This is explained in more detail
here.
The velocity gradient and SR are
equal to the Eulerian strain rate, which normalises the velocity
difference to the instantaneous length, while it is customary to use
the Lagrangian strain, which normalises the change of length to the
initial length,which is explained in more detail
here.
The transmural velocity gradient can be
measured by the strain rate imaging method, which is quicker than
tracing the endo- and epicardial borders. Measuring the transmural
systolic strain by integrating strain rate, however,
is a roundabout way of
arriving at the relative wall thickening, which can be measured
with
equal temporal resolution and much higher reliability with M-mode. In
addition, anatomical M-mode can imagine wall thickening in other
directions as well, although with only 2D grey scale frame rate.
Limitations of tissue Doppler
The main limitations of the
tissue Doppler method are:
- Noise,
reducing radial
and temporal resolution due to the need for smoothing
- Drop outs;
giving similar effects as reverberations:
- Reverberations
resulting in systematic measurement errors that may invert, reduce or
increase measurement values, depending on position. In general, it may
be assumed that segments more basal to reverberations will tend to
increased absolute values, and thus mean values in the basal segments
may be overestimated. This may be the explanation of an apparent
skewness in the strain values seen in a recent population study (152)
using tissue Doppler in the basal
segments only.
- Angle
deviation, which is
inherent in the Doppler method, but
aggravated by the geometry of strain and strain rate.
- Problems with lateral
resolution, which also are aggravated by poor alignment of the
ultrasound bean and the wall.
Random
noise
As can be seen from the equations given above, the strain rate is
derived as a difference of two velocities. But this means that the
signal (strain rate) has a noise that is the sum of the noise of the
two velocities, while the signal itself is the difference between the
same velocities. Thus, Strain rate has a far less favorable
signal-to-noise ratio that velocities, as ca be seen from the figure
below.

|

|

|
Velocity plot at one time during the heart
cycle, taken from a 4-chamber view of a normal patient. The plot shows
the distribution from the basis of the septum (left) through the apex
to the basis of the lateral wall (right). It can be seen that there is
a clear velocity
gradient along the wall, but also that the velocities are fairly
different from point to point (noise),
resulting in even more noise in the derived strain rate.
|
Velocity curves from a normal patient. It
is evident that there is some noise in the curves, but the curves can
be interpreted very well.
|
Strain rate curves from the same dataset,
showing how the data derived from the velocities multiplies the noise
in the velocity data, due to the spatial derivation process. In
this image, no smoothing is applied.
|
The Subtraction algorithm will give substantial increase in random
noise. The reason for this is the spatial derivation (
11): There
is a certain randomness
of velocity measurement, the limit of precision of measurement. This is
independent of the value measured, so lower velocities has lower signal
(velocity) to noise (variability) than higher values. Strain rate,
being the difference of two velocities (SR = (V
1 - V
2)
/ L) has a random variability that is the sum of the variability of the
two velocities measured, or twice that of velocity. The signal,
however, is the difference between the same two velocities, resulting
in a much lower signal to noise ratio, which is evident in the plots
above. there is considerable variation in velocity measurements.
By calculating the mean velocity gradient along the strain length by
linear
regression, the noise is substantially reduced, compared to the
simple subtraction algorithm (SR = (V
1 - V
2) / L)
and regression is the present method of choice. However, random noise
remains a problem, as seen in
recent
studies.
Random noise in tissue Doppler derived strain rate can be
reduced by:
- Increased
offset distance (strain length).
- Spatial averaging. Both points
will reduce spatial resolution along the ultrasound beam (depth), and
increase the susceptibility for non random noise as f.i reverberations
and also for drop outs.
- Temporal
averaging (curve
smoothing). Reduces temporal
resolution (effective frame rate), and may lead to undersampling.
- Integration
to strain. Changes the physiological
information, but the importance of this for clinical diagnosis is
uncertain. However, the temporal resolution is reduced.
- Averaging
of more than one heart cycle
(cine compound). But this also will
give some interpolation between frames, and reduce the effective frame
rate. In addition, there will be the possibility of introducing non
random noise into the compound cycle. It is argued that averaging peak
values may be better.
- reduce data to parametric images.
In
this modality, the information is reduced to qualitative information
(shortening and lengthening) or semi quantitative information (f.i. colour WMS).
Spatial smoothing

|

|
|
The effect of Strain length can clearly be
seen. Usual default offset at present is 12 mm.
|
ROI size. With offset 4 mm, the effect of
averaging more samples is evident. Usual default is 12 x 6 mm at
SL 12 mm.
|
Illustration
of how both strain length and ROI size reduces spatial resolution.
each strain length is represented by a point in the middle of the
length. In principle the radial (depth) resolution should be ROI
length + Strain length, but as shown the effect of the points outside
the ROI decreases with the distance from the ROI with the regression
method.
|
Ideally, the strain length plus ROI should be as great as possible,
giving the highest possible velocity difference to ensure the best
signal-to-noise ratio. As there may be little added information by
entering into sub segmental resolution, the resolution may be as low as
the length of one segment. This will give the
segmental
strain rate and strain, although measured along one ultrasound
beam. However, this also increases the risk of including areas of
reverberations
or
drop
outs.
This is indeed true of all methods
for spatial averaging.
Temporal smoothing

|

|
Temporal smoothing at SL=4 mm, ROI size 6x6
mm. Left no smoothing, right Gaussian
smoothing at
40 ms. Default at present is Gaussian 40 mm at SL 12 mm and ROI 12 x 6
mm. #Examples of recordings with the default settings can be seen above
at different
places.
|
Effect of temporal integration. Left
unsmoothed strain rate, right strain from the same dataset. As the
noise is random, the summation will eliminate the random variations,
resulting in smooth curves.
|
A detailed treatment of temporal filtering can be found in (
75,
77).
In the original post processing application no smoothing was
applied. That meant that the original studies were done by visually
correcting for noise. Later temporal smoothing, and in the latest
software (from about 2002), the Gaussian
temporal smoothing
is
implemented. In addition, the strain rate algorithm has moved from a
simple subtraction algorithm back to the original (
14)
regression
method, resulting in a more robust strain rate estimate.
The experience from the HUNT study (
153),
seem to show that strain rate by tissue Doppler is a more noisy method
for strain rate than the others, including
segmental
strain, giving higher peak strain rate (probably due to noise
peaks) and wider standard deviations (variability) as shown in the
comparison
study. Strain did not differ,
showing that the integration to strain is efficient in eliminating
random noise.
Averaging more
than one heart cycle
(Cine compound):
Random noise will not repeat from heart cycle to heart cycle, and thus
averaging more than one consecutive heart cycle, will eliminate random
noise. As can be seen from the recent
comparison
of methods in the HUNT study (
153),
peak systolic strain rate by tissue Doppler is much more sensitive to
noise than other methods. This can be seen from the fact that the
average (absolute) peak values are much higher then other methods,
while the standard deviations are correspondingly wider. Thus the peak
values incorporate noise. Systolic strain, on the other hand, are quite
similar, showing that the increased peak values disappears as noise is
integrated sto strain as discussed
above
and that the higher peak values are incorporating noise peaks.
On the other hand, averaging heart cycles has several disadvantages:
- The frames will not be at exactly the same time point in the
cycle, and thus corresponding franes will be from slightly different
points in the cycle, and thus averaged. This effect is in fact similar
to temporal
averaging, and results in a
similar reduction in effective frame reate, and may lead to undersampling.
- As there is significant beat to beat variation in cycle
length, later events in the heart cycle will occur at different
intervals from the R-wave. Systole and
diastole varies differently with respect to RR-interval, especially at
HR < 100 (29),
it will
especially affect the time around end systole / early diastole.This
may lead to bizarre results as seen below.
- Cine compounding by automated methods will always sample three
consecutive cycles, and if one beat is of low qualituy due to motion or
respiratory artefacts, or even as an extrasystole, the compound curves
will include the artefact.
 |

|

|
Native recording,
showing four cycles. (Healthy child, HR around 90). It is evident that
there is variation in heart rate, as it can be seen by the increasing
fusion of E and A waves.
|
Cine compound x 2, i.e. each cycle shown in
a compound of two cycles. Systolic peak velocities does not
change much, but e' wave velocities are almost halved, due to the
averaging of E waves that are at different relative positions in the
cycle.
|
Cine compound x 3, i.e. each cycle shown in
a compound of three cycles. There is not much change from
cine-compound x 2, and still e' waves are very different from the
native loops.
|

|

|

|
Velocity
curves from the septum. There is substatial noise as well as
fusion of E and A due to beat 2 being an extrasystole at the end of
cycle 2 (arrow).
|
Resulting in a high noise spike in the
strain rate signal,
|
which again leads to a substantial artefact
in the strain curve.
|

