Displacement and
velocity

|
Strain and strain
rate

|
Top, mitral annular
displacement curve, being the curve showing the
longitudinal shortening of the left ventricle.
Below, the tissue velocity curve, which is the
temporal derivative of the displacement curve.
Comparing to the volume/flow curve, it is evident
that there is more complex motions, especially n
elation to the isovolumic phases, than is evident
from the mere volume diagram to the left. |
Top, strain curve from
mid septum, showing the deformation, below the
strain rate (temporal derivative). The curves seem
to be very similar to inverted motion and velocity
curves, however, deformation will show more
regional detail as discussed here.
Remark also how the strain curve is similar to the
volume curve, showing the same pattern, while the
strain rate (temporal derivative of strain) is
similar to the flow curve (temporal derivative of
volume).
|
It has been established that the longitudinal shortening of
the left ventricle, and thus the longitudinal measures is
closest related to the stroke volume and EF, i.e. to the total
left ventricular volume change (
13,
30
-
35,
56,
59,
60,
64
- 67,
116).
Thus,
the longitudinal strain is the most important measure, and it
is also closely related to the wall thickening and thus
internal shortening as discussed
here.
It is obvious that the LV shortening and the ejection are
interrelated. In fact, the LV systolic shortening * the
circumferential area should be aproximately equal to the
stroke volume.

|
 |
Stroke volume by
Doppler flow velocity integral (VTI) and LVOT
diameter. The diameter gives the area, and the
velocity time integral gives the distance that an
object travels if it follows that velocity curve
(v =ds/dt, means that s = v
dt. Multiplied with each other, the area and VTI
gives the volume of a cylinder, equal to the
stroke volume.
|
Relation of stroke
volume and LV shortening. The volume reduction is LV shortening * LV area
at the mitral plane. As area is far higher, the
distance is far smaller than the VTI.
|
Global systolic function
measures
It should be emphasized that normal limits (A) are
not the same as cut off limits between a normal
population and a defined patient population (B).
Secondly, the diagnostic accuracy of a method is a
function of both the spread of measurements (measurement
plus biological variability - C) and the separation of
means between the two populations (D).
Peak systolic versus end
systolic measures of ventricular function.
Peak systolic measures are the measures of peak ventricular
performance, and are basically
- peak ejection velocity in the LVOT,
- peak annular systolic velocity, and
- peak global ventricular strain rate.
These occur early in systole, and may be less load dependent,
as maximum afterload is reached later in systole. They all
occur during the first part of systole, and thus are more
closely related to contractility, and especially to
contractility
changes, as shown in studies (
78,
79,
80,
223).
All such studies are really
studies in contractility changes, and thus, useful to
separate contractile states,
rather than measure contractility direct.
However, they are not completely load independent, as
increased load will result in a delayed and blunted
development of force and velocity, as opposed to the
pressure/volume relation.
End systolic measures on the other side, are measures of the
total work performed by the left ventricle during ejection.
This is influenced not only by force, but also by load
(resistance), and the ejection time (HR). They are
- stroke volume,
- Ejection fraction (and fractional shortening)
- annular displacement and
- global strain
There is, however, little evidence directly comparing
displacement / strain to velocity / strain rate at varying
load, and the few and small studies that are published seems
to indicate a very similar load response. However,
increased contractility will not to the same degree lead to
increased stroke volume, if there is no concomitant increase
in venous return, as in inotropic stimulation. Thus stroke
volume wil be maintained, but at a lower end diastolic volume.
This means end systolic measures will be less sensitive
to contractility increase as discussed
above
(
223).
 |

|
Peak systolic annular
velocity (S'), which is early systolic,
compared to peak annular displacement (MAPSE),
which is end systolic. S' is actually the peak
rate of annular displacement, and is thus closer
to contractility, while MAPSE is end systolic and
thus closer to ejection fraction. |
Peak strain rate, being early
systolic, compared to peak strain, which is
mainly end systolic, and closer even, to ejection
fraction.
|
Most of the indices above have been studied, and are
established as indices of ventricular function. However, in
addition to they all being only imaging indices, they have
different shortcomings, and to some degree slightly different
interpretation physiologically.
Even
LV elastance is an end systolic
measure, although it is taken as the real contractility. This
may be only in relation to volume, in fact.
Peak
systolic measures - contractility indices
Even though
contractility per se
cannot be measured by imaging alone, the measurement can
be approximated, but early systolic measures come close (
78,
79,
80,
223).
Peak annular systolic
velocity
Peak systolic velocity (S') was early validated as a measure
of systolic function (
37,
38,
39,
40).
Peak annular velocity occurs early in systole, and may be less
load dependent, as maximum afterload is reached later in
systole. The peak velocity is taken as an average measure of
two or four points around the mitral ring.
 |

|
Pulsed
tissue Doppler of the mitral ring. These are
the velocity traces of the longitudinal motion,
while dividing by the end diastolic length results
in an approximation to the
Lagrangian
strain rate .
|
Age
dependent peak systolic, early and late diastolic
velocity in normals from the HUNT study (165). The early diastolic
velocities are higher than the systolic, and the
decline is thus steeper, but the relation is
evident. |
The peak systolic annular velocity is useful in that it is a
better marker of systolic function than EF, and that it offers
a measure that allows direct comparison of systolic and
diastolic
function.
as they are measured by the same method.
But there are some slight limitations:
1:
Peak systolic velocity is not a direct measure of peak
rate of shortening.
Even though it seems intuitive, the comparison with peak
strain rate shows that there is a velocity component in the
peak that is a global translational motion toward the apex as
discussed
here.
This is due to the recoil force from the ejected blood.

|
 |
Peak velocities from the
myocardial wall are shown in blue and green,
showing parallel velocity curves at peak, thus
identifying the velocity as translation, which do
noe show in the difference (strain rate) curve
(red).
|
Peak velocities along the
septum, showing a slight blunting of the velocity
peak in the apex. Even if there is a translational
velocity, this will decrease in the apex where the
myocardium is pressed into the chest wall. The
strain rate curves corresponding to the regions
between the velocity curves are shown below,
indicating the this velocity is not the true
deformation rate.
|
It might be argued that the recoil velocity is also generated
by myocardial contraction, and represents a liberation of the
work done during IVC. From this argument, peak velocity might
be considered a fully legitimate measure of myocardial
performance, maybe even more legitimate that peak rate of
volume decrease.
2:
Peak systolic velocity is not always simultaneous in all
segments of the mitral annulus.
The normal pattern of annular velocities varies in the normal
subjects:

|

|

|
A fairly common pattern is a
sharp peak in the lateral annulus (cyan), and a
more rounded curve with a later peak velocity in
the septum (yellow). Thus, the divergence of the
curves in the initial
ejection phase may represent a light tilting
(rocking) of the apex toward the septum. |
A slightly different normal
pattern where the initial peak in the lateral wall
decelerates slightly, the accelerates again,
giving a later second peak. The septum shows an
even curve, but with peak velocity between the two
peaks of the lateral wall.
|
In this case peak annular
velocity is early and simultaneous in both walls. |
The varying timing of the peaks is most probably due to
different impact of the recoil force from ejection, creating a
slight rocking motion of the whole heart. AS in the example to
the left, this will mean that the lateral peak is exaggerated
due to rocking, while the early septal velocity is blunted,
and the later peak is due to this blunting.
The impact of the recoil momentum on the septal and lateral
annulus will, of course, depend on the angle between the
momentum vector (velocity vector), and the ventricular long
axis. As the aortic opening is situated in the septal part of
the LV base, the angle deviation, if any, can be expected to
be towards the septum, delivering the highest impact
laterally. This is in accordance with clinical observation,
the peak is most consistently present in the lateral wall.
However, the ejection from the right ventricle must also be
taken into consideration, being nearly simultaneous, and with
the same stroke volume (mass), only the difference in
velocities will account for the difference in momentum. The
pulmonary ostium is also situated medially, in front of the
aortic, but often with less of an angle deviation. However,
the angle will be opposite, and may counteract the aortic
momentum.
Thus, both the actual value and the timing of peak systolic
velocity can be dependent on the site where it is measured as
shown above. This, of course means that the peak annular
systolic velocity which is used as a systolic functional
parameter, measured either as one site, or as an average, is
an approximation, as the timing may differ between sites.
The correlation with EF is weaker than for MAPSE, which is not
unsurprising, EF and MAPSE being
end
systolic measures, and as such measures of the total
systolic work, S' is peak systolic, measuring peak systolic
performance.
One of the main advantages of tissue velocities is that
systolic and
diastolic
function are measured by the same method. From the
beginning, systolic function by EF was compared to diastolic
function by mitral flow, equivalent to comparing apples with
bananas. This lead to the concept of pure diastolic
dysfunction, which has later been shown to be erroneous (
202).
The correlation between systolic function S' and diastolic
function e* was found in an early study to be 0.6 over a wider
range of ventricular function (
201),
and in the HUNT study (
165)with
a
large number (N=1266) and limited to healthy subjects, the
correlation was found to be 0.59.
The correlation reflects among other things, the physiological
mechanism that much of the diastolic recoil is due to elastic
stored energy from systolic contraction (restoring forces),
but also, and most important:
that systolic and diastolic function are closely
related.
In another study (
202)
it was found that the systolic function by S' was reduced in
patients with heart failure with normal ejection fraction.
This led to a renaming of the state that up to then was called
"diastolic heart failure" to "heart failure with normal
ejection fraction". This, of course corrects the implied, but
mistaken assumption that there existed a pure diastolic
failure. However, it does not address the fundamental problem,
which is one of methodology, that EF should not been used in
normal sized or smaller ventricles.
The S' has been shown to be sensitive for reduced function in
relatives who are mutation positive, of patients with manifest
hypertrophic cardiomyopathy, despite having normal EF and no
hypertrophy (
203).
The diastolic function by tissue Doppler was similarly
decreased. It also correlates better with BNP in heart failure
than the fractional shortening (
204).
Thus, the peak systolic annular velocity is useful in that it
is a better marker of systolic function, and that it offers a
measure that allows direct comparison of systolic and
diastolic function.
Where
and
how should measurements be done?
As the peak velocities are more often higher in the lateral
than the medial, it is evident that the measurements are
different if different sites are chosen. This can be seen
from the HUNT study (
165).
This study consisted
of 673 women with a mean BP of 127/71 ,mean age of
47,3 years and BMI of 25.8 and 623 men, with mean BP of
133/77, mean age of 50.6 and BMI of 26.5. Both sexes were
normally distributed with an SD of 13.6 and 13.7 years,
respectively. 20% of both sexes were current smokers. Basic
echo findings are in accordance with other studies,
like the findings of Schirmer et al (
156,
157),
so
the study population may be assumed to be representative.
|
Anterior
|
(Antero-)lateral
|
Inferior
|
(Infero-)septal
|
PwTDI
S' (cm/s)
|
8.3
(1.9) |
8.8
(1.8)
|
8.6
(1.4)
|
8.0
(1.2)
|
cTDI
S' (cm/s)
|
6.5
(1.4)
|
7.0
(1.8)
|
6.9
(1.4)
|
6.3
(1.2)
|
Results from the
HUNT study with normal values based on 1266 healthy
individuals. Values are mean values (SD in
parentheses).
The maximal differences can be seen to be about 10%
relative, with the highest values in the lateral wall,
lowest in the septum. The reason for this, can be
partly
explained by the differences in length of the walls,
seeing that the peak strain and strain rate varies much
less.
The initial studies (
37,
38,
39)
used the average of four sites as a measure of global
systolic function. In the HUNT study, however, there were no
difference between the peak systolic velocity (S') mean of
lateral and septal, and the mean of all four points.
However, Thorstensen et al (
154)
did show that reproducibility was about 35% better using
four point average (p<0.001), in line with what was found
earlier (
40),
even
if the mean values were the same.
The common method of measuring peak
velocity at each point and then average the peak values
from two or four point, is methodologically slightly
unsound, as they may not be simultaneous:
Septal (yellow) and lateral (cyan) velocity
curves from the fisrt subject above. Peak velocities
are 6.25 cm/s septally and 7.6 cm/s
laterally. Mean of peak values are thus 6.93 cm/s.
The averaged curve of the two is shown in red, and
the peak of the average is 6.67 cm/s. Difference
here is small, but this may not always be the case.
The point from a puristic view is that if the peaks are not
simultaneous, the mean peak velocity doesn't exist in real
time (cfr. the peak-to-peak gradient of invasive aortic
stenosis measurement). Still, the method has been established
as useful, and normal values for the averages has been
established (
165).
And, as in the example above, while the early peak in the
lateral wall is exaggerated due to the rocking, the early
septal velocity is blunted. The true peak translational
velocity is seen in the average curve, and the true peak
velocity is the peak of the average curve, not the average of
the peaks.
The varying timing of the peaks is most probably due to
different impact of the recoil force from ejection, creating a
slight rocking motion of the whole heart. AS in the example to
the left, this will mean that the lateral peak is exaggerated
due to rocking, while the early septal velocity is blunted,
and the later peak is due to this blunting.
In some cases, the rocking of the apex, even if the ventricle
is normal may become completely misleading.

|
|
Rocking heart with normal
ventricular function.
|
Peak velocities have totally
different timing, and much of both of the peak
components are due to translation.The
septal peak has a component of rocking to the
right, the lateral peak a component of rocking to
the right, both may be overestimates. |
As discussed
here.
In this case, the peak velocities should be viewed with
skepticism as functional measures. The mean curve might give a
more correct estimate, although this is not validated, and is
not available in standard analysis software.
Using pulsed wave tissue Doppler, this is not an option, and
curve averaging is not standard analysis software.
Normal values for
systolic velocities of the right and left ventricle from the
HUNT study by age and gender. From (165).
|
Left
ventricle, mean of 4 walls
|
Right
ventricle (free wall)
|
|
S'
(pw TDI)
|
S'
cTDI
|
S'
(pwTDI)
|
Females
|
|
|
|
< 40 years
|
8.9
(1.1)
|
7.2
( 1.0)
|
13.0
(1.8)
|
40 - 60 years
|
8.1
(1.2)
|
6.5
(1.0)
|
12.4
(1.9)
|
> 60 years
|
7.2
(1.2)
|
5.7
(1.1)
|
11.8
(2.0) |
All
|
8.2
(1.3)
|
6.6
(1.1)
|
12.5
(1.9)
|
Males
|
|
|
|
< 40 years
|
9.4
(1.4)
|
7.6
(1.2)
|
13.2
(2.0)
|
40 - 60 years
|
8.6
(1.3)
|
6.9
(1.3)
|
12.8
(2.2)
|
> 60 years
|
8.0
(1.3)
|
6.4
(1.2)
|
12.5
(2.3)
|
All
|
8.6
(1.4)
|
6.9
(1.3)
|
12.8
(2.2)
|
Annular velocities by
sex and age. Values are mean (SD). pwTDI: Pulsed Tissue Doppler
recorded at the top of the spectrum with minimum gain, c
TDI: colour TDI. Normal range is customary defined
as mean ± 2 SD.
The study is based on 1266 healthy individuals from the
HUNT study by Dalen et al (
165).
The
age dependency of values is evident. Colour tissue Doppler
gives mean values, which are consistently lower than pulsed
wave values, as discussed
here.
It is evident that the systolic values decline with age, as do
the early
diastolic.
It is important to realise that the peak values obtained by pw
tissue Doppler are higher, due to the breadth of the spectrum,
while colour tissue Doppler gives mean velocities, thus being
modal velocities in the middle of the spectrum. However, peak
values by spectral Doppler are affected by gain settings
(increased gain - broader spectrum - higher peak values),
while colour Doppler are affected by
clutter
(stationary reverberations - zero velocity - reduced average).

