. By reducing sector size, line
density and sampling frequency, the CFM image can achieve an acceptable
frame rate. This is feasible because the region of interest for
the flow is usually only a part of the ROI for The B-mode, ,flow being
intracavitary as shown below. |
![]() |
![]() |
| Power Doppler image of the renal
circulation. The amplitude is a function of the number of scatterers,
i.e. the number of blood cells with a Doppler shift. This is shown as
the brightness (hue) of the signal. In addition, direction of flow can be
imaged by different colours (red - positive flow - towards probe, blue
- negative colours - away from probe), and still the brightness may
show the amplitude. |
Colour flow showing a large mitral regurgitation. Velocities away from the probe is shown in blue (converting to red where there is aliasing), towards the probe is red. In this image, the green colour is used to show the spread (variance) of velocities. This will also reflect areas of high velocities (high variance due to aliasing). The sector with colour flow is seen to be far smaller than the B-mode sector. The image displays the direction, extent and timing of the jet. |
![]() |
|
| Recording from a patient with apical hypertrophic cardiomyopathy. Ejection can be seen in blue, and there is a delayed, separate ejection from the apex due to delayed relaxation. There is an ordinary mitral inflow (red), but no filling of the apex in the early phase (E-wave), while the late phase (A-wave) can be seen to fill the apex. Left, a combined image in HPRF and colour M-mode. The PRF is adjusted to place two samples at thr mitral annulus and in the mid ventricle just at the outlet of the apex. The mitral filling is shown by the green arrows, and the late filling of the apex is marked by the blue arrow. In addition, theere is a dynamic mid ventricular gradient shown by the red arrow, with aliasing in the ejection signal in colur Doppler. The delayed ejection from the apex is marked by the yellow arrow (the case is described in (87). The utility of the different methods is evident: HPRF (or cw Doppler) for timing and velocity measurement, but with depth ambiguity, colour M-mode for timing and location of the different jets, direction being displayed by the colour. | |
The phase analysis is often done by the process known as autocorrelation.
This will result in a values that does not reflect the spectrum, but
only mean values in the spectrum. But if there is clutter in the region
(stationary echoes), this will be incorporated in the mean, resulting
ion lower values. In Doppler flow, this can be filered by the high pass
filter, and thus will represent a small problem. In tissue
Doppler, this may be a more
significant problem, as the velocities are only about 1/10 of the flow
values, and thus clutter may be more difficult to separate from true
velocities. Thus, a substantial amunt of clutter may reduce
autocorrelation values for tissue Doppler more than pulsed Doppler as
discussed below.
In addition, it is
customary to analyse the tissue Doppler values in native, rather than
harmonic imaging, due to the Nykvist
limitation.
Thus, there is a greater amount of clutter than if harmonic imaging had
been used, as
shown in B-mode images.
For optimal colour flow,
it is important to realise that there may, in some scanners, be an
inverse relation between the gain of colour Doppler and B-mode. (In
some scanners it is possible to adjust the priority, or to adjust the
gain settings separately). This, however, is an acquisition finction,
and not image adjustment, and thus cannot be compensated afterwards.
This is illustrated below:
![]() |
![]() |
| The
diagram to the left shows the placement of flow and tissue signals on
this intensity (amplitude) / velocity diagram. Velocity given as the
height ogf the bars, intensdity showb both by the placement on the x
axis, as well as the darkness of the bars, black being the highest
intensity. The flow signals are low intensity
but mostly high velocity, while the tissue is exclusively low
velocity, high intensity. The heart valves, however, are solid
structures which moves with the velocity of the passing blood,
resulting in high intensity signals giving a saturation of the Doppler
spectrum. A typical flow curve from the right ventricular outflow tract
is shown to the left, with the valve click. |
|
![]() |
![]() |
| Application
of a high pass filter (low velocity reject) shown schematically to the
left and in practice applied to a mitral flow curve to the right.
Velocities lower than the limits of the green bar (showing the range of
the filter) are removed seen in the dark zone in the middle of the
spectrum. The
setting rejects velocities at blood intensities below 15 - 20 cm/s,
which is too high for normal flow velocities as in this instance,
although may often be useful in continuous wave Doppler
recordings
of
high velocities in jets. |
|
![]() |
![]() |
| The
filter is adjustable and is here reduced to 10 cm/s |
|
![]() |
![]() |
| Further
reduction in the filter below 10 cm/s results in high intensity signals
becoming visible, especially in early diastole. This is tissue signals
from the mitral ring. |
|
![]() |
![]() |
| Fully
removing the filter results in a dense band of high intensity tissue
signals around the baseline. The signal is difficult to analyse, as it
has so high amplitude that the display is saturated. |
|
![]() |
![]() |
| Decreasing
the scale and gain (shown as all signals being illustrated in lighter
colour, but with the same relative placement on the x axis), and
placing the sample volume near the mitral ring,
discloses the tissue velocity curve of the ring, still taken with an
ordinary Doppler. The flow signal, having a much lower amplitude, is
removed simply by reducing the gain. |
|
![]() |
![]() |
| All
modern ultrasound machines today has separate applications for tissue
Doppler which optimises the signal for this purpose, among other things
by applying a low pass filter that removes most of the flow velocities.