|

|

|
When this artefact is shown in a
non-compounded image, it is limited to one cycle, and it it fairly
evident that this cyscle should be discarded.
|
Cine compound x 2, extends the effect to
two cycles
|
and cine-compound 3 extends the effect to
three cycles, and in this image, it is not evident which cycles are
representative.
|
Basically, If one is concerned about peak values, averaging peak values
from three cycles will be more robust that automatically creating a
cine compound cycle, and especially if cycles that deviate are
eliminated. Thus, one is not restricted to three consecutive cycles.
Cine compound may seem an attractive way of averaging three cycles, and
then being able to reduce the number of measurements to one set in a
cycle that is the average of three, but then the following caution
should be observed:
- It should only be applied to systole. Also, velocity measurements
are fairly smooth already, while strain rate is noisy and is the case
where one profits most. Thus peak systolic strain rate maight be the
one case where it it feasible and favorable. Strain curves are fairly
smooth already, and cine compound will not add anything.
- It should only be applied where there are three consecutive
cycles of good quality, but as can be seen above, the noise is not very
evident in the velocity curves, and if cine compound is applied before
conversion to strain /strain rate, the artefacts may be less evident.
Amount and method for temporal
averaging should always be reported in clinical studies.
Drop outs
Where there are drop outs, no data can be obtained, and the resulting
velocities will be zero:

|

|

|
Drop
out in the anterior wall. In this area there are no B-mode data,
indicating that there are no velocity data either (although this may
not always be the case).
|
Schematic figure corresponding to the loop
to the left. In this case there are no data in the drop out, resulting
in zero velocity, i.e. v1 = 0.
|
As is evident from the traces, the velocity
in the drop out area (cyan
curve) is zero. Below the
drop out, the velocity curve is normal. |

|

|
The effect on strain rate, however, may
seem a little paradoxical. Below the drop out, there are much higher
absolute strain rates. This can be explained by the subtraction
algorithm, (although a little simplified), as SR = (V1 - V2)
/ L, as V1 = zero, this will result in SR = - V2, i.e. the strain
rate curve becomes an inverted velocity curve. Thus, the absolute
values will be much higher that the real strain rate. This can be
seen in the transitional zone where the strain length crosses the
border between the drop out and the normal area (cyan curve).
Observe how the curve looks almost the same as the normal velocity
curve above (yellow velocity curve). The more basal measurement,
with all of the strain length within normal data, shows normal values
(yellow curve). The curved M-mode shows the distribution of the
effects. In the apex (1), there are no data for strain rate, showing no
deformation. (the whole strain length is within the drop out. In
the midwall (2), there are exaggerated strain rate values, shown by the
colour intensity (red) equivalent to the cyan curve to the left. In the
base (3) there are normal strain rate values (orange) equivalent to the
yellow curve to the left.
|
Stationary reverberations
In
stationary
reverberations, the mechanism is similar to the effect of drop
outs. The various artifacts may arise from the effect of measuring zero
velocities in some areas:

|

|

|
Stationary reverberations shown in B-mode,
the mechanism is explained in the ultrasound
section. Stationary echoes are called clutter.
|
The effect being that in the area of the
reverberation, there are zero velocities (V2), while the
velocities above (V1) and below (V3) are normal.
|
The result is that the reverberation shows
up as a stationary area of inverted colour, showing sharply in the
strain rate image. However, this is not the only effect.
|
Above the reverberation, where
all of the strain length is in normal signals, the strain rate is
normal (orange in systole, blue in diastole. In the area where the V1
end of the strain length is in the normal area, and the V2
end is in the reverberation, and thus = 0, the strain rate will
be SR = (V1 -V2) / L = V1
/ L. This is seen in the yellow
curves, looking like normal velocity and displacement curves. As
systolic shortening is negative, while velocity and displacement are
positive, there is apparent dyskinesia in this area. This area
shows inverted strain rate in the M-mode to the left. Below the
reverberation, where the apical end of the strain length is in the
reverberation, and the basal end is in the normal velocity field, the
effect will be SR = (V2 -V3) / L = -V3
/ L as shown by the deeper colour (red) and the cyan strain rate curve,
equivalent to the inverted velocity curve shown in the drop out
artifact. Thus we have apparent
hyperkinesia. Finally, the red and green curves are both from the
transitional zone, being interpolations between the other artifacts.
The green would be apparent initial dyskinesia, while the red is
apparently normal. However, as the whole area has data quality arising
from artifacts, NONE of the curves should be used, lest there be biased
post processing. The whole area should be discarded.
The problem of reverberations are even
greater in tissue Doppler, as this is done in fundamental, and not
harmonic
mode. Suppression of reverberations by harmonic imaging is not
feasible due to the low frequency giving a low
Nykvist
limit, resulting in aliasing in tissue Doppler. Although strain
rate would unwrap most of the aliasing, this would mean that separate
recordings would have to be made for velocity (fundamental) and
deformation (harmonic) imaging, instead of deformation being post
processed from velocity recordings.
Insonation angle deviation
The main limitation of tissue Doppler due
to the angle dependency is the inability to analyse anything other than
longitudinal strain in all segments of the ventricle. Tissue
Doppler may give transmural strain in the anterior and inferior
segments (crosswise), and circumferential strain in the lateral and
medial segments (tangentially), but full segmental analysis is not
possible. Thus, if the other strain directions should prove to give
added information, this means that at least combined methods would be
preferable. However, this is still not definitely proven.
Insonation angle
deviation means that the ultrasound beam deviates from the direction of
the wall at the point of measurement. This is due to different
mechanisms:
The wall is curved, while the ultrasound beams are straight and
deviates with depths in a fan, as shown in the figures below.
The probe may be placed off centre, although this may not be
apparent in the imaging plane, only in the orthogonal plane (
76).
Off centre placement of the
probe, however, may both increase angular error (76) in the segments
with best alignment, or decrease it due to curvature compensation (
42).
The angle dependency is mainly a problem in tissue Doppler, but
reduced
lateral resolution in B-mode in order to achieve a high frame rate
(or indeed, 3D images), may increase
angle
dependency in speckle tracking by reducing the ability to
track
in the lateral direction.
It is well
known that velocity measurement is dependent on the angle between the
ultrasound beam and the velocity direction (vector) – insonation angle,
and
that the measured velocity vm is reduced in proportion to
the cosine
of the angle a between the velocity vector and the ultrasound
beam as discussed in the ultrasound
and mathematics
sections.
For strain
rate and strain measurement by velocity
gradient, the
angle dependency is somewhat more
complex. In
the longitudinal direction, the longitudinal velocity gradient is
reduced by
the cosine of the insonation angle, as for velocities. But in an
incompressible
object, there is simultaneous strain in the transverse direction, in
order to keep the volume constant, and
the two
strain components are opposite and will detract from each other (1,
2,
7).
This results in
further reduction in the measured strain and strain rate (7).
This is
illustrated below.

|

|
| The true velocity
vector is shown as the straight arrow, the ultrasound beam as the
dotted line. The vector measured along the ultrasound beam is reduced
by the cosine of the angle between the true vector and the ultrasound
beam. |
An object
undergoing longitudinal shortening (negative strain). The ultrasound
beam (dotted line crosses the longitudinal direction at an angle. In
addition to the apparent reduction of the shortening by the angle,
there is simultaneous thickening in the transverse direction (positive
strain), which further detracts from the n numerical value of the
shortening.
It's important to realise
that this
double angle problem is limited to the velocity gradient method. Tissue
Doppler is still angle dependent, but segmental
strain by tissue Doppler has only
the basic limitation common to
all Doppler measurement. And using it in combination
with transverse speckle tracking,
eliminates this angle dependency
also, to the same degree as in speckle tracking (meaning that speckle
tracking is not altogether angle
independent). |
Angle deviation is most usual in the apex and base:

|

|

|

|
Angle deviation is biggest in the parts
that are most perpendicular to the ultrasound beams, as shown here, in
the apex and the base of the sigmoid septum.
|
The sigmoid septum illustrated here,
showing systolic positive
strain (blue - apparent lengthening) due to the angle being crosswise
to
the wall, in reality measuring thickening
|
Another patient with no sigmoid septum
shows no artifact in the base of the septum. In addition, as the
ventricle is not dilated, the area with apparent lengthening (blue) is
fairly small.
|
Strain rate curves from the same patient
showing it top be perfectly feasible to measure strain in the basal
part of the apical segment (red ROI and curve), the area with inverted
values (being in reality thickening) is fairly small (yellow ROI and
curve).
|
This is described in a more detailed mathematical analysis in the mathematics
section.
 |
 |

|
| Which segment that is most in alignment
with the ultrasound beam, may vary with LV shape and depth, as shown
here. |
It has been maintained that strain and
strain rate cannot be measured in the apical segments, but from the
illustrations shown above, this is not necessarily true for all
patients on a segmental level, and the apical segments may in fact be
the ones showing best alignment in some cases.
It's important to realise that this
double angle problem is limited to the velocity gradient method. Tissue
Doppler is still angle dependent, but
segmental
strain by tissue Doppler has only the basic limitation common to
all Doppler measurement. And using it in
combination
with transverse speckle tracking, eliminates this angle dependency
also, to the same degree as in speckle tracking (meaning that
speckle
tracking is not altogether angle
independent).
However, the apex seems most susceptible in the
HUNT
study, but the problem was solved by the ROI tracking the
myocardial
motion.
In post processing, the main point is to exclude segments with to great
angle deviation from analysis, at least other than parametric.
In some instances the angle problem is due to imperfect alignment
(foreshortening), if
the probe is not positioned properly over the apex. (As indeed may be
necessary to obtain an acceptable window). In that case, the angle
problem can vary along the wall as shown below:

Less than perfect
alignment with the apex, results in a, angle along the inferior wall in
this 2-chamber view. It can be seen that the wall apparently is curved,
and that the alignment is better in the basal than the apical half.
|

