DSC Calibration Wizard Resistance & Capacitance

(Step 2)

Sapphire Run

The second experiment required to obtain Tzero cell resistance and capacitance values involves heating the cell with two equal weight sapphire disks on the sample and reference platforms over the same temperature range and under the same experimental conditions as the previous Baseline Run. Since identical experimental conditions are required, you will not be prompted to enter that information again. You only need to weigh and enter the weights of the sapphire disks.

Follow the instructions on the calibration window using the information below as a guideline:

Once all of the experimental parameters are entered, press Next to start the sapphire run experiment. For information on the next step, setting up a cell constant experiment, click here more.

For information on the Experiment in Progress window or on results of the calibration experiment click here DSC Calibration Wizard Experiment.

Back to top

Related Topics