Measurements of strain and strain rate by ultrasound

by Asbjørn Støylen, dr. med.

Contact address: asbjorn.stoylen@ntnu.no






Website updated: October 2011.     This section updated: October 2011.

Links in this section have now been repaired.


Other sections:

Basic principles of ultrasound and scanner technology.

This section is important for the understanding of the basic principles described in detail in the section on measurements of strain rate by ultrasound. The priciples will also be useful to gain a basic understanding of echocardiography in general, and may b read separately even if deformation imaging is not interesting. .

Is deformation imaging useful?


This section deals with the approach to using deformation imaging by ultrasound in a practical way, as well as the accumulating clinical evidence for the utility of the methods.

Mathematics of strain and strain rate

Back to main website index

Measurements of strain and strain rate by ultrasound

 In this section, the limitations of the various methods are discussed in more detail. However, it should be emphasised that the fundamental limitations arising from the data quality, are common for any method. The mechanisms for how these limitations affect each application may differ, and thus the effects may be slightly different. In addition each method has specific limitations that are discussed under each method.

However, the main point is that areas with poor data quality should not be analysed by any method. Garbage in, garbage out!

This is especially shadows and reverberations.

Fundamental limitations of ultrasound.


Reverberations




Double reverberation in the lateral wall.
Stationary reverberation in harmonic (left) and fundamental (right) imaging. Tissue Doppler is taken in fundamental imaging, due to the Nykvist limit.

The mechanism for reverberations are described in detail in the ultrasound section.

Shadows (drop outs)



Shadow causing drop out in the entire anterior wall in two chamber view. The drop out is caused by a shadow with a certain distance from the probe aperture, for instance a lung edge or a costa below some subcutaneous fat.
Shadow very close to the probe aperture will not cause drop out, but reduces the effective aperture, resulting reduced lateral resolution, as seen in the image to the left, wile a slight repositioning of the probe removes the problem as seen to the right. However, this is not always possible, if the space between costae is narrow and the subcutaneous fat is very thin.


The mechanism for the effects of shadows is shown in more detail in the ultrasound section.

Sampling rate:

The basic problem of sampling rate is related to measurement of peak values. If the sampling rate is much lower than the rate of change of peak values, the probability of hitting the peak is low, and on average, the measurement will underestimate peak values, called undersampling. Undersampling will also increase variability, by introducing a random element in hitting the true peak value. On the other hand, if the rate of change is slowest at peak, the sampling rate does not matter much. The principle is shown below.


The effect of sampling rate. The true curve is shown left, with the sampling points shown on the curves and below the curves. To the right are the reconstructed curves, from interpolating between the values at the sampling points.  Top: low, and bottom: high sampling rate. It is evident that the sampling rate art the top fails to hit the true peak value, and the peak value shown is underestimated. In addition, hitting the true peak is a more random process, so undersampling will also increase variability. Bottom, the sampling rate is doubled, interpolating sampling points between those at the top (you may measure with a ruler, if you want) and hitting the peak fairly well. The reconstructed curve resembles the original much more closely.



In post processing tissue velocity data, the sampling frequency is equal to the frame rate. The shorter the duration of the event to be measured, the higher the frame rate ought to be. Analysis by decimating tissue Doppler images, showed that peak systolic velocities had both reduced mean values as well as reduced reproducibility, at frame rates below 70, isovolumic indices below 100 (74). However, this is not the only problem. Temporal filtering used to decrease random noise, also reduces the effective frame rate, so even if the frame rate is sufficient applying temporal filtering will then reduce peak values and increase variability (77). Tissue Doppler has a frame rate of ca 100, B-mode ca 50 in a full sector image.

Thus ideally, increased temporal filtering should be compensated by increased frame rate.

Frame rate can be increased to about 150and 90, respectively, but at the cost of lateral resolution which affects both tissue Doppler, inducing artifacts, as well as  speckle tracking in the ability to track laterally as shown also for 2D strain, which also increases angle dependency of speckle tracking.

However, it is important to realise that this is relevant for the measure of peak values. To detect the events and measure duration on colour images, the frame rate only needs to be adequate to detect changes from positive to negative values consistently 50 FPS is generally sufficient.  The temporal resolution of time measurements, however, will depend on the frame rate rate, 50 FPS gives a resolution of 20 ms.

It important that strain is less frame rate sensitive, as the rate of change is lowest at end systole. This means that reduced sampling frequency is less important. However, one study did find that frame rate had an impact on measured strains (124).

Frame rate should always be reported in clinical studies.


Integrational drift

Any method that measures instantaneous values, and then integrates these measurements to an integrated value, is prone to drift. Random errors will cancel out, but non random errors will accumulate. Drift is thus a phenomenon of the modalities of displacement and strain where values are cumulated by integration of instantaneous curves. It results from the accumulation of small non random errors. In tissue Doppler, the package acquisition is a special mechanism for drift, as described below. But, drift occurs in speckle tracking also. Here it is related to the quality of tracking as discussed below. Out of plane motion and small shifts in probe position or heart position (respiration) may contribute, as may changing reflexivity of structures due to changes in fibre direction as described below.


Drift. The cumulation of non random errors by integration leads to the curve drifting from cycle to cycle. End cycle strain however, has to be zero.

If total strain at end diastole is different to zero, the heart would either diminish or increase in size. This is obviously nonsense. (Even with minimal changes this would lead to the heart either disappearing or growing to the size of the whole body in a matter of hours. The same is true of displacement. If the end diastolic displacement is different from zero, the heart would walk away from the patient in a short time.)

And even if total strain is zero at end systole, each region also have to end with zero strain and displacement at end diastole. Else, this would be equivalent to a shape change that would turn the heart inside out in a short time. Thus, end diastolic strain has to be zero, and this condition can be used to compensate drift.

Drift can thus be compensated in post processing, from the necessary condition that strain at end diastole (end cycle) has to be zero.

Either the value can be reset at the end of each cycle. This will not compensate for drift during the heart cycle, and may be seen by a distinct step in the strain curve at end diastole. A smoother correction is to apply a linear correction during each heart cycle, by calculating the slope  of the baseline from the "step" at the end of the cycle and subtracting this slope from the baseline values. This correction is based on the assumption that the drift is linear during each heart cycle, which is an approximation, but not completely true, as seen in the example above where drift is probably more sinusoid shaped.

Resetting at the end of each heart cycle gives more ugly curves, but does only correct what is known to be false. Linear resetting will result in correction also during the heart cycle, but this is presumptuous. This principle is illustrated below.


a: normal strain curve.  b: strain curve with drift, the curve does not end at zero at end diastole. c: Resetting, by simply forcing the curve back to zero at end diastole, will not change peak systolic strain, but will prevent the drift to carry over to the next heart cycle. However, if there is drift during the heart cycle, this drift affects peak systolic strain as well, as indicated in b and c. d: Linear correction can be applied as the slope from beginning to end systole can be calculated from the end systolic drift. . e: The slope values can then be subtracted from the strain values, thus forcing the curve back to a shape without drift. This will correct also values during the heart cycle, thus giving corrected value for end systolic strain. This correction, however, is presumptuous, and rests on the assumption that the drift is linear throughout the heart cycle, which is probably not the case in general.




Strain without drift compensation. It is evident in the image to the left that the strain curve  drifts from cycle to cycle. The drift may be both downwards and upwards, in this case downwards. Drift compensation by resetting the curve to zero at the beginning of each heart cycle. This is evident by the vertical line at end diastole. Comparing to the image left, peak systolic value at cycle 1 can be seen to have the same value, as no drift is compensated during the heart cycle.The resetting prevents the drift to affect later heart cycles, so cycle 2 and 3 have similar peak systolic strain as cycle 1, far lower and stable than without compensation.
Linear drift compensation is applied linearly during the whole period. This results in a further correction, as values during each heart cycle are adjusted as well. In cycle 1, the peak systolic strain is lower (in absolute value) than without or with linear compensation. In addition, cycle2 and 3 have lower peak systolic values than with resetting at end cycle, although this may not be the real case.


Thus, linear drift compensation affects peak systolic strain unpredictably, and not necessarily in the correct direction.

Method for drift compensation should always be reported in clinical studies.

Velocity gradient

The first method for measuring deformation, (apart from the wall thickening, of course), was tissue Doppler. Specifically the colour Doppler method, gave (nearly) simultaneous values across the whole sector. The frame rate was high, at the cost of low number of lines in the sector, but giving a high temporal resolution. Experiences from pulsed tissue Doppler did show this to be necessary. The simultaneity over the whole sector was the prerequisite for measuring velocity differences, i.e. velocity gradients. (Later, the tissue Doppler has also been utilised differently, in segmental strain.

The concept of velocity gradient was introduced by Fleming et al (14). The velocity gradient is defined as the slope of the linear regression of the myocardial velocities along the M-mode line across the myocardial wall.
If velocities are linearly distributed through the wall, this is equal to the difference in endocardial and epicardial velocities divided by the instantaneous wall thickness (W).

The definition was extended by Uematsu et al (15) to include the transmural velocity gradient across the parts of the wall where the scan line is not perpendicular to the wall, by the cosine correction of the velocities. The velocity gradient measured in this way, was transmural (radial). As transmural strain rate is the rate of change in wall thickness, the strain rate is the


In other words, the velocity gradient is an estimator of the transmural strain rate, strain per time unit approximates velocity per length unit. The reason this is an approximation only, is that while strain rate is defined in relation to the initial (diastolic) wall thickness, the velocity gradient is a function of the instantaneous wall thickness as shown in the formula. The velocity gradient can also be applied to the longitudinal velocities. As the apex is stationary and the mitral plane moves (fig 6), there has to be increasing velocities as well as motion from apex to base, as shown below. In other word, there is a velocity gradient from apex to base. A general definition of the velocity gradient will then be:

which in the transverse direction will be equal to the original definition. A more detailed analysis of the velocity gradient here.