|

|
Same tissue Doppler recording
with two different gain settings. We see that peak
systolic velocity differs by 2 cm/s, and the
lowest gain setting is closest to the modal
velocity. However, the modal velocity itself,
remains unchanged by the gain setting.
|
There is a band of clutter
close to zero velocities, but as seen here, the
spectral modality makes it very easy to separate
the true and clutter velocities. However, the
clutter affects the autocorrelation velocity (red
line), giving lower velocities, but with clutter
filter this effect is removed (blue line), and the
peak value is substantially higher. Image modified
from (268).
|
Peak acceleration (??)
As acceleration precedes velocity, and is at the time of peak
rise of velocity, this should be slightly earlier than peak
velocity, and thus even less load dependent, - at least
afterload, preload dependency will still be present.
Acceleration is also more closely related to peak force by
Newton's second law (F = ma). In addition, if taking the peak
velocity to be partly a function of the recoil, it is also
caused by the pressure buildup, even more so than the peak
velocity. Thus there are hypothetical advantages in relation
to physics and physiology.
However, the concept is ill defined. Also, the temporal
derivation of acceleration from velocity will result in a less
favourable signal-to-noise ratio than the velocities.

|

|
Velocity curves from a normal
subject. The initial peak acceleration may be
defined by the slope of the tangent to the initial
velocity curve. As
illustrated, as the two curves from septum and
lateral walls are different, this will result in
different acceleration values. In addition, there
can be different ways of defining the tangent:
- A: The steepest slope of the lateral
curve
- B: The slope from nadir to the lateral
curve peak
- C: The steepest slope of the septal
curve
- D: The slope from nadir to the septal
curve breakpoint
- E: The slope from nadir to septal
curve peak
All of them reasonable, but resulting in
wildly different values (This is equal to a noise
component as shown right)
|
Real-time temporal derivation
of the two velocity curves. Due to the derivation,
the curves are fairly noisy, especially taking
into account the the velocity curves was somewhat
smoothed at the outset. It is obvious that the
peak values may be affected by the noise, incorporating
noise spikes. Averaging the septal and lateral
points, still doesn't solve the noise problem.
Also averaging can be done in two ways:
- Mean of peak values, which in this
case will be 186, and
- Peak of mean curve, in this case 163
And in addition, curve derivation and
averaging is not standard issue in analysis
software.
|
At present, the peak acceleration is less useful, being
- Closely related to peak velocity, as the peak velocity
is determined by the rate of velocity rise,
- In need of definition of concepts
- Dependent on heavy post processing, and not standard
analysis software
- while especially pw Tissue Doppler in the standard mode is a
quick, robust and online method.
Peak systolic
strain rate
Peak systolic strain rate is the peak rate of shortening. As
explained above,
this peak occurs later during the ejection phase than peak
velocity, as the strain rate algorithm subtracts the
initial velocty peak, being translation, not deformation:

|

|

|
Examining one normal subject
with early velocity peak in the lateral annulus:
|
Looking at velocities within
the wall in base an apex, the biphasic pattern
with an early peak can be seen in
both points in the lateral wall.
|
Examining the strain rate
from the entire walls between the apical and basal
points no sign of a biphasic shortening can be
seen, indicating that the lateral peak is only due
to translation, the peak being subtracted.
The peak strain rate is much later than peak
velocity in both walls.
|
The slight rocking motion affecting the timing of peak
velocities, means that there is no fixed relation between the
timing of peak strain rate and peak velocity:

|

|

|
Early peak velocity on both
sides. Slightly later peak strain rate in both
walls.
|
Early lateral peak velocity.
late septal; early septal peak strain rate, later
lateral, mean peak strain
rate later than mean peak velocity
|
Early lateral and later
septal peak velocity, earlier lateral than
septal peak strain rate, mean peak strain
rate later than mean peak velocity.
|
Thus, peak strain rate is the true measure of peak deformation
rate, i.e.
peak rate of shortening. However, this does
not mean that it is a truer measure of
peak systolic
performance, as the peak velocity incorporates the
ejection recoil due to the isovolumic pressure buildup.
But basically, as volume change is generally are
related
to longitudinal shortening, peak strain rate must be
close to peak rate of volume reduction, i.e. peak emptying
rate.
For global strain rate measures, the strain length as well as
the ROI should be as long as possible to reduce noise (
331)
as shown in the above examples and
discussed
in the measurements section.
Thus, the peak strain rate should be more related to peak
force than to peak rate of force development. However, in an
experimental invasive study, Greenberg (
80)
found a stronger correlation between both
dP/dt and
LV
elastance with strain rate than with systolic annular
velocities.
Normal values are necessary if measurements are to be used
diagnostically. In addition, they will give additional
information about physiology. In the north Tröndelag
population (HUNT) study, 1266 subjects without known heart
disease, hypertension and diabetes were randomly selected from
the total study population of 49 827, and subjects with
clinically significant findings on echocardiography (a total
of only 30) were excluded. (
153)
This is the largest normal population study of
echocardiographic strain and strain rate rate to date. End
systolic strain and peak systolic strain rate was measured by
the
combined
tissue Doppler / speckle tracking segmental strain
application of the Norwegian University of Science and
Technology, but the results were
compared
to
other methods in a subset of subjects, showing small
differences. The study consisted of 673 women with a
mean BP of 127/71 ,mean age of 47,3 years and BMI of 25.8 and
623 men, with mean BP of 133/77, mean age of 50.6 and BMI of
26.5. Both sexes were normally distributed with an SD of 13.6
and 13.7 years, respectively. 20% of both sexes were current
smokers. Basic echo findings are in accordance with
other studies, like the findings of Schirmer et al (
156,
157),
so
the study population may be assumed to be representative.
While differences between septum and lateral wall was of the
order of 10% in velocities, in deformation parameters (
153),
the
same difference was on the order of 4% in strain rate and only
1% (relative) in strain.
Normal
values for strain and strain rate per wall in the HUNT
study. From (153).
|
Anteroseptal
|
Anterior
|
(Antero-)lateral
|
Inferolateral
|
Inferior
|
(Infero-)septal
|
SR
(s-1)
|
-0.99
(0.27) |
-1.02
(0.28)
|
-1.05
(0.28)
|
-1.07
(0.27)
|
-1.03
(0.26)
|
-1.01
(0.25)
|
Strain
(%)
|
-16.0
(4.1) |
-16.8
(4.3)
|
-16.6
(4.1)
|
-16.5
(4.1)
|
-17.0
(4.0)
|
-16.8
(4.0)
|
Results from the HUNT
study (153,
165)
with normal values based on 1266 healthy individuals.
Values are mean values (SD in parentheses). The
differences between walls are seen to be smaller in
deformation parameters than in motion parameters, although
still significant due to the large numbers.
Normal
values for global strain and strain rate in the HUNT study
by age and gender. From (153).
|
Female
|
Male
|
|
End systolic strain (%)
|
Peak systolic strain rate
|
End systolic strain |
Peak systolic strain rate |
<
40 years
|
-17.9%
(2.1)
|
-1.09s-1
(0.12)
|
-16.8%
(2.0)
|
-1.06s-1
(0.13) |
40
- 60 years
|
-17.6%
(2.1)
|
-1.06s-1
(0.13) |
-18.8%
(2.2)
|
-1.01s-1
(0.12) |
>
60 years
|
-15.9%
(2.4)
|
-0.97s-1
(0.14) |
-15.5%
(2.4)
|
-0.97s-1
(0.14) |
Over
all
|
-17.4%
(2.3)
|
-1.05s-1
(0.13) |
-15.9%
(2.3)
|
-1.01s-1
(0.13) |
Values are given
as mean ( SD). The customary definition of normal values
as mean ± 2SD, giving about 95% of the normal population,
results in wider normal limits than previously shown as
cut off values in small patient studies. The values were
normally distributed, and with no clinically significant
differences between levels or walls. Values decline with
age, as does the velocity.
Normalised velocity(?)
As seen later, peak annular displacement can be normalised for left ventricular length
to derive a measure of global longitudinal strain. But this
doesn't necessarily mean that normalising velocity in the
same way gives global strain rate. Assuming that annular
velocity was the peak rate of ventricular shortening,
normalising for end diastolic length would give peak
Lagrangian strain rate, normalising for instantaneous
systolic length would give the Eulerian strain rate.
However, giving the difference in riming between the two
curves, due to the translational velocity, this is not the
case.