This has different effect
in the different parts. Basally, there is a normal strain rate curve
(yellow). Apically, the systolic strain rate is reduced to half, due to
angle distortion (red). In the midwall, there is a normal peak value,
but the systolic curve is cut off, resulting in zero values in the late
systole (cyan). as the bent area moves into the ROI.
|
This is also apparent in the curved M-mode, showing an
area of apparent a- to dyskinesia in late systole in the midwall.
|
Using timing information, especially the shifts between positive and
negative strain rate, on the other hand has been proposed to overcome
the angle limitation. This, however, may be problematic if the
alignment is less than perfect, in a way tat the angle between
the ultrasound beam and the wall varies through the heart cycle, as
shown in the midwall segment in the middle image above.
Variable
insonation angle during the heart
cycle
In individual cases, there may also be angle problems in other levels,
especially the inferior wall in the care of foreshortening. The angle
deviation may apparently vary during the heart cycle.
Two chamber view. The apex
is in fact outside
the sector to the left, and the inferior wall appears to have a break
in the midwall. This part is transverse to the ultrasound beam, and
here the measured strain rate will be wall thickening (positive strain
rate), as shown in the diagram, longitudinal shortening (negative
strain rate; orange arrows) in the apex and base, transverste
thickening in the midwall (positive strain rate; cyan arrow).
|

This has a similar effect
as in above, but in this case the alignment is better both in the
apex and the base. In the midwall, there is still normal peak strain
rate, but the systolic curve is cut off.
|

In the curved M-mode, the
area of positive strain (transverse thickening is seen to move with the
wall. The pattern may
resemble a reverberation, but doesn't last throughout the heart cycle,
and the time course is not horizontal.
|
The
motion of the distortion area suggests how to deal
with this artefact by making the ROI track the myocardial motion, as
shown here.
Tracking the ROI:
Commercial software have the option of tracking the ROI manually, but
the tracking could be done by automatic methods as in the
NTNU
application. The tracking eliminated the
systematic angle problem in the apex in the HUNT study.

The same loop as above, showing
normal strain rate curve in the base (yellow), but
abbreviated systolic strain rate curve in the midwall, as the area of
transverse strain moves into the ROI.
|

The same ROI placement in start
systole as left, but now the ROIs are made to track the myocardial
motion through the systole. Thus, the midwall curve improves,
showing normal strain rate through systole, demonstrating the the
finding in a is an artefact. It also demonstrates that tracking makes
little difference in a normal strain rate curve (yellow), except maybe
in the apex.
|
Thus, as opposed to stationary artefacts, tracking may help to keep the
ROI outside the path of moving artefacts, but of course if the ROI is
trackiong into a reverberation, the results will be worse

Normal
strain curve below
the reverberation. The ROI is stationary in space.
|

Same ROI as left, but the ROI made
to
track the myocardial motion, passing through the reverberation
during systole, and strain rate curve can be seen to be cut out and
inverted in that period.
|
Thus the value of tracking the region
of interest depends on the quality of the data.
Lateral resolution
Another problem, being related to the insonation angle, but with
mechanism similar to the drop out and reverberation issues, is arising
from the low lateral resolution of tissue Doppler. This is due to low
line density which is applied in order to achieve a high frame rate, as
discussed
in the ultrasound section. If the frame rate is around 150 FPS,
there is usually less than 20 lines in a sector. This is even more
enhanced by using the
MLA
technique where a broad transmit beam is used, and the signal
is received along more, narrower receive beams. The simultaneity of the
parallel receive beams results in signals partially being received by
the neighboring beam. In addition, as tissue Doppler is acquired in
the fundamental mode as
discussed
above,
side
lobes are more prominent, contributing to reduced lateral
resolution.

|

|
Relation between frame rate and lateral
resolution in tissue Doppler. The numbers are receive lines, this
means that in 4MLA, the number of transmit beams are one fourth. In
reality, the lines have the same width, the data are interpolated
between the lines. (Image courtesy of E Sagberg.)
|
Example of how this affects velocity
measurements, In this image, taken at 150 FPS, the four sample volumes
placed side by side in the base of the lateral wall, generates exactly
the same velocity curve, showing that the data are the same.
|
Initially the default frame rate was 150 FPS in tissue Doppler, but now
the default is intermediate (about 100). The low lateral resolution may
have effects similar to the effects of drop outs and reverberations, as
one of the velocities of the strain length may be from cavity or
pericardium as illustrated below.

|

|

|
Two
wide ultrasound beams are shown in grey, with
the middle marked in red. The velocities that are recorded within the
beam are transposed to the middle line. Strain rate is analysed
along the middle line. In the apex the angle makes the beam miss v1,
and may instead use pericardial velocity (red circle) in strain rate
analysis, thus v1 - v2 will be equal to -v2, too high value. In
the midwall, missing v2 and using pericardial velocity instead, v1 - v2
will be equal to v1, a velocity curve, inverted and too high
numerical values. Here, this is shown in the midwall, but the
base is even more prone to this artifact. In the base is shown another
artifact, the beam misses v1, and uses velocities from the cavity
instead, which are zero, being removed by the low pass filter. v1 - v2
will then again be equal to -v2, accentuated numerical values, being an
inverted velocity curve.
|
Normal myocardial
strain rate curves (orange, cyan and green ROI and curve), show a
fairly
even distribution of strain rate. In the lateral wall a
sample volume too far into the cavity (cyan ROI and curve) will give
high numerical strain rate values, for the reason shown in a, basal. A
curve too near the pericardium (red ROI and curve) shows reduced
values, due to a partial effect of the pericardial influence on v2, as
illustrated to the left; midwall.
Averaging makes this effect partial, pericardial velocity only
detracting from the numerical strain rate values. |
Pericardial
artifact as illustrated to the left in the midwall. Top, systole,
bottom, diastole. Here, near the base is seen high, inverted strain
rate values. |
The effect may
explain
why some authors have found higher strain rate values in the base than
in the rest of the ventricle in normals, which is an absolute artifact.
It may also explain
why some authors have found differential strain rate values across the
wall, which is fairly improbable as discussed
above.
Integrational drift in tissue Doppler:
In tissue Doppler, the
package
acquisition of colour Doppler, results in drift from one package to
the next, as shown below.
Integrational
drift in package acquisition. The curve is a displacement curve.
Ultrasound pulses are shot in packages of at least two pulses, and both
pulses contributes to the velocity estimate, as described in the
basic ultrasound section on colour Doppler. Thus the velocity is an
estimate for the package with a duration of 1/PRF, and the velocity of
motion is
extrapolated for the interval between packages, i.e. 1/ FR. From the
figure, is is
evident that integration of these velocity samples will deviate from
the true values in between packages: In this case there is negative
drift in
systole, positive in diastole, but the end result is unpredictable. The
repeating pattern, however, will ensure that there may be a net
positive or negative drift from heart cycle to heart cycle, which is
more or less linear. The drift within each heart cycle, however,
is not linear.
This effect is not only due to
the mathemathical integration of displacement from velocity, it is
actually also a property of tracking by tissue Doppler, as the position
of a kernel is calculated from the position and velocity in the
previous frame. (this is in reality the same thing). A frame rate of 1
KHz would to some degree eliminate this. Then each velocity could be
calculated by single acquisition i.e. from one frame to the next,
without the gaps due to FR being much lower than PRF. This has been
shown to be technically feasible (
215).
Segmental strain and strain rate.
Segmental strain and strain rate is taken to mean the deformation
measured over a complete segment, giving the average value for the
segment.
 |
 |
| Tracking of
segmental borders. The strain is the relative change in segment length,
and the strain rate the strain change per time. |
Segmental strain in six segment in four
chamber view. Schematically , the motion of the segmental borders
is shown by the arrows, decreasing from base to apex. The segmental strain is the difference
between the motion of the two ends of a segment. The curves are from a
real
example. The blue strain curves show segmental values, while the green
curves are the average of the wall for comparison.Image courtesy of H Dahlen. |
Segmental
strain has several
advantages:
- As measurements are fairly noisy, the
average of a whole segment will tend to be more robust. This will give
a high signal-to-noise ratio as discussed above.
The segmental strain is equivalent to
a strain length equal to the segment length, i.e. about 3 cm. The
segments
are the basic unit for evaluating regional wall motion score (WMS) in
the recommendations of the ASE/EAE (146),
and so far the clinical usefulness of a higher resolution has not been
demonstrated.
- Tissue Doppler measures the velocity gradient along the
ultrasound
beam, not along the segment. Increasing the strain length will reduce
noise, but the strain
length will follow the direction of the ultrasound beam, and this will
give problems where the alignment is not perfect, as discussed in the
pitfalls chapter, under the discussion on insonation
angle and lateral
resolution. Tracking the end of each segment, ensures a better
measurement of the segmental longitudinal shortening.This will make the
method less angle sensitive, as
well as more
similar to the other methods as the measurements are related to kernels
at the segmental borders.
- As long as segmental length is followed by tracking the ends of
the segment, the value will be little affected by smaller artefact's
within the segment as illustrated below.
Limitations of segmental strain
The
disadvantages may be that
- The method will be measuring only two points along the line and
thus
- be less robust than the full segmental average (provided the
data in the segment are good).
- be extremely sensitive to artifacts at the points of tracking
as seen below.
- The low resolution in the
radial (longitudinal) direction.
Sub segmental values cannot be extracted (although they could be
interpolated).
- If the tracking at one segmental border
is poor, it will affect two
segments on both sides
of the border as discussed below.
If the algorithm does not track one kernel
correctly, the strain values
will be wrong for the segments on both sides of the kernel. This is
evident in areas of drop outs or reverberations as illustrated
schematically below.