Strain rate by tissue Doppler velocity gradient

Thus, longitudinal velocity gradient, is a measure of longitudinal strain rate. However, it can be shown that this is equal to the velocity gradient over a fixed distance. Strain rate by tissue Doppler measures the velocity gradient of two points over a segment with a fixed distance (In the latest scanner software, the velocity gradient is in fact calculated by linear regression of all pixel velocities within the delta x, to reduce noise.):

This is a different algorithm from the velocity gradient, but it can be proved (here) that the two formulas result in the same ratio. The distance  is called the offset distance or strain length.


a


b
a, longitudinal velocity gradient, where v1 and v2 are two different velocities measured at points 1 and 2, and L the instantaneous length of the segment between those points. The velocity gradient is given in the formula.  b, strain rate measured by tissue Doppler, as the velocities of a segment with fixed length as shown by the formula. It can be proven that these two approaches yield the same result.

Thus:        

This is explained in more detail here. The velocity gradient and SR are equal to the Eulerian strain rate, which normalises the velocity difference to the instantaneous length, while it is customary to use the Lagrangian strain, which normalises the change of length to the initial length,which is explained in more detail here.

The transmural velocity gradient can be measured by the strain rate imaging method, which is quicker than tracing the endo- and epicardial borders. Measuring the transmural systolic strain by integrating strain rate, however, is a roundabout way of arriving at the relative wall thickening, which can be measured with equal temporal resolution and much higher reliability with M-mode. In addition, anatomical M-mode can imagine wall thickening in other directions as well, although with only 2D grey scale frame rate.

Limitations of tissue Doppler

The main limitations of the tissue Doppler method are:
  • Noise, reducing radial and temporal resolution due to the need for smoothing
  • Drop outs; giving similar effects as reverberations:
  • Reverberations resulting in systematic measurement errors that may invert, reduce or increase measurement values, depending on position. In general, it may be assumed that segments more basal to reverberations will tend to increased absolute values, and thus mean values in the basal segments may be overestimated. This may be the explanation of an apparent skewness in the strain values seen in a recent population study (152) using tissue Doppler in the basal segments only.
  • Angle deviation, which is inherent in the Doppler method, but aggravated by the geometry of strain and strain rate.
  • Problems  with lateral resolution, which also are aggravated by poor alignment of the ultrasound bean and the wall.

Random noise

As can be seen from the equations given above, the strain rate is derived as a difference of two velocities. But this means that the signal (strain rate) has a noise that is the sum of the noise of the two velocities, while the signal itself is the difference between the same velocities. Thus, Strain rate has a far less favorable signal-to-noise ratio that velocities, as ca be seen from the figure below.




Velocity plot at one time during the heart cycle, taken from a 4-chamber view of a normal patient. The plot shows the distribution from the basis of the septum (left) through the apex to the basis of the lateral wall (right). It can be seen that there is a clear velocity gradient along the wall, but also that the velocities are fairly different from point to point (noise), resulting in even more noise in the derived strain rate.
Velocity curves from a normal patient. It is evident that there is some noise in the curves, but the curves can be interpreted very well.
Strain rate curves from the same dataset, showing how the data derived from the velocities multiplies the noise in the velocity data, due to the spatial derivation process.  In this image, no smoothing is applied.

The Subtraction algorithm will give substantial increase in random noise. The reason for this is the spatial derivation (11): There is a certain randomness of velocity measurement, the limit of precision of measurement. This is independent of the value measured, so lower velocities has lower signal (velocity) to noise (variability) than higher values. Strain rate, being the difference of two velocities (SR = (V1 - V2) / L) has a random variability that is the sum of the variability of the two velocities measured, or twice that of velocity.  The signal, however, is the difference between the same two velocities, resulting in a much lower signal to noise ratio, which is evident in the plots above.  there is considerable variation in velocity measurements. By calculating the mean velocity gradient along the strain length by linear regression, the noise is substantially reduced, compared to the simple subtraction algorithm (SR = (V1 - V2) / L) and regression is the present method of choice. However, random noise remains a problem, as seen in recent studies.

Random noise in tissue Doppler derived strain rate can be reduced by:

  • Increased offset distance (strain length).
  • Spatial averaging. Both points will reduce spatial resolution along the ultrasound beam (depth), and increase the susceptibility for non random noise as f.i reverberations and also for drop outs.
  • Temporal averaging (curve smoothing). Reduces temporal resolution (effective frame rate), and may lead to undersampling.
  • Integration to strain. Changes the physiological information, but the importance of this for clinical diagnosis is uncertain. However, the temporal resolution is reduced.
  • Averaging of more than one heart cycle (cine compound). But this also will give some interpolation between frames, and reduce the effective frame rate. In addition, there will be the possibility of introducing non random noise into the compound cycle. It is argued that averaging peak values may be better.
  • reduce data to parametric images. In this modality, the information is reduced to qualitative information (shortening and lengthening) or semi quantitative information (f.i. colour WMS).

Spatial smoothing






The effect of Strain length can clearly be seen. Usual default offset at present is 12 mm.
ROI size. With offset 4 mm, the effect of averaging more samples is evident.  Usual default is 12 x 6 mm at SL 12 mm.
Illustration of how both strain length and ROI size reduces spatial resolution. each strain length is represented by a point in the middle of the length. In principle the radial (depth) resolution should be ROI length + Strain length, but as shown the effect of the points outside the ROI decreases with the distance from the ROI with the regression method.

Ideally, the strain length plus ROI should be as great as possible, giving the highest possible velocity difference to ensure the best signal-to-noise ratio. As there may be little added information by entering into sub segmental resolution, the resolution may be as low as the length of one segment. This will give the segmental strain rate and strain, although measured along one ultrasound beam. However, this also increases the risk of including areas of reverberations or drop outs. This is indeed true of all methods for spatial averaging.

Temporal smoothing






Temporal smoothing at SL=4 mm, ROI size 6x6 mm. Left no smoothing, right Gaussian smoothing at 40 ms. Default at present is Gaussian 40 mm at SL 12 mm and ROI 12 x 6 mm. #Examples of recordings with the default settings can be seen above at different places.
Effect of temporal integration. Left unsmoothed strain rate, right strain from the same dataset. As the noise is random, the summation will eliminate the random variations, resulting in smooth curves.

A detailed treatment of temporal filtering can be found in (75, 77).

In the original post processing application  no smoothing was applied. That meant that the original studies were done by visually correcting for noise. Later temporal smoothing, and in the latest software (from about 2002), the Gaussian temporal smoothing is implemented. In addition, the strain rate algorithm has moved from a simple subtraction algorithm back to the original (14) regression method, resulting in a more robust strain rate estimate.

The experience from the HUNT study (153), seem to show that strain rate by tissue Doppler is a more noisy method for strain rate than the others, including segmental strain, giving higher peak strain rate (probably due to noise peaks) and wider standard deviations (variability) as shown in the comparison study. Strain did not differ, showing that the integration to strain is efficient in eliminating random noise.

Averaging more than one heart cycle (Cine compound):

Random noise will not repeat from heart cycle to heart cycle, and thus averaging more than one consecutive heart cycle, will eliminate random noise. As can be seen from the recent comparison of methods in the HUNT study (153), peak systolic strain rate by tissue Doppler is much more sensitive to noise than other methods. This can be seen from the fact that the average (absolute) peak values are much higher then other methods, while the standard deviations are correspondingly wider. Thus the peak values incorporate noise. Systolic strain, on the other hand, are quite similar, showing that the increased peak values disappears as noise is integrated sto strain as discussed above and that the higher peak values are incorporating noise peaks.

On the other hand, averaging heart cycles has several disadvantages:
  1. The frames will not be at exactly the same time point in the cycle, and thus corresponding franes will be from slightly different points in the cycle, and thus averaged. This effect is in fact similar to temporal averaging, and results in a similar reduction  in effective frame reate, and may lead to undersampling.
  2. As there is significant beat to beat variation in cycle length, later events in the heart cycle will occur at different intervals from the R-wave. Systole and diastole varies differently with respect to RR-interval, especially at HR < 100 (29), it will especially affect the time around end systole / early diastole.This may lead to bizarre results as seen below.
  3. Cine compounding by automated methods will always sample three consecutive cycles, and if one beat is of low qualituy due to motion or respiratory artefacts, or even as an extrasystole, the compound curves will include the artefact.



Native recording, showing four cycles. (Healthy child, HR around 90). It is evident that there is variation in heart rate, as it can be seen by the increasing fusion of E and A waves.
Cine compound x 2, i.e. each cycle shown in a compound of two cycles.  Systolic peak velocities does not change much, but e' wave velocities are almost halved, due to the averaging of E waves that are at different relative positions in the cycle.
Cine compound x 3, i.e. each cycle shown in a compound of three cycles.  There is not much change from cine-compound x 2, and still e' waves are very different from the native loops.



Velocity curves from the septum.  There is substatial noise as well as fusion of E and A due to beat 2 being an extrasystole at the end of cycle 2 (arrow).
Resulting in a high noise spike in the strain rate signal,
which again leads to a substantial artefact in the strain curve.



When this artefact is shown in a non-compounded image, it is limited to one cycle, and it it fairly evident that this cyscle should be discarded.
Cine compound x 2, extends the effect to two cycles
and cine-compound 3 extends the effect to three cycles, and in this image, it is not evident which cycles are representative.

Basically, If one is concerned about peak values, averaging peak values from three cycles will be more robust that automatically creating a cine compound cycle, and especially if cycles that deviate are eliminated. Thus, one is not restricted to three consecutive cycles. Cine compound may seem an attractive way of averaging three cycles, and then being able to reduce the number of measurements to one set in a cycle that is the average of three, but then the following caution should be observed:

  1. It should only be applied to systole. Also, velocity measurements are fairly smooth already, while strain rate is noisy and is the case where one profits most. Thus peak systolic strain rate maight be the one case where it it feasible and favorable. Strain curves are fairly smooth already, and cine compound will not add anything.
  2. It should only be applied where there are three consecutive cycles of good quality, but as can be seen above, the noise is not very evident in the velocity curves, and if cine compound is applied before conversion to strain /strain rate, the artefacts may be less evident.

Amount and method for temporal averaging should always be reported in clinical studies.

Drop outs

Where there are drop outs, no data can be obtained, and the resulting velocities will be zero:




Drop out in the anterior wall. In this area there are no B-mode data, indicating that there are no velocity data either (although this may not always be the case).
Schematic figure corresponding to the loop to the left. In this case there are no data in the drop out, resulting in zero velocity, i.e.  v1 = 0.
As is evident from the traces, the velocity in the drop out area (cyan curve) is zero. Below the drop out, the velocity curve is normal.