This results in a cleaner signal. |
|
![]() |
![]() |
|
| Velocity and strain rate imaging of the same (normal) left ventricle. The colour sector can bee seen to be equal to the B-mode sector.Velocity is red in systole when all parts of the heart muscle moves toward the probe (apex) and blue in diastole. The changes are too quick to observe entirely, to make full use of the information the image has to be stopped and scrolled. | Curved anatomical M-mode (CAMM). A line is drawn from apex to base, and velocity data over time are sampled along the line and displayed in colour along a straight line. The numbers on the curve and the M-mode are included for reference and corresponds to the numbers on the B-mode image. This example shows the septum from the apex to base along one axis, and one heart cycle along the other, in a two - dimensional space - time plot. S: systole, E: early relaxation, A: atrail contraction. | Extracted velocity curves from three points in the septum. As in colour flow, the M-mode gives the depth - time - direction information, while the curves give the quantitative information. |
| A patient with a stationary reverberation
in the basal lateral wall (left), with the sampling point for both pw
and colour Doppler is indicated. . The pw spectral Doppler
(middle) can be seen to be broadened, comprising all velocities between
peak and zero. This is due to stationary pixels due to the artifact,
but a sufficient number of pixels with normal velocity (and a number of
pixels in between). Averaging the spectrum will result in too low
values, as the lowest values are due to artifact. The
autocorrelation values (right) are likewise average values, thus too
low. |
|
a
|
b
|
![]() |
![]() |
![]() |
![]() |
| Kernel displacement Displacement curve obtained by tracking through a whole heart cycle shown to the right, derived velocity curve shown below. | From two different kernels, the relative displacement and hence, strain as well as strain rate can be derived. | With kernels
at
all segmental borders, segmental motion and deformation can be tracked,
as shown to the right. |
And the length variations
of the segments between the kernels kan be followed through the heart
cycle. |
![]() |
![]() |
| Strain rate is displayeed as yellow to orange in systole (shortening) and cyan to blue in the two diastolic phases early and late filling (lengthening), but green in periods of no deformation. The changes are too quick to observe entirely, to make full use of the information the image has to be stopped and scrolled. | Combined strain rate image with one
systolic and one diastolic frame displaued in B.mode, below the CAMM
from the septum and below that the strain rate (yellow) and strain
curves from one point in septum. |
![]() |
Integration
of velocity |
![]() |
| Velocity
traces may be
considered the raw data. All other modalities are integrated or derived
from this. Here normal function is shown at normal basal velocity (6
cm/s), as well as normal decrease from base to apex. This
decrease is evident by visual assessment alone, as the distance between
the curves. The distance between the curves then is a direct visual
assessment of strain rate. Strain rate curves can be obtained by
spatial derivation of velocity: |
Displacement curves
obtained by integration of velocity. Temporal integration reduces
velocity to motion. In principle, strain could be obtained by spatial
derivation of
displacement, although not used: |
|
![]() |
Integration
of strain rate |
![]() |
| The strain rate curves are
the spatial derivative of velocity, showing the time course of the
velocity gradient. This is equivalent to the local deformation rate.
The curves, however, are noisy, shown in this unfiltered image, the
increase in random noise is a consequence of derivation. |
Strain is regional
deformation. This can be obtained both by spatial
derivation of strain (not used at
present) and temporal integration of strain rate
(used). |
| Colour M-mode (CAMM) of tissue velocities
in
fundamental (above) and harmonic (below) imaging. Slight aliasing can
be seen in native imaging in the e' wave at the base. In harmonic imaging, there is aliasing both in the S' wave, and the e' wave (double). |
Colour tissue Doppler curved M-mode in
harmonic
imaging, velocity plot (above), strain rate (below). As can be seen
there is heavy aliasing in the velocity plot, but no aliasing in strain rate imaging. |

![]()
Editor:
Asbjorn Støylen, Contact address:
asbjorn.stoylen@ntnu.no, Updated: 2010