|
 |
Basal velocity traces, and
the velocity traces normalised for end diastolic
ventricular length (Lagrangian
normalisation), results in curves that resemble
inverted velocity curves, with the same shape.
|
Comparing this with the real
velocity / strain rate plots from the same
subject, it is evident that strain rate curves
have a different shape. (In fact, in this case the
lateral strain rate curve is more rounded, the
septal with an earlier and more defined peak,
opposite of the velocity curves).
|
This means that in terms of curve shape and timing, there is
no point in normalising the velocity curves, and the
normalised velocity curves is still a slightly different
measure than strain rate.
However, where there is a large variation in ventricular size,
as in children it makes sense, giving age independent
measures (
159,
214,
288).
Strain rate is one form of size normalising, but using pulsed
tissue Doppler, normalised velocity will make the velocity
measurements useful in children of all ages.
Basically, peak strain rate is most useful for assessing
regional function, where the motion due to tethering to
neighboring segments needs to be subtracted n(although the
effect of
segment
interaction remains). With uneven segmental
contractility, the peak strain rate in different segments also
becomes non simultaneous.
Peak ejection
velocity
LVOT ejection velocity must be proportional to flow, as the
LVOT diameter is considered constant. But this means that peak
systolic velocity is actually a direct measure of peak volume
reduction rate or peak emptying rate. And, as peak shortenoing
rate (strain rate) is very close to peak emptying rate, this
means that both measures are very close to measuring the same
thing. THis was evident in a study where both measures did
show a very similar change with chenges in contractile states
(
223).
In all four subjects, the peak
velocity of LVOT flow seems to be relativelt
simultaneous with peak strain rate, consistent with
theory. Averaged curves might be
even closer.
Thus, both theoretically, empirically and experimentally, the
peak LVOT flow and peak global strain rate seems to be
measuring very much the same event, even though indifferent
measures.
But should peak LVOT velocity be normalised for heart size?
Probably not, AS strain arte already is a norlmalisation for
heart size, the flow velocity is dependent on LVOT diameter,
whic increases with heart sise. So, the LVOT velocity is in a
way a normalisation of peak flow.
Normal peak
ejection velocities from Doppler, by age and gender from the
HUNT study (165)
|
Females
|
Males
|
< 40 years |
1.01 (0.17)
|
0.99 (0.17)
|
40 - 60 years |
1.02 (0.16)
|
0.99 (0.18)
|
> 60 years |
1.01 (0.17)
|
0.96 (0.18)
|
Over all |
1.01 (0.16)
|
0.98 (0.18)
|
Values are given as mean (
SD). The customary definition of normal values as mean ±
2SD, giving about 95% of the normal population, results in
wider normal limits than previously shown as cut off values
in small patient studies.
The only difference is that peak values of peak ejection
velocity do not decline with age. AS stroke volume gpoes down
with ventricular volume by age, this must mean a corresponding
reduction in LVOT diameter.
With the appearance of new methodology, a number of new
methods for measuring left ventricular global function has
emerged. Older measures has traditionally been measurements of
the
cavity function: Stroke volume,
ejection fraction (and the M-mode equivalent shortening
fraction). Newer methods include
longitudinal
measures of wall function, as annular displacement and
velocity, as well as mean strain/strain rate, either based on
segmental measurements, or a global averaging (as global
strain form speckle tracking
2D
strain). It should be of general interest to comment on
the relationship between the methods. It is also important to
realise that while strain and strain rate are measures of
shortening per length unit, the annular velocity and
displacement are also measures of the same, but in absolute
values (i.e. not normalised for ventricular length). However,
all measures that measure relations to changes, i.e. in paired
experiments of load alterations, the normalisation will cancel
out, and displacement will behave as strain, strain rate as
velocity. Thus all experiments with systolic displacement and
velocity in relation to global changes, will pertain also to
strain and strain rate.
End
systolic measures - systolic work indexes.
End systolic measures on the other side, are measures of the
total work performed by the left ventricle during ejection.
This is influenced not only by force, but also by load
(resistance), and the ejection time (HR). They are
- stroke volume,
- Ejection fraction (and fractional shortening)
- annular displacement and
- global strain
There is, however, little evidence directly comparing
displacement / strain to velocity / strain rate at varying
load, and the few and small studies that are published seems
to indicate a very similar load response. However,
increased contractility will not to the same degree lead to
increased stroke volume, if there is no concomitant increase
in venous return, as in inotropic stimulation. Thus stroke
volume wil be maintained, but at a lower end diastolic volume.
This means end systolic measures will be less sensitive
to contractility increase (
223).
Cavity measurements of systolic
function
Fractional shortening
As M-mode was the first echo modality, the fractional
shortening of the LV cavity was the first LV systolic
functional measure by echo. The fractional shortening is
defined as
FS = (LVIDD -
LVIDS)/LVIDD thus, in fact being an one-dimensional
version of EF. Diameter is conventionally measured to the
endocardium, so the fractional shortening is more precisely
the endocardional fractional shortening. It's less accurate
than the EF when there is regional dysfunction, as the
measured fractional shortening will be generalised to the
whole ventricle. It is quite common to measure longitudinal
strain, i.e. wall or segment shortening as a measure of
longitudinal function. On the other hand the fractional
shortening of the chamber diameter is a well established
measure of global and radial function. But in the case of
hypertrophy, this may lead to completely erroneous conclusions
about the changes in radial versus global function, as shown
in the theoretical treatment below.
The relation between wall thickening and fractional shortening
is ilustrated below:
In this theoretical M-mode of
the LV, a normal ventricle has a wall thickness of 1 cm,
an internal end diastolic chamber diameter (EDD) of 4 cm,
resulting in an external diameter of 6 cm. As most of the
wall thickening is inward, with little change in outward
diameter (except in the case of differing filling
pressures on the two sides), an end systolic wall
thickness of 1.5 cm will result in a diameter shortening
of 1 cm and an end systolic chamber diameter of 3 cm.
Thus, wall thickening (WT, transmural strain) is (1.5 cm -
1 cm) / 1 cm = 50%, chamber diameter reduction is 1 cm,
fractional shortening (FS) is (4 cm - 3 cm) / 4 cm = 25%. Thus, if wall thickening
decreases due to reduced myocardial function, so do
fractional shortening as seen in the middle figure. And if
there is dilation as well, the denominator will increase,
resulting in further reduction in FSas an inverse function
of the diameter. In
LV dilation, there is usually a combination of increased
diameter and reduced wall thickening.
The erroneous comparison
between longitudinal strain and fractional
shortening:
The
incompressibility
principle tells us that as the wall shortens in the
longitudinal and circumferential direction, it has to thicken
in the transverse direction, and the relation is
geometrically
determined. Thus the longitudinal and transverse
function as measured by strain should be interrelated. Reports
about radial compensation of reduced longitudinal function
is in direct opposition
to the incompressibility principle. The problem
arises if we do not measure the same values for longitudinal
and radial function.

Compared to the normal example to the left, in
the case of concentric hypertrophy as in the middle,
the chamber diameter is reduced due to increased wall
thickness. A hypertrophy leading to a wall
thickness of 1.5 cm, will give an EDD of 3 cm. A
systolic wall thickening of 0.5 cm will then be
(2 cm - 1.5 cm) / 1.5 cm = 33%; i.e. a clear reduction
in radial function. But 1 cm diameter shortening
is FS = (3 cm - 2 cm) / 3 cm = 33%, an apparent
increase in radial function, due to geometrical
misconception! In concentric remodelling (right), the
diameter is reduced. In the case of heart failure with
reduced myocardial function, (reduced wall
thickening), the diameter reduction may cause the FS
to be normal, despite the reduced radial function.
From the reasoning above, any conclusions about radial
function based on fractional shortening in the presence of
hypertrophy may be erroneous, and the term radial function
needs to be defined. The conclusion that there is radial
compensation for reduced longitudinal function should be
reserved to the cases where WT is increased (If this is
possible, it seems theoretically impossible, as the reduced
longitudinal shortening should correspond to reduced wall
thickening due to incompressibility).
It is extremely important
that if longitudinal and "radial function" are
compared, care should be taken that the measurements
are comparable. To compare for instance fractional
shortening of the LV diameter with longitudinal strain
(wall shortening), is comparing two different
measures, and may lead to completely erroneous
conclusions as shown above, where fractional
shortening increases but wall thickening decreases.
as the same erroneous results will be obtained by the
fractional shortening as of EF, as shown
below.
And this shows that fractional shortening is not a true
measure of "radial function".
Patient with concentric
hypertrophy. Looking at the cavity, the systolic function
may appear fair.