The effect on strain rate, however, may seem a little paradoxical. Below the drop out, there are much higher absolute strain rates. This can be explained by the subtraction algorithm, (although a little simplified), as SR = (V1 - V2) / L,  as V1 = zero, this will result in SR = - V2, i.e. the strain rate curve becomes an inverted velocity curve.  Thus, the absolute values will be much higher that the real strain rate.  This can be seen in the transitional zone where the strain length crosses the border between the drop out and the normal  area (cyan curve). Observe how the curve looks almost the same as the normal velocity curve above (yellow velocity curve).  The more basal measurement, with all of the strain length within normal data, shows normal values (yellow curve).  The curved M-mode shows the distribution of the effects. In the apex (1), there are no data for strain rate, showing no deformation. (the whole strain length is within the drop out.  In the midwall (2), there are exaggerated strain rate values, shown by the colour intensity (red) equivalent to the cyan curve to the left. In the base (3) there are normal strain rate values (orange) equivalent to the yellow curve to the left.


Stationary reverberations

In stationary reverberations, the mechanism is similar to the effect of drop outs. The various artifacts may arise from the effect of measuring zero velocities in some areas:








Stationary reverberations shown in B-mode, the mechanism is explained in the ultrasound section. Stationary echoes are called clutter.
The effect being that in the area of the reverberation, there are zero velocities (V2), while the velocities above (V1) and below (V3) are normal.
The result is that the reverberation shows up as a stationary area of inverted colour, showing sharply in the strain rate image. However, this is not the only effect.


Above the reverberation, where all of the strain length is in normal signals, the strain rate is normal (orange in systole, blue in diastole. In the area where the V1 end of the strain length is in the normal area, and the V2 end is in the reverberation, and thus = 0,  the strain rate will be SR = (V1 -V2) / L = V1 / L. This is seen in the yellow curves, looking like normal velocity and displacement curves. As systolic shortening is negative, while velocity and displacement are positive, there is apparent dyskinesia in this area.  This area shows inverted strain rate in the M-mode to the left.  Below the reverberation, where the apical end of the strain length is in the reverberation, and the basal end is in the normal velocity field, the effect will be SR = (V2 -V3) / L = -V3 / L as shown by the deeper colour (red) and the cyan strain rate curve, equivalent to the inverted velocity curve shown in the drop out artifact. Thus we have apparent hyperkinesia.  Finally, the red and green curves are both from the transitional zone, being interpolations between the other artifacts. The green would be apparent initial dyskinesia, while the red is apparently normal. However, as the whole area has data quality arising from artifacts, NONE of the curves should be used, lest there be biased post processing.  The whole area should be discarded.


The problem of reverberations are even greater in tissue Doppler, as this is done in fundamental, and not harmonic mode. Suppression of reverberations by harmonic imaging is not feasible due to the low frequency giving a low Nykvist limit, resulting in aliasing in tissue Doppler. Although strain rate would unwrap most of the aliasing, this would mean that separate recordings would have to be made for velocity (fundamental) and deformation (harmonic) imaging, instead of deformation being post processed from velocity recordings.


Insonation angle deviation

The main limitation of tissue Doppler due to the angle dependency is the inability to analyse anything other than longitudinal strain in all segments of the ventricle. Tissue Doppler may give transmural strain in the anterior and inferior segments (crosswise), and circumferential strain in the lateral and medial segments (tangentially), but full segmental analysis is not possible. Thus, if the other strain directions should prove to give added information, this means that at least combined methods would be preferable. However, this is still not definitely proven.

Insonation angle deviation means that the ultrasound beam deviates from the direction of the wall at the point of measurement. This is due to different mechanisms:
The wall is curved, while the ultrasound beams are straight and deviates with depths in a fan,  as shown in the figures below.
The probe may be placed off centre, although this may not be apparent in the imaging plane, only in the orthogonal plane (76). Off centre placement of the probe, however, may both increase angular error (76) in the segments with best alignment, or decrease it due to curvature compensation (42).

The angle dependency is mainly a problem in tissue Doppler, but reduced lateral resolution in B-mode in order to achieve a high frame rate (or indeed, 3D images), may increase angle dependency in speckle tracking by reducing the ability to track in the lateral direction.

It is well known that velocity measurement is dependent on the angle between the ultrasound beam and the velocity direction (vector) – insonation angle, and that the measured velocity vm is reduced in proportion to the cosine of the angle a between the velocity vector and the ultrasound beam as discussed in the ultrasound and mathematics sections.



For strain rate and strain measurement by velocity gradient, the angle dependency is somewhat more complex. In the longitudinal direction, the longitudinal velocity gradient is reduced by the cosine of the insonation angle, as for velocities. But in an incompressible object, there is simultaneous strain in the transverse direction, in order to keep the volume constant, and the two strain components are opposite and will detract from each other (1, 2, 7). This results in further reduction in the measured strain and strain rate (7).  This is illustrated below.





The true velocity vector is shown as the straight arrow, the ultrasound beam as the dotted line. The vector measured along the ultrasound beam is reduced by the cosine of the angle between the true vector and the ultrasound beam. An object undergoing longitudinal shortening (negative strain). The ultrasound beam (dotted line crosses the longitudinal direction at an angle. In addition to the apparent reduction of the shortening by the angle, there is simultaneous thickening in the transverse direction (positive strain), which further detracts from the n numerical value of the shortening.
It's important to realise that this double angle problem is limited to the velocity gradient method. Tissue Doppler is still angle dependent, but segmental strain by tissue Doppler has only the basic limitation common to all Doppler measurement. And using it in combination with transverse speckle tracking, eliminates this angle dependency also, to the same degree as in speckle tracking (meaning that speckle tracking is not altogether angle independent).

Angle deviation is most usual in the apex and base:









Angle deviation is biggest in the parts that are most perpendicular to the ultrasound beams, as shown here, in the apex and the base of the sigmoid septum.
The sigmoid septum illustrated here, showing systolic  positive  strain (blue - apparent lengthening) due to the angle being crosswise to the wall, in reality measuring thickening
Another patient with no sigmoid septum shows no artifact in the base of the septum. In addition, as the ventricle is not dilated, the area with apparent lengthening (blue) is fairly small.
Strain rate curves from the same patient showing it top be perfectly feasible to measure strain in the basal part of the apical segment (red ROI and curve), the area with inverted values (being in reality thickening) is fairly small (yellow ROI and curve).



This is described in a more detailed mathematical analysis in the mathematics section.



Which segment that is most in alignment with the ultrasound beam, may vary with LV shape and depth, as shown here.

It has been maintained that strain and strain rate cannot be measured in the apical segments, but from the illustrations shown above, this is not necessarily true for all patients on a segmental level, and the apical segments may in fact be the ones showing best alignment in some cases.


It's important to realise that this double angle problem is limited to the velocity gradient method. Tissue Doppler is still angle dependent, but segmental strain by tissue Doppler has only the basic limitation common to all Doppler measurement. And using it in combination with transverse speckle tracking, eliminates this angle dependency also, to the same degree as in speckle tracking (meaning that speckle tracking is not altogether angle independent).


However, the apex seems most susceptible in the HUNT study, but the problem was solved by the ROI tracking the myocardial motion.


In post processing, the main point is to exclude segments with to great angle deviation from analysis, at least other than parametric.

In some instances the angle problem is due to imperfect alignment (foreshortening), if the probe is not positioned properly over the apex. (As indeed may be necessary to obtain an acceptable window). In that case, the angle problem can vary along the wall as shown below:


Less than perfect alignment with the apex, results in a, angle along the inferior wall in this 2-chamber view. It can be seen that the wall apparently is curved, and that the alignment is better in the basal than the apical half.

This has different effect in the different parts. Basally, there is a normal strain rate curve (yellow). Apically, the systolic strain rate is reduced to half, due to angle distortion (red). In the midwall, there is a normal peak value, but the systolic curve is cut off, resulting in zero values in the late systole (cyan). as the bent area moves into the ROI.

This is also apparent in the curved M-mode, showing an area of apparent a- to dyskinesia in late systole in the midwall.

Using timing information, especially the shifts between positive and negative strain rate, on the other hand has been proposed to overcome the angle limitation. This, however, may be problematic if the alignment is less than perfect, in a way tat the angle between the ultrasound beam and the wall varies through the heart cycle, as shown in the midwall segment in the middle image above.


Variable insonation angle during the heart cycle

In individual cases, there may also be angle problems in other levels, especially the inferior wall in the care of foreshortening. The angle deviation may apparently vary during the heart cycle.


        
Two chamber view. The apex is in fact outside the sector to the left, and the inferior wall appears to have a break in the midwall. This part is transverse to the ultrasound beam, and here the measured strain rate will be wall thickening (positive strain rate), as shown in the diagram, longitudinal shortening (negative strain rate; orange arrows) in the apex and base, transverste thickening in the midwall (positive strain rate; cyan arrow).

This has a similar effect as in above, but in this case the alignment is better both in the apex and the base. In the midwall, there is still normal peak strain rate, but the systolic curve is cut off.

In the curved M-mode, the area of positive strain (transverse thickening is seen to move with the wall. The pattern may resemble a reverberation, but doesn't last throughout the heart cycle, and the time course is not horizontal.

The motion of the distortion area suggests how to deal with this artefact by making the ROI track the myocardial motion, as shown here.

Tracking the ROI:

Commercial software have the option of tracking the ROI manually, but the tracking could be done by automatic methods as in the NTNU application. The tracking eliminated the systematic angle problem in the apex in the HUNT study.




The same loop as above, showing normal strain rate curve in the base (yellow), but abbreviated systolic strain rate curve in the midwall, as the area of transverse strain moves into the  ROI.

The same ROI placement in start systole as left, but now the ROIs are made to track the myocardial motion  through the systole. Thus, the midwall curve improves, showing normal strain rate through systole, demonstrating the the finding in a is an artefact. It also demonstrates that tracking makes little difference in a normal strain rate curve (yellow), except maybe in the apex.