|
 |
Effect of a reverberation on the border
between the apical lateral and the midwall lateral segment. A
kernel in this area will not track, as illustrated by the
arrow. The next border between the basal and midwall segment
moves normally, leading to an exaggerated shortening of the midwall
segment, while the basal segment shortens normally. (The segmental
strain in the apex is the difference between the apical motion (zero)
and the apparent motion ( near zero) in the reverberation, the
midwall strain is the difference between the apparent motion ( near zero) in the
reverberation
and the (normal) motion of the border below.) This is evident by
the curves (compare to the average curve: The apical curve shows little
strain, the midwall curve shows far more than the wall average, and the
basal shows average strain). Compare with the image above.
Image courtesy of H Dahlen.
|
The kernel is in a reverberation in the
lateral wall, and will not track. In this example the reverberation is
in the border between the basal and midwall segment. the shortening of
the basal segment is exaggerated, and the shortening of the midwall
segment is reduced.
|
Thus, one kernel
tracking poorly will then lead to two segments being discarded, giving
a high discard percentage. This was seen in the HUNT study with
automated analysis, we consider this an
advantage of
the study, leading to
little contamination of the data by artifacts, thus ensuring the data
to be "clean". However, it is a d
isadvantage
of
the method,
leading to a lower feasibility. However, in clinical studies, the
feasibility was around 80%, and in addition showin added diagnostic
value to B-mode (
128).
But basically
a high discard rate ensures higher quality of the
studies.
Segmental strain by tissue Doppler
Instead of calculating strain rate along one ultrasound beam, it has
been proposed (
44)
to calculate
strain rate
from the velocity differences at the segmental interfaces and segment
length along the wall. (Any points will do actually), as shown below.
This ensures that tracking is done at the real segmental borders, and
makes the method less dependent on the alignment of the ultrasound
beam and the myocardial wall. The method is still angle dependent, but
this is the ordinary angle dependency of Doppler, not the angle
dependency of
strain
and strain rate.
The method has not been implemented in clinical use.
Segmental
strain rate, measured by tissue
Doppler, but by
segmental velocities that do not lie on one ultrasound beam, while
strain length is measured along the wall, between the velocity
points.
Tracking is possible, by calculating the displacement from one frame to
the next, along each ultrasound line, and thus, the segment length.
This was a theoretical approach in the time when only tissue Doppler
was available for tracking.
The advantages of segmental strain are present also when using kernel
tracking with kernels at the segment borders using either speckle
tracking or combined tissue Doppler and speckle tracking. Both those
methods are in addition angle independent, as the segment orientation
follows the myocardium, and the strain is simply calculated along the
length of the segment as it is.
Speckle tracking in grey scale
images.
Diving humpback
whale. Each
humpback has an unique (speckle) pattern on the
underside
of the tail (and flukes). Thus each individual can be identified by its
speckle pattern. Photographs at different times and places can thus
track the wandering of each individual all over the area it wanders,
without recourse to anything else than the pattern.
- Speckle
tracking! (In grey whale images). This is thus a method with low frame
rate, giving mainly the
extent of wandering over a long time period (the sampling intwerval).
The basic principle of speckle tracking is based on the interference of
the reflected ultrasound giving rise to an irregular - random -
speckled
pattern. The random distribution of the
speckles ensures that each
region of the myocardium has an unique pattern, a
fingerprint, just as in the whales. The speckles follow the
motion of the
myocardium so when the
myocardium moves from one frame to the next, the position of this
fingerprint will shift slightly, remaining fairly constant.
Thus, if a region (kernel) is defined in one frame, a search algorithm
will be able to recognise the lie sized and -shaped area with the most
similar speckle pattern in the next frame, within a defined search area
(fig. 18c), and hence, to
find the new position of the kernel (
26).
This has
been shown to be feasible in flow (
94)
and strain rate imaging (
95).
The basics of speckle formation and speckle tracking is given in more
details
here.

|
|
Typical
speckle patterns in
the myocardium, demonstrating the differences between different areas
(kernels). The difference in the pattern is the basis for speckle
tracking.
|
Speckle
tracking. the speckle pattern from the first frame (red) has moved to a
new position (green) in the next,
and can be recognised by a
search within the search area.
|

|

|

|
| Kernel
displacement. Following the kernel through a whole heart cycle,
will lead to a displacement curve shown to the right. Temporal
derivation (displacement per time, or frame by frame displacement
divided
by the time between frames) results in the derived velocity curve shown
below. |
From
two different kernels, the relative displacement and hence, strain as
well as strain rate can be derived. The strain obtained by simply
subtracting the two displacements and dividing by the end diastolic
distance is the Lagrangian strain.
To obtain the Eulerian
strain rate, the correction
has to be applied for each frame.
|
If Kernels are placed at the segmental
borders, the result will be segmental
strain and strain rate in six segments
per plane.
|
The advantage of this method is that it tracks in two
dimensions, along
the direction of the wall, not
along the ultrasound beam, and thus is considered angle independent.

Longitudinal speckle
tracking, with kernels at the segmental borders in four chamber view.
|

Longitudinal speckle
tracking, but done crosswise in parasternal long axix view.
|
In principle,
pure speckle tracking is direction independent,
and can track crosswise. This means
true longitudinal strain, as the length will follow the "tilting" of
the segment as well as the shortening as seen from the example above.
In
addition, the B-mode has a far better lateral resolution than tissue
Doppler.
However, the speckle pattern will not repeat perfectly. This is due to
both true out of
plane motion (rotation and torsion relative to apical planes and
longitudinal deformation relative to short axis planes) and to small
changes in the interference pattern. But the frame
to frame change is small, and the approach to recognition is
statistical, the basic algorithms are shown
here.
The greater the change from frame to frame is, the poorer tracking.
This may be especially a concern in higher heart rates (f.i. in stress
echo), where the number of frames per heart cycle decreases. This could
be compensated by increasing frame rate, but at the cost of lower line
density and lateral resolution. Lateral resolution is important in
delineating the speckles in the lateral direction. If the lateral
resolution is low, the
interpolation will result in a "smeared" picture as shown
here,
with speckles
that
are
nor so easily tracked in the lateral direction. In addition the
lateral resolution decreases in depth with sector probes, making
lateral tracking at greater depths doubtful. The poorer the lateral
resolution, the poorer tracking in the lateral direction, and the more
angle dependent the method becomes.
Thus,
as the lateral resolution is far less than
the axial, the two directions are not equal. This means that tracking
in the longitudinal direction is better than the lateral, so the method
is angle dependent to some degree. And increasing the frame rate (for
instance to compensate for high heart rate) reduces the lateral
resolution even more, reducing the angle independence of the speckle
tracking method even further.
Speckle tracking has some fundamental limitations as well, partially
the same, and partially more or less complementary to the Tissue
Doppler limitations.
Fundamental imitations of speckle
tracing
The speckle tracking method has fundamental limitations, just as tissue
Doppler
- Low frame rate, which may lead to undersampling.
Shorter events
like the isovolumic phases may disappear
all together, and peak values may be reduced due to under sampling,
especially diastolic velocities and strain rate. This limitation is
most
important for measuring peak values in diastole,
not so much in systolic strain rate, and systolic strain has least
frame rate sensitivity of all.
- Poorer tracking at higher heart rates may render the method less
useful in f.i. stress echo.
- Increased angle dependency with increased frame rate due to reduced
lateral resolution, and also with
increasing depth in the sector due to decreasing line density with
depth.
In Speckle tracking, this reduces the ability to track
in the lateral direction. The
problem is less than in tissue Doppler, but "angle independence" is
an oversimplification.
- Reverberations is just as great a problem in speckle tracking.
where the echo doesn't move, the application doesn't track
- Drop outs is just as great problem in speckle tracking, where
there is no B-mode data, the application doesn't track.
- Drift
is also a problem, due to imperfect
tracking, as shown below.