|

|
Wall thickness 17 mm,
EDD 40 mm, Fractional shortening was 35%,
however, wall thickening only 28%
|
Ejection fraction
Based on Nuclear or X-ray contrast studies, the first measures
was measurements of cavity reduction in systole, i.e. the
stroke volume. While this may be the most important
result of cardiac
pumping, it confers little information about the state of the
heart itself. A dilated ventricle can maintain stroke volume,
but it is reduced in terms of the left ventricle volume, and
may have a severely reduced contractility. Thus stroke volume
should be normalised for end diastolic volume, to obtain
Ejection fraction:
Ejection fraction is still
the most widely used measure of systolic left ventricular
function today. This is mainly due to the vast amount of
prognostic information from earlier studies, and the
prognostic interventions that are geared to a cut off
point in EF. Even so, EF has been shown to be a poor
prognosticator even in heart failure, when patients
without dilation is included (
227).
In assessing EF, it should be emphasized, however, that EF
is not a direct measure of myocardial function, as it
measures the cavity, not the myocardial deformation. At
best, it could be characterised as an indirect measure.
Does this matter? Yes. If we look at a few examples:
Again, the cavity approach works very well in dilated
hearts, but not in eccentric hypertrophy:
Classic view of
ejection fraction. In a dilated ventricle (right),
with thin wall, both wall thickening and
longitudinal shortening are reduced. The cavity
volume is increased, so the EF is reduced, even if
the stroke volume may be maintained. As the
ventricle dilates crosswise, the stroke volume is
maintained with a shorter MAPSE, thus the
longitudinal strain is also reduced, as shown below.
As we see, in dilated ventricles, there is a
correlation between longitudinal shortening and EF, as
explained
below.
However, this correlation is not present in
hypertrophic ventricles (
190).
But this is due to the fact that in concentric
geometry (hypertophy or remodelling) EF doesn't
measure systolic function at all.
The erroneous use of EF in
concentric geometry.
concentric
geometry, EF will not give true measures of systolic
function at all!
In concentric
hypertrophy (middle), as often seen in pressure
overload, the wall may be thickened, and the cavity
volume is usually reduced.
Concentric hypertrophy reduces the cavity volume. Absolute
wall thickening may often be preserved, while relative
wall thickening is reduced as in the example of rectional
shortening above. Then the longitudinal shrtening will
also be reduced. I have pointed this out concerning
fractional shortening as seen above, the reasoning was
taken further into three dimensions by MacIver (
228).
EF has been shown to be more related to absolute than
relative wall thickening (
229),
and may be unchanged or even increased, but stroke
volume is reduced, which also indicates systolic
dysfunction. This is the same finding as in
FS.
The same patient
as above.
EDV about 100 ml, EF about 55%. Again the systolic
function may appear normal, looking only at the
cavity. However, looking at long axis shortening, it
appears severely depressed.

|
 |
-
which is confirmed, systolic mitral annular
excursion is 5 mm and peak systolic
annular velocity is < 3 cm/s
|
In concentric remodelling, as in the atrophy of ageing,
where LV mass is unchanged. but ventricle size is reduced,
the EF will fare just as poorly. Wall thickness may be
unchanged or increased, but as the myocardial mass/volume
is reduced less than cavity size, the myocardial wall /
cavity volume ratio is decreased. Again, stroke volume is
reduced, EF may be normal. Absolute wall thickening may be
reduced, but relative wall thickening more, and the
longitudinal shortening is reduced in proportin to
relative wall thickening.
The annular displacement has been shown to be more
sensitive than EF in predicting events in heart failure (
36,
192)
and hypertension (
193).
But alas, interventional studies using echocardiography as
secondary outcome, persists in using only EF, instead of
including newer measures for direct comparison of the
ability in predicting clinical outcome as well as
establishing cut off values for intervention, as studies
are driven by investigators with little knowledge of
echocardiography. This is illustrated below.
|
Normal
left
ventricle
|
Dilated
left
ventricle
|
Concentric
hypertrophy
|
Concentric
remodelling
|
|
Diastole
|
Systole
|
Diastole |
Systole |
Diastole |
Systole |
Diastole |
Systole |
LV
length (cm) |
9.5
|
7.7
|
11
|
9.8
|
11
|
10.6
|
8.5
|
7.8
|
LV
outer diameter
(cm) |
6.0
|
5.7
|
7.5
|
7.1
|
6.5
|
6.18
|
4.5
|
4.2
|
Wall thickness (cm)
|
0.95
|
1.43
|
0.6
|
0.7
|
1.7
|
2.0
|
0.95
|
1.2
|
LV
Inner diameter
(cm) |
5.1
|
4.1
|
6.3
|
5.7
|
3.1
|
2.1
|
2.6
|
1.9
|
LV
cavity volume
|
123
|
78
|
228
|
167
|
55
|
24
|
30
|
15
|
Stroke
volume (ml)
|
|
45
|
|
61
|
|
31
|
|
15
|
Ejection
fraction
(%)
|
|
62
|
|
27
|
|
56
|
|
51
|
Fractional shortening
(%)
|
|
30
|
|
10
|
|
32
|
|
27
|
Wall
thickening
(%) |
|
50
|
|
20
|
|
20
|
|
25
|
Longitudinal
shortening
(%)
|
|
21
|
|
10
|
|
3
|
|
8
|
|
|
15
|
|
7
|
|
14
|
|
13
|
All measures are
calculated from a geometrical model of a half
ellipsoid with wall thickness in the apex being half
of the sides, all measures are calculated from the
input measures of LV length, outer diameter, wall
thickness and wall thickening.
While cavity parameters are preserved or even increased in
the concentric geometry, all systolic wall deformation
measures (longitudinal, circumferential and transmural
strain)are reduced.
It has been shown by speckle tracking observational
studies in various hypertrophic states, that all three
pricipal strain may be reduced, whil ejection fraction is
preserved (
230,
231,
232,
233).
Thus, the EF or FS is a measure that actually only works
with dilation of the ventricles and becomes erroneous in
the cases of reduced EDV.
EF is a
geometrical concept, and only works in some
geometries. Asd both the modelling
and the sudies cited above shows, the systolic finction
may be reduced in all directions despite a normal EF.
Because this has been poorly recognised, it has lead
to some fairly bizarre results. As systolic function has
been measured by EF, and diastolic function with mitral
flow parameters, the hypothesis of "isolated diastolic
heart failure" has been proposed. At the outset, measuring
systolic and diastolic function by different measures with
different sensitivity, is methodological nonsense in any
case.
This has been realised, ad the term is now substituted
with the term "Heart failure with normal or
prteserved ejection fraction" (
HFNEF or HFPEF).
But
as EF as a measure