Thus, as opposed to stationary artefacts, tracking may help to keep the ROI outside the path of moving artefacts, but of course if the ROI is trackiong into a reverberation, the results will be worse



Normal strain curve below the reverberation. The ROI is stationary in space.

Same ROI as left, but the ROI made to track the myocardial motion, passing through the reverberation during systole, and strain rate curve can be seen to be cut out and inverted in that period.

Thus the value of tracking the region of interest depends on the quality of the data.
 



  Lateral resolution

Another problem, being related to the insonation angle, but with mechanism similar to the drop out and reverberation issues, is arising from the low lateral resolution of tissue Doppler. This is due to low line density which is applied in order to achieve a high frame rate, as discussed in the ultrasound section. If the frame rate is around 150 FPS, there is usually less than 20 lines in a sector. This is even more enhanced by using the MLA technique where a broad transmit beam  is used, and the signal is received along more, narrower receive beams. The simultaneity of the parallel receive beams results in signals partially being received by the neighboring beam. In addition, as tissue Doppler is acquired in the fundamental mode as discussed above, side lobes are more prominent, contributing to reduced lateral resolution.








Relation between frame rate and lateral resolution in tissue Doppler.  The numbers are receive lines, this means that in 4MLA, the number of transmit beams are one fourth. In reality, the lines have the same width, the data are interpolated between the lines.  (Image courtesy of E Sagberg.)
Example of how this affects velocity measurements, In this image, taken at 150 FPS, the four sample volumes placed side by side in the base of the lateral wall, generates exactly the same velocity curve, showing that the data are the same.


Initially the default frame rate was 150 FPS in tissue Doppler, but now the default is intermediate (about 100). The low lateral resolution may have effects similar to the effects of drop outs and reverberations, as one of the velocities of the strain length may be from cavity or pericardium as illustrated below.




Two wide ultrasound beams are shown in grey, with the middle marked in red. The velocities that are recorded within the beam are transposed to the middle line.  Strain rate is analysed along the middle line. In the apex the angle makes the beam miss v1, and may instead use pericardial velocity (red circle) in strain rate analysis, thus v1 - v2 will be equal to -v2, too high value.  In the midwall, missing v2 and using pericardial velocity instead, v1 - v2 will  be equal to v1, a velocity curve, inverted and too high numerical values.  Here, this is shown in the midwall, but the base is even more prone to this artifact. In the base is shown another artifact, the beam misses v1, and uses velocities from the cavity instead, which are zero, being removed by the low pass filter. v1 - v2 will then again be equal to -v2, accentuated numerical values, being an inverted velocity curve.
Normal myocardial strain rate curves (orange, cyan and green ROI and curve), show a fairly even distribution of strain rate. In the lateral wall  a sample volume too far into the cavity (cyan ROI and curve) will give high numerical strain rate values, for the reason shown in a, basal. A curve too near the pericardium (red ROI and curve) shows reduced values, due to a partial effect of the pericardial influence on v2, as illustrated to the left; midwall. Averaging makes this effect partial, pericardial velocity only detracting from the numerical strain rate values. Pericardial artifact  as illustrated to the left in the midwall. Top, systole, bottom, diastole. Here, near the base is seen high, inverted strain rate values.

The effect may explain why some authors have found higher strain rate values in the base than in the rest of the ventricle in normals, which is an absolute artifact. It may also explain why some authors have found differential strain rate values across the wall, which is fairly improbable as discussed above.


Integrational drift in tissue Doppler:



In tissue Doppler, the package acquisition of colour Doppler, results in drift from one package to the next, as shown below.

Integrational drift in package acquisition. The curve is a displacement curve. Ultrasound pulses are shot in packages of at least two pulses, and both pulses contributes to the velocity estimate, as described in the basic ultrasound section on colour Doppler. Thus the velocity is an estimate for the package with a duration of 1/PRF, and the velocity of motion is extrapolated for the interval between packages, i.e. 1/ FR. From the figure, is is evident that integration of these velocity samples will deviate from the true values in between packages: In this case there is negative drift in systole, positive in diastole, but the end result is unpredictable. The repeating pattern, however, will ensure that there may be a net positive or negative drift from heart cycle to heart cycle, which is more or less linear. The drift within  each heart cycle, however, is not linear.

This effect is not only due to the mathemathical integration of displacement from velocity, it is actually also a property of tracking by tissue Doppler, as the position of a kernel is calculated from the position and velocity in the previous frame. (this is in reality the same thing). A frame rate of 1 KHz would to some degree eliminate this. Then each velocity could be calculated by single acquisition i.e. from one frame to the next, without the gaps due to FR being much lower than PRF. This has been shown to be technically feasible (215).

Segmental strain and strain rate.

Segmental strain and strain rate is taken to mean the deformation measured over a complete segment, giving the average value for the segment.


Tracking of segmental borders. The strain is the relative change in segment length, and the strain rate the strain change per time. Segmental strain in six segment in four chamber view.  Schematically , the motion of the segmental borders is shown by the arrows, decreasing from base to apex. The segmental strain is the difference between the motion of the two ends of a segment. The curves are from a real example. The blue strain curves show segmental values, while the green curves are the average of the wall for comparison.Image courtesy of H Dahlen.

Segmental strain has several advantages:
  1. As measurements are fairly noisy, the average of a whole segment will tend to be more robust. This will give a high signal-to-noise ratio as discussed above. The segmental strain is equivalent to a strain length equal to the segment length, i.e. about 3 cm. The segments are the basic unit for evaluating regional wall motion score (WMS) in the recommendations of the ASE/EAE (146), and so far the clinical usefulness of a higher resolution has not been demonstrated.
  2. Tissue Doppler measures the velocity gradient along the ultrasound beam, not along the segment. Increasing the strain length will reduce noise, but the strain length will follow the direction of the ultrasound beam, and this will give problems where the alignment is not perfect, as discussed in the pitfalls chapter, under the discussion on insonation angle and  lateral resolution. Tracking the end of each segment, ensures a better measurement of the segmental longitudinal shortening.This will make the method less angle sensitive, as well as more similar to the other methods as the measurements are related to kernels at the segmental borders.
  3. As long as segmental length is followed by tracking the ends of the segment, the value will be little affected by smaller artefact's within the segment as illustrated below.

Limitations of segmental strain

The disadvantages may be that
  1. The method will be measuring only two points along the line and thus
    1. be less robust than the full segmental average (provided the data in the segment are good).
    2. be extremely sensitive to artifacts at the points of tracking as seen below.
  2. The low resolution in the radial (longitudinal) direction. Sub segmental values cannot be extracted (although they could be interpolated).
  3. If the tracking at one segmental border is poor, it will affect two segments on both sides of the border as discussed below. 
If the algorithm does not track one kernel correctly, the strain values will be wrong for the segments on both sides of the kernel. This is evident in areas of drop outs or reverberations as illustrated schematically below.


Effect of a reverberation on the border between the apical lateral and the midwall lateral segment. A kernel  in this area will not track, as illustrated by the arrow.  The next border between the basal and midwall segment moves normally, leading to an exaggerated shortening of the midwall segment, while the basal segment shortens normally. (The segmental strain in the apex is the difference between the apical motion (zero) and  the apparent motion ( near zero) in the reverberation, the midwall strain is the difference between the apparent motion ( near zero) in the reverberation and the (normal) motion of the border below.) This is evident by the curves (compare to the average curve: The apical curve shows little strain, the midwall curve shows far more than the wall average, and the basal shows average strain).  Compare with the image above. Image courtesy of H Dahlen.
The kernel is in a reverberation in the lateral wall, and will not track. In this example the reverberation is in the border between the basal and midwall segment. the shortening of the basal segment is exaggerated, and the shortening of the midwall segment is reduced.

Thus, one kernel tracking poorly will then lead to two segments being discarded, giving a high discard percentage. This was seen in the HUNT study with automated analysis, we consider this an advantage of the study, leading to little contamination of the data by artifacts, thus ensuring the data to be "clean". However, it is a disadvantage of the method, leading to a lower feasibility. However, in clinical studies, the feasibility was around 80%, and in addition showin added diagnostic value to B-mode (128). But basically a high discard rate ensures higher quality of the studies.

Segmental strain by tissue Doppler

Instead of calculating strain rate along one ultrasound beam, it has been proposed (44) to calculate strain rate from the velocity differences at the segmental interfaces and segment length along the wall. (Any points will do actually), as shown below. This ensures that tracking is done at the real segmental borders, and makes the method less dependent on the alignment of the ultrasound beam and the myocardial wall. The method is still angle dependent, but this is the ordinary angle dependency of Doppler, not the angle dependency of strain and strain rate. The method has not been implemented in clinical use.

Segmental strain rate, measured by tissue Doppler, but by segmental velocities that do not lie on one ultrasound beam, while strain length is measured along the wall, between the velocity points. 

Tracking is possible, by calculating the displacement from one frame to the next, along each ultrasound line, and thus, the segment length.

This was a theoretical approach in the time when only tissue Doppler was available for tracking. The advantages of segmental strain are present also when using kernel tracking with kernels at the segment borders using either speckle tracking or combined tissue Doppler and speckle tracking. Both those methods are in addition angle independent, as the segment orientation follows the myocardium, and the strain is simply calculated along the length of the segment as it is.


Speckle tracking in grey scale images.


Diving humpback whale. Each humpback has an unique (speckle) pattern on the underside of the tail (and flukes). Thus each individual can be identified by its speckle pattern. Photographs at different times and places can thus track the wandering of each individual all over the area it wanders, without recourse to anything else than the pattern.  - Speckle tracking! (In grey whale images). This is thus a method with low frame rate, giving mainly the extent of wandering over a long time period (the sampling intwerval).

The basic principle of speckle tracking is based on the interference of the reflected ultrasound giving rise to an irregular - random - speckled pattern. The random distribution of the speckles ensures that each region of the myocardium has an unique pattern, a fingerprint, just as in the whales.  The speckles follow the motion of the myocardium so when the myocardium moves from one frame to the next, the position of this fingerprint will shift slightly, remaining fairly constant. Thus, if a region (kernel) is defined in one frame, a search algorithm will be able to recognise the lie sized and -shaped area with the most similar speckle pattern in the next frame, within a defined search area (fig. 18c), and hence, to find the new position of the kernel (26). This has been shown to be feasible in flow (94) and strain rate imaging (95).