|

|

|
Drop
out affecting speckle tracking. The application cannot track where
there are no tissue data, in
this case in the anterior wall and the application doesn't track (the
markings don't move). The inferior wall seems to track normally.
|
In
this case, the segments that do not track, are excluded by both
automated software (2D strain ) and visually, thus giving no colour in
the M-mode and no curves.
|
Reverberation
in the lateral wall affecting speckle tracking. AS is visually evident,
the application does not
track across the reverberation, thus the two segments apical to the
reverberations are seen as akinetic, the basal as hyperkinetic. All
shortening is seen in the basal segment.
|
Drift in speckle tracking - kernel
slippage
If the tracking is less than perfect, the kernel will slip slightly
from one frame to the next;
"kernel
slippage". Out of
plane motion and small shifts in probe position or heart position
(respiration) may contribute, as
may
changing reflexivity of structures due to changes in fibre
direction. Then the tracking will start at the new,
slightly off position in the next frame. This effect will tend to be
non random, as the
imperfect tracking usually will result in the kernel moving a shorter
distance than the tissue, and with further slippage, there may be a
cumulated drift during the cycle as illustrated below:
Kernel slippage in speckle
tracking. The kernel is
defined in frame 1, indicated by the red rectangle. In the next frame,
due to out of plane motion, or simply changes in reflectivity some of
the speckles disappear or have lower intensity in the next frame due to
complete or partial out of plane motion in the B-mode
image. Then the kernel may find a slightly different area as the
new kernel position. (Especially if the tracking is done by the sum
of absolute differences where the identification rests with the
summed intensity within the kernel area). In frame 2, the true kernel
motion is identified by the dark grey rectangle, the tracking, however,
identifies the new position as the red rectangle. Some of the speckles
above the kernel have decreased in intensity, while the speckles below
have all increased. In frame 3, further
changes in speckle visibility results in further slippage, i.e.
slippage in relation to frame 2, which then is a larger cumuated
slippage from frame 1. Two speckles from frame 2 above the kernel
have disappeared, four speckles have decreased in intensity. Two
speckles below the kernel have increased. The true position of the kernel from frame 1 is indicated by the
light grey rectangle, the position of the red kernel from frame 2 by
the dark grey rectangle, and the tracking by the red rectangle.
This is a fundamental property of
speckle tracking, and the drift from start of cycle to end of
cycle may actually be
used as a criterion for
quality of
speckle tracking. And even more advanced comparing tracking
forwards and backwards through the whole cycle , f.i. by cross
correlation. It may be less with a higher frame rate. (Although that
will lead to more angle dependency). If the speckle tracking is used
for calculating a velocity field as the primary variable, as in
2D strain,
the integration to displacement an
strain will result in further drift by cumulating small errors. In
addition
undersampling
is a property of
low frame rate, i.e. B-mode. This reduces peak velocity, and the peak
values is even more reduced if
smoothing
is applied before integration as it is in 2D strain. (In tissue
Doppler, smoothing is applied to velocity and strain rate, but the
integration is done on raw, unsmoothed data, as can be seen from
this
example.)
Segmental
strain, measuring the displacement
of kernels directly, and calculating the strain from segment length, if
free from the drift due to mathemathical integration of velocity /
strain rate, as here displacement and strain is measured directly.
However, the drift by kernel slippage from imperfect
tracking in speckle tracking, or from
package
acquisition in tissue Doppler is still present.
Quality evaluation of speckle tracking
Basically, the quality of the speckle tracking can be assessed
visually, by evaluating how well the kernels follow the tissue motion.
This is facilitated by slowing the replay, or
stopping and stepping the
frames (a method that is strongly recommended in stress echo evaluation
as well). In the basic application, this is feasible, as each kernel
tracks
independently. (This is not always the case in more advanced commercial
applications, as will be discussed later.) Independent tracking also
gives the possibility to replace a kernel that is not optimally placed:

|

|
The kernel is in a reverberation in the
lateral wall, and will not track, thus both the segment below and above
the reverberation will show artefacts.
|
Adjusting the position of the kernel
manually, allows speckle tracking despite the reverberation, if the
kernel
remains outside the reverberation during the whole heart cycle.
|
Thus the position of the kernels can be adjusted. If this is not
feasible however, the segments can be discarded from analysis.
Additional software may give automated quality control. The method used
in the HUNT study checks quality by tracking forwards and backwards,
comparing the tracing both ways by cross correlation (
151),
or by the difference
between backward tracking and start of forward tracking (
127,
128,
151).
It still remains unsolved
which method (including visual assessment) is best.
Segmental strain by speckle
tracking
This method, with kernels placed as
shown, will result in measurement of
segmental
strain and strain rate, segment length being defined as a straight
line. This means that there will be less angle dependency than in
segmental strain rate by tissue Doppler, and thus the tracking may be
better than the above method for segmental strain.
The advantage is the same as in
segmental
strain
generally, being little angle dependent, robust against noise and
little affected by sub segmental artifacts. And kernels may be replaced
to avoid areas of reverberations or drop outs, if possible.
The
fundamental
limitations of
speckle tracking, however, apply.
Segmental strain by combined use of
tissue Doppler and
Speckle tracking.
Modern ultrasound equipment has the
capability of acquiring second harmonic grey scale images with an
acceptable frame rate of 40 - 50 FPS and good lateral resolution,
simultaneously with tissue Doppler data. This opens the possibility of
tracking along the ultrasound beam by tissue Doppler, while
tracking transverse to the ultrasound beam by speckle tracking (
124)
in the grey scale data.
The fundamental advantages as well as limitations of
segmental
strain remains.
 |
|

|
| Combined search by tissue Doppler
and speckle tracking. The kernels are shown as the small,
round, yellow circles. The longitudinal search area along the
ultrasound beam by tissue Doppler is shown in red. The lateral search
area by speckle tracking is shown in white. |
The result is tracking of segmental
borders. The strain is the relative change in segment length, and the
strain rate the strain change per time.
|
Display of segmental strain rate from the
six segments.
|
In addition, the combined method has additional advantages:
- As longitudinal strain is the main issue so far, the advantage
specifically of the combined method is that longitudinal tracking is
done with the high frame rate of tissue Doppler. This may give an
additional benefit, at least in strain rate.
- Doing only transverse search by speckle tracking simplifies the
search algorithm, limiting the search area to a sector extending in the
radial direction and thus reducing the time for the speckle
search.
- The transverse tracking by speckle tracking eliminates the
problems of both insonation
angle and low
lateral resolution of tissue
Doppler alone. Thus, the method is just as angle independent
as speckle tracking, as opposed to the segmental
strain by tissue Doppler alone.
- If the
tracking of tissue (lateral tracking) is poor, the measurements will be
similar to the segmental
strain by tissue
Doppler, meaning that it will be somewhet angle dependent, but
still less than the velocity grasdient method.
- Finally, it utilises the full
dataset inherent in the combined image.
The combined method can be used in different
ways to analyse strain rate
imaging (
127):
- Segmental strain by speckle tracking alone
- Segmental strain by Combined tissue Doppler and speckle tracking
- Strain rate by longitudinal velocity
gradient,
by placing an ROI and strain length in mid segment (in end diastole)
- Letting the ROI remain stationary or
- Letting the ROI follow the segment being tracked by the
combined method
3b is similar to ordinary strain rate by tissue Doppler, 3a is
improved, as the present applications only offer the manual tracking as
a
possibility, while the combined method gives the strain rate. The
tracking has been shown to be advantageous in the apical segments, in a
comparative
study.
This method has
already been shown clinically useful in stress echo (
128),
giving a sensitivity of peak
systolic strain rate for ischemia of 84% and an AUC of 0.9, compared
to coronary angiography, and with a feasibility
at peak stress of 80% of segments. It is also the method used for the
HUNT study (
153),
for automated
analysis. In this study methods 2, 3a and 3b was compared as well, and
compared to 2D strain,
comparisons
shown
below, showing little differences between mean values and normal
variations.
The limitations of the combined method is the same as for echo in
general issues related to image quality, and the general limitations of
segmental strain.
It has been considered a problem that the combined method
is not
commercially available, at present it is a research tool at the
Norwegian University of Science and technology. It has been criticized
by some, that the results are not useful, as they are not transferable.
I consider this criticism only
partially valid.
- Firstly, this is a research tool so far for doing
research into and comparing speckle tracking and tissue Doppler. And
the speckle tracking is "pure" speckle tracking, not advanced
applications utilising a lot of computing to achieve results. The
application makes it possible to compare tissue
Doppler, speckle tracking and the combination directly (151),
and as shown below.
- The HUNT substudy showed little bias between methods.
So far, the two data sets could be combined in different ways, not
limited to the application described above. Further research should be
undertaken to assess this.
It seems rather absurd in the long run, that having
access to high quality grey scale tissue data as well as high frame
rate tissue Doppler data in the same loop, the quality of measurements
will improve by discarding one of the data sets.
2D Strain (AFI) by speckle tracking.
The application known as 2D strain or AFI (automated functional
imaging) by GE Vingmed, is generally seen as a speckle tracking
method, which actually is the basic method. Tracking is
done by the same method, (
sum
of absolute differences). However, the method also has implemented:
User friendly interface (if strain rate imaging by tissue Doppler had
had the same interface from the start, it would probably have gained
more acceptance). Some may call it
seductively
user friendly.
A high degree of smoothing, making the results very robust (but may
have disadvantages)
Method specific processing that imposes specific limitations.
With a greater number of kernels, distributed
both along and across the
wall, each kernel can be tracked individually,
and displacement and velocity can be measured in two dimensions, both
longitudinally and transversally for
each
(
73).
From this, differential motion
- i.e. deformation - can in principle be measured, both in the
longitudinal and transverse direction. The smaller the kernel, the less
certain will the tracking be, but this can be compensated by selection
of kernels on the basis of a stable pattern from one frame to
next. One method of insuring stable tracking is to discard
kernels that are
not present in a sufficient number of frames. In the same way, kernels
that does not move can be discarded, reducing the influence of
reverberations.However, the dependence on recognising stable kernels
from
one frame to next, makes the method even more frame rate sensitive.