of systolic function in the case of small, hypertrophic
ventricles is meaningless, the concepts are still
dubious, the emphasis of an erroneously normal EF remains.
The problem with both strain AND
EF in eccentric hypertrophy
In eccentric hypertrophy, the problem reverses, but in this
case it even affects strain. Basically, in eccentric
hypertrophy, the VV mass increases, but hte wall cavity ratio
remains normal (
146).
This
means that the ventricle enlarges mainly outwards, but as a
more or less normally thick wall (at least relative) surrounds
a much larger cavity, the mass has to increase. This is seen
in different states:
- As compensation for volume load in valvular
regurgitation. In this case, the end diastolic volume
increases, but so does the total stroke volume. In this
case the EF remains normal or even super normal. The
strains can thus also be expected to remain normal as long
as myocardial function is.
- In athletes heart. In this case there is eccentric
hypertrophy in response to endureance training, i.e. the
demands on the ventricle during near maximal performance.
The ventricle increases both in diameter and length, and
hence, in end diastolic volume.
Diagram illustrating
how eccentric hypertophy, with a larger ventricle with
normal wall thickness, will show reduced shortening
fraction due to the larger denominator, even if wall
thickening is normal. (Same wall thickening, larger end
diastolic diameter, same absolute, but less relative diameter shortening). The finding will be the
same in three dimensions, a larger ventricle with normal
strains will still show reduced EF, even if stroke
volume remains the same (Larger EDV, same stroke volume
= same absolute, but less relative volume
reduction; i.e. lower EF). This, however, is an over
simplification, unchanged stroke volume in a larger
ventricle will show reduced strains as discussed below.
As long as stroke volume
remains constant, the absolute wall thickening and
shortenings (strains) will also remain constant, but the
relative wall shortening will be less due to a longer
ventricle (i.we. lower longitudinal strain). As relative
wall shortening decreases, this may be seen to give a
lower wall thickening as well, as the two are interrelated
due to the incompressibility principle. If the wall
thickening is unchanged in terms of absolute amount of
wall thickening, the thickening is spread over a larger
surface of a longer and wider ventricle, and thus is
reduced in terms of relative wall thickening. And in fact
in a larger ventricle, even absolute wall thickening may
be reduced, as a lower amount of wall thickening wil
result in the same volume reduction when spread over a
larger surface.
Thus; in eccentric hypertrophy, the strains
will decrease, even with unchanged stroke volume.
Eccentric hypertrophy with
unchanged stroke volume. In the larger ventricle to the
left, the same stroke volume as to the right, can be
maintained by a smaller longitudinal shortening (MAE) due
to the wider ventricle as explained here.
This will result in a lower longitudinal strain. And the
strain is even more reduced as this smaller MAE is
relative to a greater LV length. But less
longitudinal shortening will also result in less wall
thickening, thus all strain are reduced. However, even
this is still an over simplification, as this is reasoned
without the compensatory regulatory mechanisms.
Thus, with an increased EDV and unchanged stroke volume, the
EF and FS may be reduced, and so may strains, due to the
larger ventricle.
However, athletes usually have downregulated heart rate as
well, which increases diastolic filling, and thus the
stroke volume.
In
the
intact body, the cardiac output is regulated according
to the circulatory need, by both autonomic balance and other
vasoactive and volume regulatory mechanisms affecting both
filling and resistance, contractility and heart rate
simultaneously.
This means that athletes at rest (having no need of increased
resting cardiac output - only of increased cardiac output
reserve) will
have unchanged cardiac output, a larger LV, with lower heart
rate and increased stroke volume. But as this is a result of a
lower sympathetic tone, the ejection fraction and strains
may still be
reduced, although to a lesser degree than if HR was unchanged,
and may be unpredictable.
For evaluation of systolic myocardial function in eccentric
hypertrophy, myocardial systolic velocities or strain rate
would be more appropriate, although the litterature is fairly
scarce, at least in comparing with normals. At least, the rate
of shortening should be less affected by the total stroke
volume, but afterload may still be a confounder, in general
athletes may have lower blood pressure, but also a lower
sympathetic tone.
Wall measurements -
long axis systolic function.
Wall thickening is a measure of systolic deformation. It can
be assessed semi quantitatively in B-mode.
Wall
motion score index (WMSI) by B.mode, being the
average of wall motion score of all evaluable segments
becomes a measure of global function, and has been
shown to correlate with EF in infarcted ventricles (
40).
It
has also been shown to be similar in sensitivity to reduced
function (and infarct size) to global strain (
189).
However, the index is useless unless there is regional
differences. Any dilated cardiomyopathy will show
hypokinesia in all segments, giving a WMSI of 2, regardless
of EF.
Wall thickening measured in M-mode, however, is only available
in limited segments, and can only be generalised to global
measures if the ventricle is symmetric. In addition, the
wall thickening is mainly a function of the long axis
shortening, due to the
incompressibility
of the heart muscle.
Systolic long axis
shortening
The
systolic long axis
function is measured by any means of any longitudinal
motion or deformation. I.e. Long axis shortening measured by
mitral annulus motion or
global strain, or shortening velocity
/ rate by
mitral annulus
velocity or
global strain rate.
It has been established that the longitudinal
shortening of the left ventricle, and thus the
longitudinal measures is closest related to the stroke
volume and EF, i.e. to the total left ventricular
volume change (
30
- 35,
56,
59,
60,
64
- 67,
116).
Longitudinal
systolic strain of the left ventricle is
shortening, normalised for diastolic length
(similar to EF, which is volume decrease (stroke
volume) normalised for end diastolic volume). As
longitudinal shortening describes most of the
actual ejection work, , there is a strong relation
between EF and longitudinal strain. Thus, it may
seem that the longitudinal fibres (or force
components) are the main contributors to the
ejection work, i.e. the isotonic part of the work.
Mitral annular systolic displacement
(MAPSE)