The basics of speckle formation and speckle tracking is given in more details here.






Typical speckle patterns in the myocardium, demonstrating the differences between different areas (kernels).  The difference in the pattern is the basis for speckle tracking.
Speckle tracking. the speckle pattern from the first frame (red) has moved to a new position (green) in the next,
and can be recognised by a search within the search area.





Kernel displacement.  Following the kernel through a whole heart cycle, will lead to a displacement curve shown to the right. Temporal derivation (displacement per time, or frame by frame displacement divided by the time between frames) results in the derived velocity curve shown below. From two different kernels, the relative displacement and hence, strain as well as strain rate can be derived. The strain obtained by simply subtracting the two displacements and dividing by the end diastolic distance is the Lagrangian strain. To obtain the Eulerian strain rate, the correction has to be applied for each frame.
If Kernels are placed at the segmental borders, the result will be segmental strain and strain rate in six segments per plane.


The advantage of this method is that it tracks in two dimensions, along the direction of the wall, not along the ultrasound beam, and thus is considered angle independent.  


Longitudinal speckle tracking, with kernels at the segmental borders in four chamber view.

Longitudinal speckle tracking, but done crosswise in parasternal long axix view.

In principle, pure speckle tracking  is direction independent, and can track crosswise. This means true longitudinal strain, as the length will follow the "tilting" of the segment as well as the shortening as seen from the example above. In addition, the B-mode has a far better lateral resolution than tissue Doppler.

However, the speckle pattern will not repeat perfectly. This is due to both true out of plane motion (rotation and torsion relative to apical planes and longitudinal deformation relative to short axis planes) and to small changes in the interference pattern. But the frame to frame change is small, and the approach to recognition is statistical, the basic algorithms are shown here. The greater the change from frame to frame is, the poorer tracking. This may be especially a concern in higher heart rates (f.i. in stress echo), where the number of frames per heart cycle decreases. This could be compensated by increasing frame rate, but at the cost of lower line density and lateral resolution. Lateral resolution is important in delineating the speckles in the lateral direction. If the lateral resolution is low,  the interpolation will result in a "smeared" picture as shown here, with speckles that are nor so easily tracked in the lateral direction. In addition the lateral resolution decreases in depth with sector probes, making lateral tracking at greater depths doubtful. The poorer the lateral resolution, the poorer tracking in the lateral direction, and the more angle dependent the method becomes.

Thus, as the lateral resolution is far less than the axial, the two directions are not equal. This means that tracking in the longitudinal direction is better than the lateral, so the method is angle dependent to some degree. And increasing the frame rate (for instance to compensate for high heart rate) reduces the lateral resolution even more, reducing the angle independence of the speckle tracking method even further.

Speckle tracking has some fundamental limitations as well, partially the same, and partially more or less complementary to the Tissue Doppler limitations.

Fundamental imitations of speckle tracing

The speckle tracking method has fundamental limitations, just as tissue Doppler

  • Low frame rate, which may lead to undersampling. Shorter events like the isovolumic phases may disappear all together, and peak values may be reduced due to under sampling, especially diastolic velocities and strain rate. This limitation is most important for measuring peak values in diastole, not so much in systolic strain rate, and systolic strain has least frame rate sensitivity of all.
  • Poorer tracking at higher heart rates may render the method less useful in f.i. stress echo.
  • Increased angle dependency with increased frame rate due to reduced lateral resolution, and also with increasing depth in the sector due to decreasing line density with depth. In Speckle tracking, this reduces the ability to track in the lateral direction. The problem is less than in tissue Doppler, but "angle independence" is an oversimplification.
  • Reverberations is just as great a problem in speckle tracking. where the echo doesn't move, the application doesn't track
  • Drop outs is just as great problem in speckle tracking, where there is no B-mode data, the application doesn't track.
  • Drift is also a problem, due to imperfect tracking, as shown below.




Drop out affecting speckle tracking. The application cannot track where there are no tissue data, in this case in the anterior wall and the application doesn't track (the markings don't move). The inferior wall seems to track normally.
In this case, the segments that do not track, are excluded by both automated software (2D strain ) and visually, thus giving no colour in the M-mode and no curves.
Reverberation in the lateral wall affecting speckle tracking. AS is visually evident, the application does not track across the reverberation, thus the two segments apical to the reverberations are seen as akinetic, the basal as hyperkinetic. All shortening is seen in the basal segment.

Drift in speckle tracking - kernel slippage

If the tracking is less than perfect, the kernel will slip slightly from one frame to the next; "kernel slippage". Out of plane motion and small shifts in probe position or heart position (respiration) may contribute, as may changing reflexivity of structures due to changes in fibre direction. Then the tracking will start at the new, slightly off position in the next frame. This effect will tend to be non random, as the imperfect tracking usually will result in the kernel moving a shorter distance than the tissue, and with further slippage, there may be a cumulated drift during the cycle as illustrated below:


Kernel slippage in speckle tracking. The kernel is defined in frame 1, indicated by the red rectangle. In the next frame, due to out of plane motion, or simply changes in reflectivity some of the speckles disappear or have lower intensity in the next frame due to complete or partial out of plane motion in the B-mode image.  Then the kernel may find a slightly different area as the new kernel position. (Especially if the tracking is done by the sum of absolute differences where the identification rests with the summed intensity within the kernel area). In frame 2, the true kernel motion is identified by the dark grey rectangle, the tracking, however, identifies the new position as the red rectangle. Some of the speckles above the kernel have decreased in intensity, while the speckles below have all increased. In frame 3, further changes in speckle visibility results in further  slippage, i.e. slippage in relation to frame 2, which then is a larger cumuated slippage from frame 1.  Two speckles from frame 2 above the kernel have disappeared, four speckles have decreased in intensity. Two speckles below the kernel have increased. The true position of the kernel from frame 1 is indicated by the light grey rectangle, the position of the red kernel from frame 2 by the dark grey rectangle, and the tracking by the red rectangle.




This is a fundamental property of speckle tracking,  and the drift from start of cycle to end of cycle may actually be used as a criterion for quality of speckle tracking. And even more advanced comparing tracking forwards and backwards through the whole cycle , f.i. by cross correlation. It may be less with a higher frame rate. (Although that will lead to more angle dependency). If the speckle tracking is used for calculating a velocity field as the primary variable, as in 2D strain, the integration to displacement an strain will result in further drift by cumulating small errors. In addition undersampling is a property of low frame rate, i.e. B-mode. This reduces peak velocity, and the peak values is even more reduced if smoothing is applied before integration as it is in 2D strain. (In tissue Doppler, smoothing is applied to velocity and strain rate, but the integration is done on raw, unsmoothed data, as can be seen from this example.)

Segmental strain, measuring the displacement of kernels directly, and calculating the strain from segment length, if free from the drift due to mathemathical integration of velocity / strain rate, as here displacement and strain is measured directly. However, the drift by kernel slippage from imperfect tracking in speckle tracking, or from package acquisition in tissue Doppler is still present.



Quality evaluation of speckle tracking

Basically, the quality of the speckle tracking can be assessed visually, by evaluating how well the kernels follow the tissue motion. This is facilitated by slowing the replay, or stopping and stepping the frames (a method that is strongly recommended in stress echo evaluation as well). In the basic application, this is feasible, as each kernel tracks independently. (This is not always the case in more advanced commercial applications, as will be discussed later.) Independent tracking also gives the possibility to replace a kernel that is not optimally placed:


The kernel is in a reverberation in the lateral wall, and will not track, thus both the segment below and above the reverberation will show artefacts.
Adjusting the position of the kernel manually, allows speckle tracking despite the reverberation, if the kernel remains outside the reverberation during the whole heart cycle.

Thus the position of the kernels can be adjusted. If this is not feasible however, the segments can be discarded from analysis. Additional software may give automated quality control. The method used in the HUNT study checks quality by tracking forwards and backwards, comparing the tracing both ways by cross correlation (151), or by the difference between backward tracking and start of forward tracking (127, 128, 151). It still remains unsolved which method (including visual assessment) is best.

Segmental strain by speckle tracking

This method, with kernels placed as shown, will result in measurement of segmental strain and strain rate, segment length being defined as a straight line. This means that there will be less angle dependency than in segmental strain rate by tissue Doppler, and thus the tracking may be better than the above method for segmental strain.

The advantage is the same as in segmental strain generally, being little angle dependent, robust against noise and little affected by sub segmental artifacts. And kernels may be replaced to avoid areas of reverberations or drop outs, if possible.

The fundamental limitations of speckle tracking, however, apply.

Segmental strain by combined use of tissue Doppler and Speckle tracking.

Modern ultrasound equipment has the capability of acquiring second harmonic grey scale images with an acceptable frame rate of 40 - 50 FPS and good lateral resolution, simultaneously with tissue Doppler data. This opens the possibility of tracking along the ultrasound beam by tissue Doppler,  while tracking transverse to the ultrasound beam by speckle tracking (124) in the grey scale data. The fundamental advantages as well as limitations of segmental strain remains.



Combined search by tissue Doppler and speckle tracking. The kernels are shown as the small, round, yellow circles.  The longitudinal search area along the ultrasound beam by tissue Doppler is shown in red. The lateral search area by speckle tracking is shown in white. The result is tracking of segmental borders. The strain is the relative change in segment length, and the strain rate the strain change per time.
Display of segmental strain rate from the six segments.