2-dimensional
strain by speckle tracking. Each red point represents a kernel
for speckle tracking. Velocity and displacement decreases from base to
apex, and the differential motion along the segment gives longitudinal
strain and strain rate. As the true direction of the motion is tracked
in this instance, the transverse component can also be tracked, and the
differential motion from kepi- to endocardium can also be tracked.,
giving transmural strain and strain rate.
|

2D strain in practice. The
midwall line is used for the longitudinal strain, being an average of
all points in the wall. The ROI follows the wall, the limits can be
seen diverging in systole, converging i n diastole, giving the
transmural strain and strain rate at the same time. The colours show
longitudinal strain rate, green is shortening and red is lengthening.
|

In order to make the
speckle tracking more robust, values are averaged over a whole segment.
|
However, this is one of the main limitations of the data in the bottom:
The smaller the kernels, the greater the uncertainty of position, and
the more noise. Thus, a liberal amount of smoothing has to be done.
Averaging a large number of kernels may make tracking more
robust, although this reduces the number of useful speckles in each
kernel. This can be done in various ways and combinations. With more
than one layer of kernels across the wall, the longitudinal
measurements can be averaged from all layers, giving a transmural
average. Longitudinal averaging can be done along one segment, giving
the segmental average. This can also be done in a more sophisticated
way, by spatial interpolation along the wall. This will result in a
gradual effect of spatial smoothing, although the extent of the
smoothing is less easily discerned. Thus, the tracking of a region of
the ROI, is not speckle tracking alone, but also extrapolation of the
AV-plane motion
i.e. virtual tracking.
This is discussed
below.
The method starts with generating a velocity field across the sector
(velocity being the displacement between one frame and the next, /
1/FR), more or less as in tissue Doppler, but the velocity vectors
being angle independent. Thus the displacement ad strain rate and
strain can be derived in the same way.
 |
 |
Longitudinal
velocity
|
Longitudinal
displacement
|
 |
 |
Longitudinal
strain rate
|
Longitudinal
strain
|
Thus, it is evident that both tissue Doppler and 2D strain will give
four modalities from one dataset:

|

|
2D strain derived displacement, velocity,
strain rate and strain. Note the smoothed curves.
|
Tissue Doppler derived velocity,
displacement, strain rate and strain. Note that these are unsmoothed
curves.
|
Quality evaluation of tracking in 2D
strain
The first step in processing will be the quality control of the
tracking, as described above. The application has built in automated
assessment of the quality of tracking, with methods similar to the ones
above. Thus the application will suggest acceptance or rejection of the
segments in the view. However, the segmental quality assessment is
compared to the average, there are no absolute criteria, so tracking
should still be evaluated visually, and segments that do not rack,
should be excluded manually.

|

|

|
Visualisation of the tracking. Observe how
the bullets in the midline follows the myocardial motion. However, due
to the smoothing
function of the
application, this
may be virtual
tracking, being
extrapolated from AV-plane motion, thus
the true tracking of the local tissue may be difficult to assess.
|
Loop
from another patient. Tracking seems fair, visually.
|
Automated
quality assessment accepting the tracking results from all segments.
The assessment is presented as a suggestion, and manual acceptance is
required, i.e. the accept button has to be pushed in order to process.
|

|

|
In this case, the application suggested to
accept all segments, which would result in the values as shown.
The poor tracking in the apical lateral segment does not reflect in the
values, due to the spatial spline
smoothing in the application.
|
The same loop as left, but manually
override of the automated quality assessment, excluding two segments
before pushing the accept button,
lateral basal as it exaggerated the tracking by tracking side
lobes, and the apical lateral, as the tracking was poor due to near
field reverberations.
|
We did an initial evaluation of an earlier version of
this application in February 2004, comparing the longitudinal motion
and deformation measurements by this application with those obtained by
tissue Doppler, in separate images. The study consisted o0f 20 patients
with a wide range of function.

|

|
| Strain
rate and
strain, comparison of 2D strain and Tissue Doppler. There is a
considerable spread between methods, but most probable due to
variability of especially of tissue Doppler. There 2D strain gives
lower values than
DTI, and this tendency increases with increasing strain rate/strain.
The term "CEB" meaning "computerized eye balling" was an early term to
describe the application. |
When measurements was sorted in quartiles, Concordance was only
between 27 and 34%. Feasibility was the same with 2D strain and TVI.
Further investigation was not undertaken at that time, as the
application was modified in later versions. Other authors have found a
much better correspondence between TDI and
2D strain (
73),
with correlations of
0.94 and 0.96 for strain rate and strain, respectively. However, as
seen by the curves in the figure below, both data sets are analysed by
the 2Dstrain software, and thus subject to the same high degree of
smoothing, so the results do not reflect independent analysis.
From a validation study where
tissue Doppler and 2D strain derived strain rate (left) and strain
(right) values were compared. However, as can be seen
from these curves, both curves are very smoothed and concordant. Thus,
much of the concordance must be assumed to be due to smoothing, as both
methods were processed by the 2D strain software, and not by
independent analysis software. Adapted from Modesto 2006 (73).
This is illustrated below:
The lines looking smoother, is a
function of the averaging function
used in the algorithm, the application will do the same to tissue
Doppler data.
Strain rate
curves from speckle tracking and tissue Doppler from the same cine -
loop. The same smoothing is applied to both, showing that
smoothing of the curves is not the result of the robustness of the
algorithm, but of specific temporal and spatial smoothing applied by
the application. The curves differ
somewhat (but not too much), as strain rate is calculated with
different
angle and lateral resolution.
Another study by Cho et al (
148)
finds only correlations of longitudinal strain by 2DS and TVI with MR
tagging of 0.51 and 0.40, respectively. This may reflect the real
precision of both methods (and of MR tagging as well?) but then the
correlation between the methods cannot be higher.
Limitations of 2D
strain
The main limitations of 2D strain are the
general
ones, as well as the ones specific to
speckle
tracking.
In addition, 2D strain has method specific limitations related to:
- Curvature
dependency, due to the technicalities of the specific applications,
which may give too high values in the apex.
- Smoothing,
relying heavily on AV-plane motion,
- which may give strain values even where there are none, and may
reduce sensitivity for reduced regional function
- Makes the tracking more difficult to assess visually
Reverberations and drop outs are illustrated
above.
Those examples are all taken from
2D strain, but shows the general principle.
Curvature
dependency of 2D strain
The longitudinal values that are obtained by the 2D strain
application are curvature dependent, as shown below:

|
 |
Curvature
dependency of strain measurement. If the ROI is curved, the midwall line
will move inwards, and thus shorten, even if there is no shortening of
the segment. This will result in an apparent shortening of the segment
itself, adding to the real longitudinal shortening. This curvature
effect is dependent on the curvature, the width and the widening of the
ROI.
|
Curvature
dependency of strain in 2D strain by speckle tracking. The two images are processed from the
same loop, to the right, care was taken to straighten out the ROI
before processing, while the left was using the default ROI. In both
analyses the application accepted all segments. It can be seen that the
apical strain values are far higher in the right than in the left image
(27 and 21% vs 19 and 17%). However, the curvature of the ROI
even affects the global
strain, as also
discussed above in the basic
section.
|
The width and thickness as well as curvature of the ROI is non
standard, defined ad hoc. In addition, the ROI width is uniform from
base to apex, while the myocardium is thinner in the apex, giving a
discrepancy between ROI width and wall thickness. As the curvature
effect is also a function of ROI width, this may add to the curvature
effect. This effect may account for the observed base-to-apex gradient
of
strain values observed in some studies. The combined method (and indeed
tracking of segment length by speckle tracing alone without TDI by the
same application) is curvature independent as shown
below.
It may be the reason
why some authors find a
base-to-apex gradient in the strain values obtained by this
application, while we did not in the
HUNT study.
Another instance of the
curvature dependency of measured values. The left image is processed
with fairly straight ROI in the apex. The middle image is the same loop
processed with more curved ROI, in both cases the application suggested
acceptance of the tracking in all segments. AS opposed to the
above example, in this case the global strain is severely affected by
the ROI shape as well (.15.7% vs - 20%). To the right is shown another
loop from the same patient, centered on the right ventricle. In this
case, the values of the septum is quite similar to the values in the
left image, but differs from the middle image. The interesting thing is
that the global strain itself is different from the mean strain
calculated from the segmental values.
This may affect the regional strain as well, as the curvature
dependency may assign higher values to akinetic segments as shown below:

|
Small
apical infarct. Admitted with a history of pain, but free of pain and
with normal ECG, but
elevated Troponin (analysis
results not ready till he had new pain) at
the time of admittance. This Echo at admittance was initially
considered normal, even though by retrospective evaluation there is a
small area of hypokinesia in the apex. He then had recurrent pain
after a few hours, with
ST-elevation.
|
 |
 |
Tissue Doppler based strain rate and strain
showing hypokinesia in the apex (yellow and red curves) , peak systolic
strain of - 5% and -8%,
strain rate of - 0.35 and -0.8 s-1 both segments with post
systolic shortening, as contrasted with normal
deformation in the base (green and cyan). This is an indication of a
small ischemic
insult at the time of the first pain episode as also shown by the
troponin results.
|
Angiography at the time of recurrent pain
showed a tight LAD stenosis (top), confirming the strain findings, it
was treated with PCI and stent (bottom)
in the same procedure. Strain and strain rate values were normal after
one week.
|