|
 |
Long axis shortening of the ventricle
equals the mitral annular systolic displacement. |
Longitudinal
M-mode through the mitral ring, displaying the
displacement of the mitral ring. The total
systolic displacement (MAPSE; mitral annular plane
systolic excursion) can be measured.
If the MAPSE is divided by the end diastolic
length of the ventricle (which, in fact is a
spatial derivation), it will give a measure of the
strain of the wall. The global strain of the left
ventricle is an average of more points of the
wall. The longitudinal (Lagrangian) strain
during systole is thus MAPSE /LD. |
Mitral annular plane systolic displacement or excursion
(MAPSE), and mitral annular systolic velocities, are
measurements of total ventricular shortening and shortening
velocity:
Mitral annular dexcursion can be measured by B-mode, M-mode or
tissue Doppler:

|

|
MAPSE by M-mode. In
this case the MAPSE was 14 mm in the septal site
and 16 mm in the lateral, giving an average of 15.
|
MAPSE by tissue Doppler
showing an MAPSE of about 15 mm.
|
The annular measurements reflect the total shortening of the
ventricle, and are thus measures of
global
longitudinal function.
The mitral annular systolic descent has had many names: The
mitral annular excursion (MAE)
(
31,
35,
37,
40)
has been used for a long time. Atrioventricular plane descent
(AVPD) (
30,
32,
34,
36)
is incorrect, as the term also comprises the tricuspid part,
and while tricuspid displacement and velocity can be measured
(and is higher than in the left ventricle) , it is usually
measured only in one point, and the relative weights for the
measurements is unclear.
However, the term
TAPSE for the tricuspid systolic
annular excursion has been firmly established. In order to
remain consistent in nomenclature, the corresponding term
MAPSE
for Mitral Annular Plane Systolic Excursion is in increasing
use. Thus, it might be the best term, and it still retains the
specificity that AVPD lacks.
The longitudinal shortening has been
shown to be very closely related to ejection fraction when
comparing different patients with normal or reduced left
ventricular function (
30,
31,
32,
34,
35,
36,
40,
64),
as
illustrated below:
When the left ventricle
dilates, the volume increases, and the stroke volume can
be maintained by a smaller fraction (Ejection fraction) of
the total (end diastolic) volume. At the same time, the
cross sectional area increases, so the volume can be
maintained by a smaller stroke length.
The relation between MAPSE
and EF has shown a correlation of 80 - 90%. However, the
relation only holds in dilated ventricles. In normal
ventricles, the MAPSE is related to the stroke volume (
59,
60,
116).
In
left ventricular hypertrophy, the MAPSE is reduced despite
preserved EF, and there is no correlation (
190).
In addition, the MAPSE is reduced in ventricles with
normal ejection fraction , the so-called HFNEF (
191),
i.e. despite normal ejection fraction.
The annular displacement has been shown to be more
sensitive than EF in predicting events in heart failure (
36,
192)
and hypertension (
193),
indicating that it is a more precise measure of systolic
function, that the cavity measurements. This may be due to
the shortcoming of EF in small ventricles / hypertrophy.
There is also a trend towards a better correlation with
infarct size than EF (
150).
Also, the MAPSE correlates better with BNP in heart
failure, that the fractional shortening (
204).
Newer studies, using MR, have found MAPSE to be a strong
predictor for MACE, both in general and in hypertension (
433,
434).
Thus, the MAPSE is a more
all round useful measure of longitudinal function than
EF.
There has been some arguments for measuring MAPSE only
during ejection, i.e. excluding the isovolumic phases (
194).
The value will be a little lower, and the main advantage
seems to be that
post
systolic shortening, not being part of the systolic
work, will be eliminated.
Systolic annular
displacement of the septal point. There is a small
shortening in the isovolumic contraction phase
(IVC), and post
systolic motion (PSM) after AVC, so the
systolic MAE is lower than the total MAE.
However, the total shortening is probably related to the
total ventricular size. This means that small ventricles
has a lower MAPSE, even if similar in relation to the
total length. This also means a lower stroke volume, of
course, from a smaller ventricle. So the relation MAPSE x
cross sectional area = SV still holds. However, this means
that some of the variations in MAPSE are due to heart
size, not heart function, which mans that the relation
with heart function has a reduced explained variance.
Theoretically, this means that the annular displacement
should be normalised for heart size, which also is the
case when using
global strain
instead, being relative shortening. This is definitely
necessary in children (
159,
214,
288),
where
the variation in heart size is great, the advantage in
adults, where variation in heart size is less (and less
than the difference betweeen normal and pathological) is
not documented.
Normal
values for MAPSE by age from the HUNT study
Interestingly, MAPSE is not gender dependent, unlike
global strain (
417):
Age distribution of MAPSE in the HUNT study.
There is a clear, significant reduction by age, but
no significant difference between genders.
In the HUNT echo study (
417),
there was a significant correlation between MAPSE and
age of 50%, and a weak (12%), but significant
correlation with BSA. This correlation, however, was
not sufficient to induce significant gender
difference. Age related values are shown in the figure
above, and table below:
Age
|
< 40
|
40 - 60
|
> 60
|
All
|
MAPSE (cm)
|
1.73 (0.20)
|
1.58 (0.23)
|
1.40 (0.22)
|
1.58 (0.25)
|
Age dependent values for
MAPSE from four walls, standard deviations in
parentheses
Previously, in dilated ventricles a cut off value of
< 1cm would identify an EF < 50% (
30,
31,
32,
34,
35).
However, MAPSE is reduced also in HFpEF (
191).
An empiric cut off of 1 cm would tend to exclude
serious pathology, but may give a sklight reduction in
sensitivity < 40 years, and slightly reduced
specificity > 60 years, as seen by the plot below.
Scatterplot of MAPSE by age. The previous
linear cut off of 1 cm will serve to exclude
serious pathology. The age dependent cut off of
mean - 2 SD, will give a higher sensitivity in
the youngest age group, and the highest
specificity in the hioghest age group. This can
also be seen from the boxplot above.
Where should measurements be
done?
Displacement is higher in the lateral than the septal part
of the mitral annulus (
154,
418,
435).
Thus, it is evident that the measurements are different if
different sites are chosen.
All studies have used the average of four points: septal
and lateral in the four chamber view, and anterior and
inferior in the two-chamber view. Thus the average is
fairly robust, representing a global average.
The HUNT study, however, shows that the mean of two walls
from four chamber view, four walls from four- and two
chamber views and six walls from all three views were less
than 1 mm (1.56 vs 1.58 vs 1.51 cm) (418). The variability
was higher in six wall average, probably due to variation
in defining the mitral ring point. Two- and four wall
averages had the same variability, but reproducibility has
been shown top be better by four walls ( reduced by about
25% by using four points instead of one (
154).
However, it has been shown that even in regional
dysfunction due to infarcts, MAPSE is globally depressed
in all points, without regional differeences between close
and remote points (
40,
436).
The reason for this is discussed under
regional
function.
Thus, it may not matter which wall is used, as long as it
is consistent, newer studies using lateral MAPSE have
shown good prognostic predictive value (
433,
434).
Normalised displacement.
Both annular displacement (MAPSE) and annular systolic
velocity can be normalised for (divided by) the length of
the ventricle or wall.

|

|
Systolic strain is
normalised MAPSE. The normalised MAPSE for
this ventricle with an end diastolic length of
9.2 cm and an MAPSEE of 15 mm is 15 / 92
= 16.3. THis corresponds to a longitudinal
strain of -16.3%. |
Compare with global
strain, in this case the global strain was
16.1%, giving a good comparison. However, the
two methods are different, as this method
normalises for the length of the curved wall,
and the actual values are dependent
on
the curvature (especially in the apex)
of the segments. |
However,
there is so far no agreement of how this normalisation
should be done. It is evident that the shortest length is
the mid length of the ventricle as seen above. Strain on
the other hand, is a myocardial measure, and should be
measured along the myocardium, i.e. wall length. But as
shown below, this can be done in various ways. In relation
to M-mode or tissue Doppler, the natural measure might be
along a straight line from apex to the mitral annulus. In
relation to speckle tracking, however, there may be
possibilities to measure along the wall. But still there
is a choice between midwall and endocardial. And finally,
the normalisation of MAPSE makes the new parameter much
more sensitive to
foreshortening.
 |