In addition, the combined method has additional advantages:
  1. As longitudinal strain is the main issue so far, the advantage specifically of the combined method is that longitudinal tracking is done with the high frame rate of tissue Doppler. This may give an additional benefit, at least in strain rate.
  2. Doing only transverse search by speckle tracking simplifies the search algorithm, limiting the search area to a sector extending in the radial direction and thus reducing the time for the speckle search.
  3. The transverse tracking by speckle tracking eliminates the problems of both insonation angle and low lateral resolution of tissue Doppler alone. Thus, the method is just as angle independent as speckle tracking, as opposed to the segmental strain by tissue Doppler alone.
  4. If the tracking of tissue (lateral tracking) is poor, the measurements will be similar to the segmental strain by tissue Doppler, meaning that it will be somewhet angle dependent, but still less than the velocity grasdient method.
  5. Finally, it utilises the full dataset inherent in the combined image.
The combined method can be used in different ways to analyse strain rate imaging (127):
  1. Segmental strain by speckle tracking alone
  2. Segmental strain by Combined tissue Doppler and speckle tracking
  3. Strain rate by longitudinal velocity gradient, by placing an ROI and strain length in mid segment (in end diastole)
    1. Letting the ROI remain stationary or
    2. Letting the ROI follow the segment being tracked by the combined method
3b is similar to ordinary strain rate by tissue Doppler, 3a is improved, as the present applications only offer the manual tracking as a possibility, while the combined method gives the strain rate. The tracking has been shown to be advantageous in the apical segments, in a comparative study.

This method  has already been shown clinically useful in stress echo (128), giving a sensitivity of peak systolic strain rate for ischemia of 84% and an AUC of 0.9, compared to coronary angiography, and with a feasibility at peak stress of 80% of segments. It is also the method used for the HUNT study (153), for automated analysis. In this study methods 2, 3a and 3b was compared as well, and compared to 2D strain, comparisons shown below, showing little differences between mean values and normal variations.

The limitations of the combined method is the same as for echo in general issues related to image quality, and the general limitations of segmental strain.


It has been considered a problem that the combined method is not commercially available, at present it is a research tool at the Norwegian University of Science and technology. It has been criticized by some, that the results are not useful, as they are not transferable. I consider this criticism only partially valid.
  1. Firstly, this is a research tool so far for doing research into and comparing speckle tracking and tissue Doppler. And the speckle tracking is "pure" speckle tracking, not advanced applications utilising a lot of computing to achieve results. The application makes it possible to compare tissue Doppler, speckle tracking and the combination directly (151), and as shown below.
  2. The HUNT substudy showed little bias between methods.
So far, the two data sets could be combined in different ways, not limited to the application described above. Further research should be undertaken to assess this.

It seems rather absurd in the long run, that having access to high quality grey scale tissue data as well as high frame rate tissue Doppler data in the same loop, the quality of measurements will improve by discarding one of the data sets.



2D Strain (AFI) by speckle tracking.

The application known as 2D strain or AFI (automated functional imaging) by GE Vingmed, is generally seen as a speckle tracking method, which actually is  the basic  method. Tracking is done by the same method, (sum of absolute differences). However, the method also has implemented:

User friendly interface (if strain rate imaging by tissue Doppler had had the same interface from the start, it would probably have gained more acceptance). Some may call it seductively user friendly.
A high degree of smoothing, making the results very robust (but may have disadvantages)
Method specific processing that imposes specific limitations.

With a greater number of kernels, distributed both along and across the wall, each kernel can be tracked individually, and displacement and velocity can be measured in two dimensions, both longitudinally and transversally for each (73).


 From this, differential motion - i.e. deformation - can in principle be measured, both in the longitudinal and transverse direction. The smaller the kernel, the less certain will the tracking be, but this can be compensated by selection of kernels on the basis of a stable pattern from one frame to next.  One method of insuring stable tracking is to discard kernels that are not present in a sufficient number of frames. In the same way, kernels that does not move can be discarded, reducing the influence of reverberations.However, the dependence on recognising stable kernels from one frame to next, makes the method even more frame rate sensitive.

2-dimensional strain by speckle tracking.  Each red point represents a kernel for speckle tracking. Velocity and displacement decreases from base to apex, and the differential motion along the segment gives longitudinal strain and strain rate. As the true direction of the motion is tracked in this instance, the transverse component can also be tracked, and the differential motion from kepi- to endocardium can also be tracked., giving transmural strain and strain rate.

2D strain in practice. The midwall line is used for the longitudinal strain, being an average of all points in the wall. The ROI follows the wall, the limits can be seen diverging in systole, converging i n diastole, giving the transmural strain and strain rate at the same time. The colours show longitudinal strain rate, green is shortening and red is lengthening.

In order to make the speckle tracking more robust, values are averaged over a whole segment.

However, this is one of the main limitations of the data in the bottom: The smaller the kernels, the greater the uncertainty of position, and the more noise. Thus, a liberal amount of smoothing has to be done. Averaging a large number of kernels may make tracking more robust, although this reduces the number of useful speckles in each kernel. This can be done in various ways and combinations. With more than one layer of kernels across the wall, the longitudinal measurements can be averaged from all layers, giving a transmural average. Longitudinal averaging can be done along one segment, giving the segmental average. This can also be done in a more sophisticated way, by spatial interpolation along the wall. This will result in a gradual effect of spatial smoothing, although the extent of the smoothing is less easily discerned. Thus, the tracking of a region of the ROI, is not speckle tracking alone, but also extrapolation of the AV-plane motion i.e. virtual tracking. This is discussed below.


The method starts with generating a velocity field across the sector (velocity being the displacement between one frame and the next, / 1/FR), more or less as in tissue Doppler, but the velocity vectors being angle independent. Thus the displacement ad strain rate and strain can be derived in the same way.

Longitudinal velocity
Longitudinal displacement
Longitudinal strain rate
Longitudinal strain


Thus, it is evident that both tissue Doppler and 2D strain will give four modalities from one dataset:


2D strain derived displacement, velocity, strain rate and strain. Note the smoothed curves.
Tissue Doppler derived velocity, displacement, strain rate and strain. Note that these are unsmoothed curves.



Quality evaluation of tracking in 2D strain

The first step in processing will be the quality control of the tracking, as described above. The application has built in automated assessment of the quality of tracking, with methods similar to the ones above. Thus the application will suggest acceptance or rejection of the segments in the view. However, the segmental quality assessment is compared to the average, there are no absolute criteria, so tracking should still be evaluated visually, and segments that do not rack, should be excluded manually.




Visualisation of the tracking. Observe how the bullets in the midline follows the myocardial motion. However, due to the smoothing function of the application, this may be virtual tracking, being extrapolated from AV-plane motion, thus the true tracking of the local tissue may be difficult to assess.
Loop from another patient. Tracking seems fair, visually.
Automated quality assessment accepting the tracking results from all segments. The assessment is presented as a suggestion, and manual acceptance is required, i.e. the accept button has to be pushed in order to process.



In this case, the application suggested to accept all segments, which would result in the values as shown.  The poor tracking in the apical lateral segment does not reflect in the values, due to the spatial spline smoothing in the application.
The same loop as left, but manually override of the automated quality assessment, excluding two segments before pushing the accept button, lateral basal as it exaggerated the tracking by tracking side lobes, and the apical lateral, as the tracking was poor due to near field reverberations.


We did an initial evaluation of an earlier version of this application in February 2004, comparing the longitudinal motion and deformation measurements by this application with those obtained by tissue Doppler, in separate images. The study consisted o0f 20 patients with a wide range of function.



Strain rate and strain, comparison of 2D strain and Tissue Doppler. There is a considerable spread between methods, but most probable due to variability of especially of tissue Doppler. There 2D strain gives lower values than DTI, and this tendency increases with increasing strain rate/strain. The term "CEB" meaning "computerized eye balling" was an early term to describe the application.

When measurements was sorted in quartiles, Concordance was only between 27 and 34%. Feasibility was the same with 2D strain and TVI. Further investigation was not undertaken at that time, as the application was modified in later versions. Other authors have found a much better correspondence between TDI and 2D strain (73), with correlations of 0.94 and 0.96 for strain rate and strain, respectively. However, as seen by the curves in the figure below, both data sets are analysed by the 2Dstrain software, and thus subject to the same high degree of smoothing, so the results do not reflect independent analysis.


From a validation study where tissue Doppler and 2D strain derived strain rate (left) and strain (right) values were compared. However, as can be seen
from these curves, both curves are very smoothed and concordant. Thus, much of the concordance must be assumed to be due to smoothing, as both
methods were processed by the 2D strain software, and not by independent analysis software. Adapted from Modesto 2006 (73).


This is illustrated below:

The lines looking smoother, is a function of the averaging function used in the algorithm, the application will do the same to tissue Doppler data.


Strain rate curves from speckle tracking and tissue Doppler from the same cine - loop.  The same smoothing is applied to both, showing that smoothing of the curves is not the result of the robustness of the algorithm, but of specific temporal and spatial smoothing applied by the application. The curves differ somewhat (but not too much), as strain rate is calculated with different angle and lateral resolution.

Another study by Cho et al (148) finds only correlations of longitudinal strain by 2DS and TVI with MR tagging of 0.51 and 0.40, respectively. This may reflect the real precision of both methods (and of MR tagging as well?) but then the correlation between the methods cannot be higher.

Limitations of 2D strain

The main limitations of 2D strain are the general ones, as well as the ones specific to speckle tracking.
In addition, 2D strain has method specific limitations related to:
  • Curvature dependency, due to the technicalities of the specific applications, which may give too high values in the apex.
  • Smoothing, relying heavily on AV-plane motion,
    • which may give strain values even where there are none, and may reduce sensitivity for reduced regional function
    • Makes the tracking more difficult to assess visually

Reverberations and drop outs are illustrated above. Those examples are all taken from 2D strain, but shows the general principle.

Curvature dependency of 2D strain

The longitudinal values that are obtained by the 2D strain application are curvature dependent, as shown below:


Curvature dependency of strain measurement. If the ROI is curved, the midwall line will move inwards, and thus shorten, even if there is no shortening of the segment. This will result in an apparent shortening of the segment itself, adding to the real longitudinal shortening. This curvature effect is dependent on the curvature, the width and the widening of the ROI.
Curvature dependency of strain in 2D strain by speckle tracking. The two images are processed from the same loop, to the right, care was taken to straighten out the ROI before processing, while the left was using the default ROI. In both analyses the application accepted all segments. It can be seen that the apical strain values are far higher in the right than in the left image (27 and 21% vs 19 and 17%).  However, the curvature of the ROI even affects the global strain, as also discussed above in the basic section.