|

|
2D strain of the same recording (B-mode
loops without TDI). The curved M-mode gives apical strain of -14
and -15%, i.e. borderline normal. This was the default ROI.
|
Adjusting the ROI making the apical
segments straighter, reduces apical strain to -9 and -11% (borderline
abnormal)
|
Both images were made with default (medium)
spatial smoothing, but the values did not change more than 1% by
reducing smoothing to minimum. In this case, the curvature effect is
probably more important than the smoothing,
although both factors may contribute.
|

|

|
 |
Inferior infarct in a rather foreshortened
view, resulting in a spherical image. By visual assessment this infarct
is akinetic in the basal segment
|
Strain by tissue Doppler, showing systolic
akinesia in the basal segment (cyan curve) - mark how the ROI is placed
to avoid the lower part of the segment where there is angle
discrepancy), and normal strain in the apical segment (yellow) and the
anterior
wall (red).
|
Strain
by 2D strain, showing borderline reduced, but still viable strain of -
12% in the
basal segment. IN this case, the near akinetic segment has a strain
that mainly is due to the inward motion as described in the diagram above.
In addition, the
ROI, being the same all the way around, overestimated the wall
thickness in the infarct, also contributing to the curvature dependent
strain, which is dependent on the ROI width. In this case, the effect
is due to the curvature, not smoothing,
reducing smoothing did not reduce strain in the infarct zone at all.
|
.
The curvature dependency of 2D strain
is a parallel to the angle
dependency
of tissue Doppler.
Smoothing in 2D strain
There is a liberal amount of temporal smoothing. In addition there is
built in a spline or polynomial smoothing along the whole region of
interest (ROI). The AV plane is the heaviest feature that is tracked,
and contributes the most to the motion, which is then distributed along
the whole ROI by a curve fitting
along the mid ROI curve, resulting in a
smoother transition from segment to segment, distributing the
deformation along the ROI. Thus, the 2D strain application does not
reflect pure speckle tracking, but also a great amount of model
fitting. The spline smoothing is weighted, being least in the basal
segments, most in the apex, which may be a way to compensate for some
of the curvature effects in the apex, but that means tat adjustments in
the base will affect measured strains in the apex.
The
spatial smoothing, however, is adjustable. By default, the smoothing is
medium, and can be adjusted to both maximum and minimum. However, in
earlier versions of the software, the regulation of the smoothing did
not carry over to the measured segmental values displayed in the quad
screen view. Thus, segmental values remained medium smoothed, and
values with less smoothing had to be taken from the traces. This
problem seems to have been fixed in the latest version of the software
(2011), but this means the software version should be taken into
account. Also, the adjustment seems to be small, compared to the total
amount of smoothing.
This is illustrated below.

|

|
| Spatial
smoothing in 2D speckle tracking strain.
The
smoothing, using longitudinal information from the AV-plane motion
results in strain values even in areas where there
is no speckles. This means that there is movewment of the points of the
ROI, in areas without speckles. This is due to the information from the
AV-plane motion being "splined" along the ROI, and the motion of the
points is due to this, not to local tracking. Thus, there is only
"virtual tracking", and the true local tracking cannot be fully
assessed . |
Segmental values from this tracking. Left:
medium (default) spatial smoothing, right: Minimum. This was done
in the latest software version (2011), where values and traces carry
over to segmental values after adjustment of smoothing, but the effect
is small, compared to the total amount of smoothing.
|
This example was obtained by manual override of the automated
positioning of the ROI, as well as manual override of the automated
quality
check, which suggested rejection of all segments. However, the
application
did track, as can be seen
above, left, and these images are to show the general principle.

|

|
Inferior
infarct in two chamber view, being akinetic in the basal inferior wall,
although with considerable passive motion due to tethering
|
2D
strain
(left) vs. tissue Doppler (right)
in an inferior infarct, analysed from the same cine loop
recording. The akinesia in the base is missed due to smoothing.
In this case, as there is dropout of the whole anterior wall, the
smoothing may be harder in the remaining three segments. Also, in
basal infarcts, the effect on AV-plane motion is less (40)
as shown above.
|

|

|
Inferior
infarct. Hypokinesia of the basal segment. Not immediately evident.
|
Strain and strain rate. Basal hypokinesia
and post systolic shortening (yellow). Also normal curves in the
inferior apex
as well as in the anterior wall (red and cyan).
|

|

|
Same infarct as above. Tracking shows poor
tracking in the basal and midwall segments have poor thickening due to
poor tracking. The anterior wall is less visible. The segments are not
approved for analysis.
|
Longitudinal strain. The apical anteior
segment shows reduced strain, but this is due to poor tracking. The
basal segment does not show reduced systoloic strain. However, looking
at the curve, the infarcted segment does show post systolic shortening,
so the infarct is still indicated.
|
Lateral tracking in 2D strain
Transverse displacement an velocity can also be derived, but as this
will be the segmental average, this value has little meaning, the
velocity and displacement increases from epicardium to endocardium. It
is the displacement and velocity
gradient that is of interest, i.e. transverse strain and strain rate.
However, this can also be calculated by this method:
|
Longitudinal |
Transverse
|
Strain
rate
|
 |
 |
Strain
|
 |

|
Longitudinal and
transverse strain derived from speckle tracking. It can be seen
that in this case the differential tracking in the transverse direction
is poor in the basal segments, thus underestimating transverse
thickening in this healthy subject. (Images with better resolution may
be seen above).
This is one of the fundamental limitations of speckle tracking as
discussed
previously.
Transmural and
circumferential strain.
As speckle tracking is partially angle
independent, it may be applied to the
short axis as well. The main concern about tracking in short axis
views, however, is the long axis motion. This means that there is
between 1 and 1.5 cm out of plane motion of the base, and about half
that in the midwall. That means that the tissue present in end diastole
is not the same as in end systole. This also means that
the speckles that are tracked do not
represent physical myocardial points. Thus, the meaning of
transmural and circumferential strain becomes slightly dubious.
However, this do not only pertain to 2D strain. As shown
above,
this is the same problem even in
parasternal
M-mode. (Which, despite this, has worked well for 50 years). However,
this remains
a caveat when new measures are added. In the question of rotation,
especially torison, the spiral course of the longitudional fibres may
even cause the displacement to cause the fibres to be traced as
rotating around the cavity centre.

|

|

|
Parasternal long axis image. The
longitudinal motion of the basis is evident.
|
End diastolic parasternal long axis
image. The yellow line crosses the ventricle near the middle.
|
End systolic parasternal long axis image
from the same loop. The yellow line crosses the ventricle much
closer to the base transecting a different part of the myocardium.
|
The speckles may be the endocardial
borders, or even the fibres
that may run in spiral. Thus, in the base, the physiological meaning of
the obtained values is questionable.
Accepting the validity of speckle tracking in short axis views, it then
allows tracing of transmural and circumferential strain.
Transmural
strain is
wall
thickening, and the tracking in the transmural direction
will be dependent on the resolution, which is better along the
ultrasound beam than laterally. The physiological meaning of
circumferential strain, shouold be midwall
circumferential shortening, which actually is nothing more than

* midwall
fractional shortening as reasoned
above.

|

|

|
2D strain applied to
short axis image. Again this can be seen to track in two dimensions,
the thickness following the wall thickening, and the mid line in the
ROI Showing midwall circumferential shortening.
|
Transmural strain. In this image
the application only measures between 10 and 15% transmural strain,
while the true values in a normal person as this may be as high as 40
- 50%. This is probaly due mainly to a too thick ROI (default),
although poor lateral tracking combined with smoothing
may contribute.
|
Circumferential strain from the same
processing. In this image about 15%, which is closer to
normal. This, however, does not mean that the circumferential strain is
more reliable, it means that the thickness error in the ROI is
compensated by an underestimation of the cavity volume. It's equivalent
to the fractional
shortening increasing in
hypertrophy, despite reduced wall thickening. (Actally
circumferential strain = * midwall FS. )
|
Width of the ROI
Transmural strain all
thickness and wall thickening. But in the 2D strain application, this
means ROI width as shown below.

|

|
Normal ventricle in short axis view.
|
The loop can be used to
generate an anatomical M-mode, the line is skewed to avoid the
papillary muscles. On this M-mode the following values were measured:
LVIDD: 53mm ,LVIDS: 36mm, giving a FS of 32%, IVDS 7 mm, IVSS 11mm,
giving a wall thickening of 57%, LVPWD 8mm, LVPWS 11mm, giving a wall
thickening of 38%, and a mean wall thickening of 48%.
|
Below are shown transmural strain by 2D strain with different ROI
width. The images are all processed from the loop above, and
endocardium traced
in standard manner. In reprocessing, only ROI width was changed without
changing the initial contour. All ROI's were accepted by the analysis
software for all segments:

|

|

|

|

|

|
Transmural strain with narrow ROI setting.
Tracking seems fair, except perhaps in the lateral wall.
This is due to drop out of the wall, not hypokinesia, as seen to the
right, where tracking is far better. However: Given poor tracking
in the lateral wall, mean WT = 57%. This is due to the tracking in the
septum, which is good, but where the wall thickening is absurd, 77 and
91!; normal endocardial motion in absolute terms, gives a
too high relative wall thickening in percent of the narrow ROI.
|
Transmural strain with default (medium) ROI
setting. Tracking
seems fair. Mean
WT = 58%.
|
Transmural strain with a wide ROI setting. Tracking
seems fair. Mean WT =
37%, because a normal endocardial motion in absolute terms will result
in a low
percentage of the too wide ROI.
|
The measured wall thickening is evidently
as expected a function of diastolic ROI width, as expected. Compare
also mean and the relevant segments with the values above
|
It is evident that the transmural strain is extremely sensitive to the
ROI width. This is pertinent to long axis analysis also, as the
curvature
in the apical
segments will lead to an increased susceptibility of the ROI width.
This may be some of the reason why Becker et al (
212)
found transmural strain even in
segments with total transmurality of scars, and not tethering as
presumed.
Repeatability of 2D
strain.
Basically, the 2D strain application, due to a high amount of
smoothing,
should have a high
repeatability, as shown
here. However,
this will
only be the case as long as the tracing is done in the same manner each
time, in the same loops. This means a very standardised endocardial
tracing, and a standardised ROI width. As shown above, the values are
extrmely dependent on the ROI, both
curvature
and
width
of the ROI. Utilising the automated features of the application will
ensure this, but will not necessarily ensure the correct shape
and width of the ROI, and hence, not necessarily the correct
values either. In a study (
208)
where repeated measurements in
the same loops was compared for different centres, the 95% limits of
agreement were -11.4% to +11.8%, but with very little bias. Repeated
recordings within one hour (presumably by the same observer), had
limits of agreement of -9.6 to + 9.7%.
Both segmental strain and 2D strain have been compared for longitudinal
strain, and compared
to tissue Doppler (
151,
153)
as shown in
this
table. Both seem
to agree fairly well. In addition variability of strain rate (but not
strain) is lower by both methods
than by tissue Doppler. However, both Segmental strain and 2D strain
use automatic
segmentation, this may be some of the reason for better repeatability,
not
speckle tracking vs. tissue Doppler per se. However, the higher
variability of strain rate by velocity gradient, shows this method to
have a somewhat higher noise componenet. Feasibility of both methods
is reported to be between 70 and 80% of segments (lower in the HUNT
study,but this is due to the aim of the study, to provide normal values
as free as possible from artefacts.
Summary of differences and
limitations
of different methods.
It is important to be aware of the limitations of each method. It
should also be emphasized that different methods are not necessarily
directly comparable, and may yield different normal values and cut
offs, due to the different ways parameters are measured. One of the
fundamental differences stem from the different geometrical assumptions
that are present as shown below:
Differences in geometry between methods. The fairly invariable outer LV
contour is shown in heavy black. The diastolic inner contour, segmental
borders, kernel positions and measurement lines are shown in light
black. Systolic inner contour, segmental borders, kernel positions and
measurement lines are shown in red. Left: Segmental strain by tracking of
kernels at segmental borders. It can be seen that the main deformation
is measured along the longitudinal axis of each segment. As the wall
thickens, the longitudinal mid line of the segments moves inwards, but
in the basal and mid wall segments this does not add to the shortening
as the angle does not change much. In the apex, however, the
angle of the center line changes, contributing to the segmental
shortening when it is measured by this method, however, the effect is
slight. To the right is shown the geometric assumptions of the 2D
strain method. The ROI uses an assumption of equal thickness from
base to apex, and the mid line moves with the thickening of the
contour. The segment length is measured along the curved line,
and both the curvature and the angle contributes to the shortening of
the segment mid line as it moves inward. Thus, the shortening
(strain) might be expected to be higher in the apical segments by
this method, as well as being dependent on the curvature, especially in
the apex. (However, this effect may be masked by the high degree
of smoothing inherent in the application, which may distribute the
differences between segments. Ultrasound beams are shown in blue,
illustrating the alignment problem of this method, thus resulting
in lower values in segments that are poorly aligned.
The main limitation of any echo method is the ones related to data
quality.
AS discussed under each method;
- The fundamental limitations related to all methods are the ones
arising from:
- Tissue Doppler, having the advantage of high frame rate has
additional limitations related to:
- Speckle tracking (in any form), being less angle dependent has
additional limitations relating to:
- Frame
rate, with the risk of
undersampling.
- Tracking
quality at high HR
- With decreasing lateral resolution the method becomes more
angle dependent, although to less degree than tissue Doppler.
- Segmental strain, being robust and giving the opportunity of
utilising both
tissue Doppler and speckle
tracking and eliminating the angle problem, has the additional
problem of:
- The 2D strain application, being robust and user friendly, has
the additional problems of:
- Smoothing,
relying
heavily on AV-plane motion,
- which may give strain values even where there are none, and
may reduce sensitivity for reduced regional function
- Makes the tracking more difficult to assess visually
- Curvature
dependency, due to the
technicalities of the specific applications, which may give too high
values in the apex.
- ROI
width seems to be critical,
espacially in transmural strain.
Comparison between methods in HUNT
In a recent study of normal values (
153)
we have compared the different methods for deformation measurement in a
subset of 57 patients :
- The combined tissue Doppler - speckle tracking method described above,
- Longitudinal velocity
gradient from
tissue Doppler without
tracking of the ROI, this is similar to the longitudinal velocity
gradient by commercial software, although in this application obtained
by the experimental software.
- Longitudinal velocity gradient with
tracking of the ROI, which can be done, althoufgh only approximate, by
manual adjustment in commercial software, and
- Speckle tracking with the 2D strain
application.
The velocity gradient is analysed by customised software, but the basic
principle is exactly the same as in commercial software (EchoPAC),
except allowing for automated analysis and automated tracking. The
results were as follows:
|
Method
1: segment length by TDI and ST
|
Method
2: Velocity gradient (stationary ROI)
|
Method
3: Dynamic velocity gradient (tracked ROI)
|
Method
4: 2D strain (AFI)
|
|
Peak
Strain rate
|
End
systolic Strain
|
Peak
Strain rate |
End
systolic Strain |
Peak
Strain rate |
End
systolic Strain |
Peak
Strain rate |
End
systolic Strain |
| Apical |
-1.12 (0.27)
|
-18.0 (3.6)
|
-1.46 (0.85)
|
-14.6 (9.0)
|
-1.31 (0.73)
|
-17.2 (9.1)
|
-1.12 (0.37)
|
-18.7 (6.6)
|
Midwall
|
-1.08 (0.22)
|
-17.2 (3.2)
|
-1.29 (0.56)
|
-18.2 (7.4)
|
-1.40 (0.58)
|
-16.9 (7.1)
|
-0.99 (0.23)
|
-18.3 (4.7)
|
Basal
|
-1.03 (0.24)
|
-17.2 (3.5)
|
-1.71 (0.94)
|
-19.6 (9.3)
|
-1.59 (0.74)
|
-17.1 (8.6)
|
-1.12 (0.36)
|
-18.0 (6.2)
|
Mean
|
-1.08 (0.25
|
-17.4 (3.4)
|
-1.45 (0.79)
|
-17.7 (8.5)
|
-1.43 (0.67)
|
-16.7 (8.1)
|
-1.07 (0.33)
|
-18.4 (5.9)
|
Comparison between methods.
Standard deviations in parentheses. Thanks to Eirik Nestaas, MD, PhD
for discovering a typographical error in this table, that now has been
corrected (bold types).
Looking at the findings, it is evident that the tissue Doppler methods
gives close to 30% higher peak strain rate values that the two other
methods.
This is probably due to a higher random noise component in tissue
Doppler, rather than the opposite, too low peak values due to under
sampling in the two other methods. This is evident from two reasons:
- Tissue Doppler derived strain rate shows a far wider standard
deviations
- Integrated strain from strain rate eliminates the differences,
showing that the noise is random.
Thus the
tissue Doppler is more sensitive to noise than other methods.
However, systolic strain values were very similar with all methods
(except in the apex for TDI with fixed ROI), showing that the smoothing
that is a function of temporal integration eliminates this problem, and
basically in strain measurements tissue Doppler is as reliable as other
methods, although still with somewhat higher standard deviations.
Another thing is also evident: Tracking the ROI in tissue Doppler
results in equal strain values in apex, midwall and base, as in the
other two applications, while no tracking yields lower values in the
apex. This is due to the
variable
angle in the apical segments, as the segment becomes shorter, they
also become more crosswise. so there is an advantage by tracking,
but only in the apical
segments. There was no difference in strain rate, only in strain, but
as peak strain rate is early in the systole, while peak strain is
(near) end systolic the effects of tracking may be greater. The reason
is probably poorer alignment in end systole if the ROI is not tracked.
In a study of the sensitivity of strain rate imaging in stress echo (
128),
no significant difference was
found between the segment length and the dynamic velocity gradient,
despite the higher noise component in velocity gradient, nor between
peak systolic strain rate versus end systolic strain.
Finally it seems that the combined method and 2D strain gives almost
the same results,
, although
there was some statistical differences, these were small and of
little clinical importance. Standard
deviations, as a measure of
variability were also comparable, in the combined method probably due
to low noise because of low spatial resolution, in the 2DS probably due
to smoothing.
But
it seems
that the
normal
values are transferable,
and for strain between methods.
In a reproducibility study (154)
the
two methods also had similar inter observer reproducibility (different
recordings and analysers).
Other authors have found a base
to apex gradient (highest strain in the apex) in strain values with the
2D application (
155),
but this may
be due to the
curvature
effect
in the apex, although MR studies report the same finding. Strain
measurement, however, is still dependent on the analysing method,
and
even the definition may vary, depending on how the strain length
is
defined.