The width and thickness as well as curvature of the ROI is non standard, defined ad hoc. In addition, the ROI width is uniform from base to apex, while the myocardium is thinner in the apex, giving a discrepancy between ROI width and wall thickness. As the curvature effect is also a function of ROI width, this may add to the curvature effect. This effect may account for the observed base-to-apex gradient of strain values observed in some studies. The combined method (and indeed tracking of segment length by speckle tracing alone without TDI by the same application) is curvature independent as shown below.  It may be the reason why some authors find a base-to-apex gradient in the strain values obtained by this application, while we did not in the HUNT study.


Another instance of the curvature dependency of measured values. The left image is processed with fairly straight ROI in the apex. The middle image is the same loop processed with more curved ROI, in both cases the application suggested acceptance of the tracking in all segments.  AS opposed to the above example, in this case the global strain is severely affected by the ROI shape as well (.15.7% vs - 20%). To the right is shown another loop from the same patient, centered on the right ventricle. In this case, the values of the septum is quite similar to the values in the left image, but differs from the middle image. The interesting thing is that the global strain itself is different from the mean strain calculated from the segmental values.

This may affect the regional strain as well, as the curvature dependency may assign higher values to akinetic segments as shown below:


Small apical infarct. Admitted with a history of pain, but free of pain and with normal ECG, but elevated Troponin (analysis results not ready till he had new pain) at the time of admittance.  This Echo at admittance was initially considered normal, even though by retrospective evaluation there is a small area of hypokinesia in the apex.  He then had recurrent pain after a few hours, with ST-elevation.

Tissue Doppler based strain rate and strain showing hypokinesia in the apex (yellow and red curves) , peak systolic strain of - 5% and -8%, strain rate of - 0.35 and -0.8 s-1 both segments with post systolic shortening, as contrasted with normal deformation in the base (green and cyan). This is an indication of a small ischemic insult at the time of the first pain episode as also shown by the troponin results.
Angiography at the time of recurrent pain showed a tight LAD stenosis (top), confirming the strain findings, it was treated with PCI and stent (bottom) in the same procedure. Strain and strain rate values were normal after one week.




2D strain of the same recording (B-mode loops without TDI). The curved M-mode gives apical strain of  -14 and  -15%, i.e. borderline normal. This was the default ROI.
Adjusting the ROI making the apical segments straighter, reduces apical strain to -9 and -11% (borderline abnormal)
Both images were made with default (medium) spatial smoothing, but the values did not change more than 1% by reducing smoothing to minimum. In this case, the curvature effect is probably more important than the smoothing, although both factors may contribute.




Inferior infarct in a rather foreshortened view, resulting in a spherical image. By visual assessment this infarct is akinetic in the basal segment
Strain by tissue Doppler, showing systolic akinesia in the basal segment (cyan curve) - mark how the ROI is placed to avoid the lower part of the segment where there is angle discrepancy), and normal strain in the apical segment (yellow) and the anterior wall (red).
Strain by 2D strain, showing borderline reduced, but still viable strain of - 12% in the basal segment. IN this case, the near akinetic segment has a strain that mainly is due to the inward motion as described in the diagram above. In addition, the ROI, being the same all the way around, overestimated the wall thickness in the infarct, also contributing to the curvature dependent strain, which is dependent on the ROI width. In this case, the effect is due to the curvature, not smoothing, reducing smoothing did not reduce strain in the infarct zone at all.
.

The curvature dependency of 2D strain is a parallel to the angle dependency of tissue Doppler.

Smoothing in 2D strain

There is a liberal amount of temporal smoothing. In addition there is built in a spline or polynomial smoothing along the whole region of interest (ROI). The AV plane is the heaviest feature that is tracked, and contributes the most to the motion, which is then distributed along the whole ROI by a curve fitting along the mid ROI curve, resulting in a smoother transition from segment to segment, distributing the deformation along the ROI. Thus, the 2D strain application does not reflect pure speckle tracking, but also a great amount of model fitting. The spline smoothing is weighted, being least in the basal segments, most in the apex, which may be a way to compensate for some of the curvature effects in the apex, but that means tat adjustments in the base will affect measured strains in the apex.


The spatial smoothing, however, is adjustable. By default, the smoothing is medium, and can be adjusted to both maximum and minimum. However, in earlier versions of the software, the regulation of the smoothing did not carry over to the measured segmental values displayed in the quad screen view. Thus, segmental values remained medium smoothed, and values with less smoothing had to be taken from the traces. This problem seems to have been fixed in the latest version of the software (2011), but this means the software version should be taken into account. Also, the adjustment seems to be small, compared to the total amount of smoothing.


This is illustrated below.



Spatial smoothing in 2D speckle tracking strain. The smoothing, using longitudinal information from the AV-plane motion results in strain values even in areas where there is no speckles. This means that there is movewment of the points of the ROI, in areas without speckles. This is due to the information from the AV-plane motion being "splined" along the ROI, and the motion of the points is due to this, not to local tracking. Thus, there is  only "virtual tracking", and the true local tracking cannot be fully assessed . Segmental values from this tracking. Left: medium (default) spatial smoothing, right: Minimum.  This was done in the latest software version (2011), where values and traces carry over to segmental values after adjustment of smoothing, but the effect is small, compared to the total amount of smoothing.

This example was obtained by manual override of the automated positioning of the ROI, as well as manual override of the automated quality check, which suggested rejection of all segments. However, the application did track, as can be seen above, left, and these images are to show the general principle.



Inferior infarct in two chamber view, being akinetic in the basal inferior wall, although with considerable passive motion due to tethering
2D strain (left) vs. tissue Doppler (right) in an inferior infarct, analysed from the same cine loop recording.  The akinesia in the base is missed due to smoothing. In this case, as there is dropout of the whole anterior wall, the smoothing may be harder in the remaining three segments.  Also, in basal infarcts, the effect on AV-plane motion is less (40) as shown above.




Inferior infarct. Hypokinesia of the basal segment. Not immediately evident.
Strain and strain rate. Basal hypokinesia and post systolic shortening (yellow). Also normal curves in the inferior apex as well as in the anterior wall (red and cyan).



Same infarct as above. Tracking shows poor tracking in the basal and midwall segments have poor thickening due to poor tracking. The anterior wall is less visible. The segments are not approved for analysis.
Longitudinal strain. The apical anteior segment shows reduced strain, but this is due to poor tracking. The basal segment does not show reduced systoloic strain. However, looking at the curve, the infarcted segment does show post systolic shortening, so the infarct is still indicated.

Lateral tracking in 2D strain

Transverse displacement an velocity can also be derived, but as this will be the segmental average, this value has little meaning, the velocity and displacement increases from epicardium to endocardium. It is the displacement and velocity gradient that is of interest, i.e. transverse strain and strain rate. However, this can also be calculated by this method:

Longitudinal Transverse




Strain
rate




Strain

Longitudinal and transverse strain derived from speckle tracking.  It can be seen that in this case the differential tracking in the transverse direction is poor in the basal segments, thus underestimating transverse thickening in this healthy subject. (Images with better resolution may be seen above).

This is one of the fundamental limitations of speckle tracking as discussed previously.

Transmural and circumferential strain.

As speckle tracking is partially angle independent, it may be applied to the short axis as well. The main concern about tracking in short axis views, however, is the long axis motion. This means that there is between 1 and 1.5 cm out of plane motion of the base, and about half that in the midwall. That means that the tissue present in end diastole is not the same as in end systole. This also means that the speckles that are tracked do not represent physical myocardial points. Thus, the meaning of transmural and circumferential strain becomes slightly dubious. However, this do not only pertain to 2D strain. As shown above, this is the same problem even in parasternal M-mode. (Which, despite this, has worked well for 50 years). However, this remains a caveat when new measures are added. In the question of rotation, especially torison, the spiral course of the longitudional fibres may even cause the displacement to cause the fibres to be traced as rotating around the cavity centre.




Parasternal long axis image. The longitudinal motion of the basis is evident.
End diastolic parasternal long axis image.  The yellow line crosses the ventricle near the middle.
End systolic parasternal long axis image from the same loop.  The yellow line crosses the ventricle much closer to the base transecting a different part of the myocardium.

The speckles may be the endocardial borders, or even the fibres that may run in spiral. Thus, in the base, the physiological meaning of the obtained values is questionable.


Accepting the validity of speckle tracking in short axis views, it then allows tracing of transmural and circumferential strain. Transmural strain is wall thickening, and the tracking in the transmural direction will be dependent on the resolution, which is better along the ultrasound beam than laterally. The physiological meaning of circumferential strain, shouold be midwall circumferential shortening, which actually is nothing more than * midwall fractional shortening as reasoned above



2D strain applied to short axis image. Again this can be seen to track in two dimensions, the thickness following the wall thickening, and the mid line in the ROI Showing midwall circumferential shortening.
Transmural strain. In this image the application only measures between 10 and 15% transmural strain, while the true values in a normal person as this may be as high as 40 - 50%. This is probaly due mainly to a too thick ROI (default), although poor lateral tracking combined with smoothing may contribute.
Circumferential strain from the same processing.   In this image about 15%, which is closer to normal. This, however, does not mean that the circumferential strain is more reliable, it means that the thickness error in the ROI is compensated by an underestimation of the cavity volume. It's equivalent to the fractional shortening increasing in hypertrophy, despite reduced wall thickening. (Actally circumferential strain = * midwall FS. )


Width of the ROI

Transmural strain all thickness and wall thickening. But in the 2D strain application, this means ROI width as shown below.



Normal ventricle in short axis view.
The loop can be used to generate an anatomical M-mode, the line is skewed to avoid the papillary muscles. On this M-mode the following values were measured: LVIDD: 53mm ,LVIDS: 36mm, giving a FS of 32%, IVDS 7 mm, IVSS 11mm, giving a wall thickening of 57%, LVPWD 8mm, LVPWS 11mm, giving a wall thickening of 38%, and a mean wall thickening of 48%.

Below are shown transmural strain by 2D strain with different ROI width. The images are all processed from the loop above, and endocardium traced in standard manner. In reprocessing, only ROI width was changed without changing the initial contour. All ROI's were accepted by the analysis software for all segments:







Transmural strain with narrow ROI setting. Tracking seems fair, except perhaps in the lateral wall.  This is due to drop out of the wall, not hypokinesia, as seen to the right, where tracking is far better.  However: Given poor tracking in the lateral wall, mean WT = 57%. This is due to the tracking in the septum, which is good, but where the wall thickening is absurd, 77 and 91!; normal  endocardial motion in absolute terms, gives a too high relative wall thickening in percent of the narrow ROI.
Transmural strain with default (medium) ROI setting. Tracking seems fair.  Mean WT = 58%.
Transmural strain with a wide ROI setting. Tracking seems fair.  Mean WT = 37%, because a normal endocardial motion in absolute terms will result in a low percentage of the too wide ROI.
The measured wall thickening is evidently as expected a function of diastolic ROI width, as expected. Compare also mean and the relevant segments with the values above

It is evident that the transmural strain is extremely sensitive to the ROI width. This is pertinent to long axis analysis also, as the curvature in the apical segments will lead to an increased susceptibility of the ROI width. This may be some of the reason why Becker et al (212) found transmural strain even in segments with total transmurality of scars, and not tethering as presumed.

Repeatability of 2D strain.

Basically, the 2D strain application, due to a high amount of smoothing, should have a high repeatability, as shown here. However, this will only be the case as long as the tracing is done in the same manner each time, in the same loops. This means a very standardised endocardial tracing, and a standardised ROI width. As shown above, the values are extrmely dependent on the ROI, both curvature and width of the ROI. Utilising the automated features of the application will ensure this, but will not necessarily ensure the correct shape and  width of the ROI, and hence, not necessarily the correct values either. In a study (208) where repeated measurements in the same loops was compared for different centres, the 95% limits of agreement were -11.4% to +11.8%, but with very little bias. Repeated recordings within one hour (presumably by the same observer), had limits of agreement of -9.6 to + 9.7%.


Both segmental strain and 2D strain have been compared for longitudinal strain, and compared to tissue Doppler (151, 153) as shown in this table. Both seem to agree fairly well. In addition variability of strain rate (but not strain) is lower by both methods than by tissue Doppler. However, both Segmental strain and 2D strain use automatic segmentation, this may be some of the reason for better repeatability, not speckle tracking vs. tissue Doppler per se. However, the higher variability of strain rate by velocity gradient, shows this method to have a somewhat higher noise componenet. Feasibility of both methods is reported to be between 70 and 80% of segments (lower in the HUNT study,but this is due to the aim of the study, to provide normal values as free as possible from artefacts.


Summary of differences and limitations of different methods.

It is important to be aware of the limitations of each method. It should also be emphasized that different methods are not necessarily directly comparable, and may yield different normal values and cut offs, due to the different ways parameters are measured. One of the fundamental differences stem from the different geometrical assumptions that are present as shown below:


Differences in geometry between methods. The fairly invariable outer LV contour is shown in heavy black. The diastolic inner contour, segmental borders, kernel positions and measurement lines are shown in light black. Systolic inner contour,
segmental borders, kernel positions and measurement lines are shown in red. Left: Segmental strain by tracking of kernels at segmental borders. It can be seen that the main deformation is measured along the longitudinal axis of each segment. As the wall thickens, the longitudinal mid line of the segments moves inwards, but in the basal and mid wall segments this does not add to the shortening as the angle does not change much. In the apex, however,  the angle of the center line changes,  contributing to the segmental shortening when it is measured by this method, however, the effect is slight.  To the right is shown the geometric assumptions of the 2D strain method.  The ROI uses an assumption of equal thickness from base to apex, and the mid line moves with the thickening of the contour.  The segment length is measured along the curved line, and both the curvature and the angle contributes to the shortening of the segment mid line as it moves inward. Thus, the shortening (strain)  might be expected to be higher in the apical segments by this method, as well as being dependent on the curvature, especially in the apex.  (However, this effect may be masked by the high degree of smoothing inherent in the application, which may distribute the differences between segments.  Ultrasound beams are shown in blue, illustrating the alignment problem of this method,  thus resulting in lower values in segments that are poorly aligned.

The main limitation of any echo method is the ones related to data quality.

AS discussed under each method;
  • The fundamental limitations related to all methods are the ones arising from:
  • Tissue Doppler, having the advantage of high frame rate has additional limitations related to:
  • Speckle tracking (in any form), being less angle dependent has additional limitations relating to:
    • Frame rate, with the risk of undersampling.
    • Tracking quality at high HR
    • With decreasing lateral resolution the method becomes more angle dependent, although to less degree than tissue Doppler.
  • Segmental strain, being robust and giving the opportunity of utilising both tissue Doppler and speckle tracking and eliminating the angle problem, has the additional problem of:
  • The 2D strain application, being robust and user friendly, has the additional problems of:
    • Smoothing, relying heavily on AV-plane motion,
      • which may give strain values even where there are none, and may reduce sensitivity for reduced regional function
      • Makes the tracking more difficult to assess visually
    • Curvature dependency, due to the technicalities of the specific applications, which may give too high values in the apex.
    • ROI width seems to be critical, espacially in transmural strain.

Comparison between methods in HUNT

In a recent study of normal values (153) we have compared the different methods for deformation measurement in a subset of 57 patients :
  1. The combined tissue Doppler - speckle tracking method described above
  2. Longitudinal velocity gradient from tissue Doppler without tracking of the ROI, this is similar to the longitudinal velocity gradient by commercial software, although in this application obtained by the experimental software.
  3. Longitudinal velocity gradient with tracking of the ROI, which can be done, althoufgh only approximate, by manual adjustment in commercial software, and
  4. Speckle tracking with the 2D strain application.

The velocity gradient is analysed by customised software, but the basic principle is exactly the same as in commercial software (EchoPAC), except allowing for automated analysis and automated tracking. The results were as follows:


Method 1: segment length by TDI and ST
Method 2: Velocity gradient (stationary ROI)
Method 3: Dynamic velocity gradient (tracked ROI)
Method 4: 2D strain (AFI)

Peak Strain rate
End systolic Strain
Peak Strain rate End systolic Strain Peak Strain rate End systolic Strain Peak Strain rate End systolic Strain
Apical -1.12 (0.27)
-18.0 (3.6)
-1.46 (0.85)
-14.6 (9.0)
-1.31 (0.73)
-17.2 (9.1)
-1.12 (0.37)
-18.7 (6.6)
Midwall
-1.08 (0.22)
-17.2 (3.2)
-1.29 (0.56)
-18.2 (7.4)
-1.40 (0.58)
-16.9 (7.1)
-0.99 (0.23)
-18.3 (4.7)
Basal
-1.03 (0.24)
-17.2 (3.5)
-1.71 (0.94)
-19.6 (9.3)
-1.59 (0.74)
-17.1 (8.6)
-1.12 (0.36)
-18.0 (6.2)
Mean
-1.08 (0.25
-17.4 (3.4)
-1.45 (0.79)
-17.7 (8.5)
-1.43 (0.67)
-16.7 (8.1)
-1.07 (0.33)
-18.4 (5.9)
Comparison between methods. Standard deviations in parentheses. Thanks to Eirik Nestaas, MD, PhD for discovering a typographical error in this table, that now has been corrected (bold types).
Looking at the findings, it is evident that the tissue Doppler methods gives close to 30% higher peak strain rate values that the two other methods. This is probably due to a higher random noise component in tissue Doppler, rather than the opposite, too low peak values due to under sampling in the two other methods. This is evident from two reasons:

  1. Tissue Doppler derived strain rate shows a far wider standard deviations
  2. Integrated strain from strain rate eliminates the differences, showing that the noise is random.

Thus the tissue Doppler is more sensitive to noise than other methods. However, systolic strain values were very similar with all methods (except in the apex for TDI with fixed ROI), showing that the smoothing that is a function of temporal integration eliminates this problem, and basically in strain measurements tissue Doppler is as reliable as other methods, although still with somewhat higher standard deviations.

Another thing is also evident: Tracking the ROI in tissue Doppler results in equal strain values in apex, midwall and base, as in the other two applications, while no tracking yields lower values in the apex. This is due to the variable angle in the apical segments, as the segment becomes shorter, they also become more crosswise.  so there is an advantage by tracking, but only in the apical segments. There was no difference in strain rate, only in strain, but as peak strain rate is early in the systole, while peak strain is (near) end systolic the effects of tracking may be greater. The reason is probably poorer alignment in end systole if the ROI is not tracked.

In a study of the sensitivity of strain rate imaging in stress echo (128), no significant difference was found between the segment length and the dynamic velocity gradient, despite the higher noise component in velocity gradient, nor between peak systolic strain rate versus end systolic strain.

Finally it seems that the combined method and 2D strain gives almost the same results, , although there was some statistical differences, these were small and of little clinical importance. Standard deviations, as a measure of variability were also comparable, in the combined method probably due to low noise because of low spatial resolution, in the 2DS probably due to smoothing. But it seems that the normal values are transferable, and for strain between  methods. In a reproducibility study (154) the two methods also had similar inter observer reproducibility (different recordings and analysers).

Other authors have found a base to apex gradient (highest strain in the apex) in strain values with the 2D application (155), but this may be due to the curvature effect in the apex, although MR studies report the same finding. Strain measurement, however, is still dependent on the analysing method,  and even the definition  may vary, depending on how the strain length is defined.

Tracking in RF data:

AS the reflected ultrasound signal really consists of not only information about the amplitude and wavelength, but about the actual waveform being reflected, this information can be extracted. From these data, both grey scale information and Doppler data can be calculated. (In fact this is what is done by the scanner, before giving what is known as "raw data".) Extracting the RF data themselves requires far more storage, and thus computational power in post processing, and has so far been slow. However, the RF data can also be used for tracking, in a matter similar to speckle tracking. Both cross correlation, normalised cross correlation, sum of absolute differences and sum of squared differences has been shown feasible (129). The method has been validated in both phantom (130) and animal experiments (131). This method has the advantage of being angle independent, as well as having the raw material for both tissue Doppler and grey scale image formation. Theoretical considerations indicates that this is advantageous in dealing with reverberation artifacts as well, compared to clutter filtering, but this remains to be shown. The clinical feasibility is so far not clear, as the method is time consuming and demanding in computational power.

Next section: is deformation imaging useful?



Back to main website index.

References:




Editor: Asbjørn Støylen Contact address: asbjorn.stoylen@ntnu.no, Updated: 2011