|
Diastolic and
systolic images of the heart. Systolic
shortening of the left ventricle relative to
diastolic length, is the systolic strain of
the ventricle. The longitudinal strain
during systole is thus:
However,
it is also evident that as the wall shortens,
it also thickens, to conserve the volume.
Heart muscle is generally assumed to be
incompressible.
|
Strain being (L - L0)
/ L0
may still not be
unambiguous, as shown below. Both the strain
length, L0 and the shortening (L - L0) will be different
when measured along a skewed line (red) and even
longer along a line following the wall curvature
(blue). As both strain length and shortening
increase when the curved line is used, the ratio
will not be as affected, but still, L0
will increase more than than the shortening.
This is shown in a practical example below.
|
Normalising for wall length will be easiest using the
straight line distance from apex to the mitral ring in
diastole, this has recently been done in the HUNT study (
417).
as illustrated below. Using the straight line is more
robust than the curved line as shown below, this is more
arbitrary, and also more robust than using the mid
ventricular length down to the middle of the mitral
ostium. The normalised MAPSE is the LV shortening divided
by the wall length, i.e. shortening per wall length,
and thus a measure of wall strain. The mean of two, four
or six walls will then be the global strain. The mean of
two walls is 9.4 cm, of four and six 9.5 cm, thus the
means are fairly similar as shown
here
(
386).
The means of MAPSE from two, four or six walls are
likewise les than 1 mm (
417),
so the mean GLS from two, four or six walls is also fairly
similar.
Illustration of normalised MAPSE based on
straight line wall length.
As for MAPSE, there was little
difference between the mean of four walls from
four- and two-chamber views and six walls from all
three standard views (15.9% vs 16.3%, respectively.
The
normalised annular displacement will be a measure of the
global strain, making it less dependent on ventricular
size (and thus, body size), at least in children. Recent
studies in children has shown normalised displacement to
be an age independent measure of systolic performance (
159,
214,
288),
i.e.
in the instance where the variation in LV size is greatest
in the normals. The study in children (
159)
did show better correlation with EF over a wide range of
pathology and age. In a small study in normal adults, it
has shown better correlation with EF (
217),
which may be an indication that it removes variability due
to LV size. However, introducing another measure (LV
length) will increase the measurement variability of the
composite parameter.
In the HUNT study, we did show that the measurement
variability was far less (less than 3%) than the
biological variability (around 15%). Thus the variability
was mainly biological. However, the main source of
biological variability was age, not body size in adults.
Thus the variability, measured by relative standard
deviation was 16% for MAPSE and 14% for normalised
MAPSE.
AS the diagnostic accuracy of a measure is mainly
dependent on:
- The variability of the measurements and
- The separation of the means of the healthy and
patient population
As variability of methods are similar,
there is so far still no evidence that normalisation
of MAPSE increases diagnostic accuracy. Only in
children, where the variation in body size is greater,
should normalisation be applied. The separation of
means cannot be assessed from the normal material
alone. On the other hand, small studies in selected
patient groups will not be sufficient. Only diagnostic
and prognostic studies in large, diverse populations,
with head to head comparisons between Global strain
and MAPSE will determine if GLS has any advantage.
The reason MAPSE and normalised MAPSE has the same
variability, is that age, not size is the main source
of variation.
MAPSE and normalised MAPSE vs. age. The
correlations are fairly similar.
Correlation with Age was -50% for MAPSE, -41% for
normalised MAPSE, with BSA only 12% for MAPSE, and
-23% for normalised MAPSE (
417).
Linear regression showed the two parameters to be
independent. There is a slightly lower correlation
with age for normalised MAPSE, As both MAPSE (
417)
and LV length (
386)
decrease with age, this is expected, but this effect
is not sufficient to reduce the relative standard
deviation of normalised MAPSE, compared to MAPSE to a
clinically important degree.
Four walls had lower variability than six.
The age and gender related values for MAPSE and
Normalised MAPSE were:
|
Women
|
Men
|
Age
|
MAPSE (cm)
|
Normalised
MAPSE (%)
|
MAPSE (cm) |
Normalised
MAPSE (%) |
< 40
|
1.73 (2.0)
|
18.1 (2.0)
|
1.72 (0.22)
|
16.5 (2.0)
|
40 - 60
|
1.58 (0.23)
|
17.0 (2.2)
|
1.58 (0.26)
|
15.4 (1.9)
|
> 60
|
1.33 (0.22)
|
14.8 (2.1)
|
1.45 (0.23)
|
14.9 (1.9)
|
All
|
1.58 (0.26)
|
17.0 (2.5)
|
1.58 (0.25)
|
15.5 (2.0)
|
As age, not body size was the main source of
variability, it is not so surprising that normalising
for heartlength did not reduce over all variability.
More surprising was the finding that normalsing for LV
length actually
increased body size
dependency. This was the opposite of expected.
Calving iceberg, Marguerite Bay,
Antarctica
There was a gender difference as well, with females
had a higher normalised MAPSE, as opposed to pure
MAPSE, which was not significantly different between
genders. By linear regression, the gender difference
was an effect of the difference in BSA, only the BSA
was significant, thus, there was no independent gender
effect.
Finally, as shown above, the Global strain will differ
depending on the choice of reference as shown below.
Illustration on how
the choice of reference length will affect the
strain value. The curved lines,representing the
longest wall measurements, will give the lowes
GLS value, the straight lines will be in
between, while the mean ventricular length will
be the shortest, and thus give the highest
strain value.
Global strain
Global strain and strain rate, may in theory be taken as
global measures of ventricular function. It's, however,
important to realise that different applications measures
strain in different ways. It has been shown that
strain
measurements vary between vendors (
373
- 383).
This of course means that global
strain has no meaning as a universal measure of LV
function, only in relation to each manufacturer.
Global strain can be achieved simply by measuring and
averaging the strain/strain rate in all segments of the
ventricle. However, there is one caveat:
Commercial software may give segmental values for six
segments in each imaging plane, resulting in a total
of 18 segmental values.
However,
this results in equal weight given to all myocardial
levels, despite there being much less myocardium in the
apical level. In order to ensure that the
average value gives similar weight to all parts of the
myocardium, only four segments in the apical level should
be included, as recommended by ASE/EAE (
146).
If
not,
the global measures may be misleading. (This is doubtful
in the
global strain
measurement by
2D
strain). The global strain by this application also
is somewhat
processing
dependent.
Global strain by speckle tracking has been introduced as a
new measure of global left ventricular function (
147).
This
compensates for the shortcomings of ejection fraction,
being both more correct in the case of small or
hypertrophic ventricles, and more sensitive (
149).
In
the 2D strain application, it should be noted that the
application relies heavily on the AV plane motion, and
then distributes the motion along the wall. By this
method, regional artifacts as drop outs and reverberations
will have less impact, which is an
advantage in
measuring global function. (As it may be a
disadvantage in
regional function, as the same smoothing may
reduce the sensitivity to regional reduced function).
It is unclear whether this application actually corrects
for the reduced amount of myocardium in the apex, giving
at the outset 6 segments per view, or 18 segments in
total. Bull's eye plots seem to show 17 segments, but
whether this is carried over to the calculation of global
strain is uncertain.
Global longitudinal strain by this method, has shown a
trend to be more sensitive to infarct size and correlate
better with infarct mass than EF. Global longitudinal
strain is thus a measure of wall shortening, normalised
for the length of the wall, as length is measured along
the curvature. Whether this allows sufficiently for the
reduced amount of myocardium in the apex, seems unclear,
as the referred study included 33 anterior and only 7
inferior infarcts. Annulus displacement had a slightly
less diagnostic accuracy than global strain, but whether
this was significant is less clear.
Normalising the annular
displacement for LV length (see below), did not
show ovious improvement in diagnostic ability, in this
group (
150,
417).
However,
annular displacement normalised for LV length
IS a measure of
longitudinal strain. Recent studies in children has shown
normalised displacement to be an age independent measure
of systolic performance (
159,
214,
288),
i.e. in the instance where the variation in LV size is
greatest in the normals.
In the HUNT study we have previously measured global
strain by the
combined
segmental method (
153).
In the new study, we assessed the over all variability by
global strain by this method by the relative standard
deviation, and found this to be 24%, higher than both
MAPSE and normalised MAPSE (
417).
Thus GLS by another method had no less variability than
MAPSE. It correlated negatively with age (R=-27%) and with
BSA (R=-27%). There was also a significant gender
difference, due to body size.
GLS showing the same gender dependency and
negative BSA relation as normalised MAPSE, thus
showing that the principle of normalising for length
in GLS is independent of method.
Thus, it's the
principle of normalising LV
shortening for LV length in GLS, and not the method that
induces the systematic error of increased and negative
body size dependency. Also GLS had no less variability
than MAPSE. Thus it is more dubious that GLS actually
confers better accuracy than MAPSE, although the large
direct comparative studies (addressing the separation of
means) still are lacking.
Findings are summarised in the following picture:
As shown by the boxplot, no significant gender
differences in MAPSE, but in normalised MAPSE and
GLS (women highest). All three measures are age
dependent. Lower panels shows weak positive
correlation between MAPSE and BSA, stronger negative
correlations between BSA and normalised MAPSE and
GLS. Gender differences only due to differences in
BSA, no independent contribution.
Why is GLS/ normalised MAPSE more BSA dependent
than MAPSE?
This relation to body size is due to the fact that
normalisation for length, corrects only for one dimension
of the heart size. However, we have found that the ratio
between LV length and LV diameter is the same for all body
sizes (
386),
although it varies with length. The Global strain
(MAPSE/L) is related to stroke volume, which is a function
of MAPSE x LV external surface. Thus, a longer ventricle
also has a larger cross sectional surface. This means that
the cross sectional area
increases by the square
of the diameter, and so does the stroke veolume, even
without any increase in MAPSE. On the other hand, as LV
length and diameter are proportional, With unchanged
MAPSE, the Normalised MAPSE and Global strain
decreases
proportional to the increase in length. Thus,
normalising for length only, induces a systematic error,
exaggeration of the heart size (or body size) effect:
Simplified and exaggerated example of the
inverse relation between heart size and normalised
MAPSE: We have previously found that the ratio
between LV external diameter and length is constant
across the spectrum of body sizes. In this example
the LV2 (left) is twice the LV1 (right). This then
means that as the length L2 is twice tle L1, the
external diameter D2 has to be twice the D1. With an
incompressible myocardium, the myocardial volume
inside the outer contour has to be constant through
the heart cycle. Then the SV is cross sectional
external area x MAPSE as shown by the green volumes,
thus the SV increases by the square of the LV
diameter, without any increase in MAPSE. Given equal
MAPSE, the area and SV in the largest ventricle thus
has to be 4 times that of the smallest ventricle.
Given the same MAPSE, thisthe is normalised to
a shorter length in the smallest ventricle, (I.e.
the shortest length L1 in ventricle 1, and the
longest length L2 in ventricle 2), and thus ,
on the other hand decreases by the inverse
of the increase in length. Then the strain of
the larges ventricle is half that of the smallest,
for the same MAPSE, hence, the inverse relation
between heart size (and body size) and global
strain. The gender difference is solely a function
of difference in body size.
Normal systolic strain and strain
rate values are method dependent
Normal values are often considered necessary if
measurements are to be used diagnostically. The mean ±2 SD
constitutes the 95% propability prediction interval for a
healthy subject. In addition, they will give additional
information about physiology.
In the north Tröndelag population (HUNT) study, 1266
subjects without known heart disease, hypertension and
diabetes were randomly selected from the total study
population of 49 827, and subjects with clinically
significant findings on echocardiography (a total of only
30) were excluded. (
153)
This is the largest normal population study of
echocardiographic strain and strain rate rate to date. End
systolic strain and peak systolic strain rate was measured
by the
combined
tissue Doppler / speckle tracking segmental strain
application of the Norwegian University of Science
and Technology:

|

|
Segmental
strain by tracking kernels at the segmental
borders, calculating strain as relative
segment length shortening, and strain rate as
the rate of strain
|
The
resulting strain rate curves.
|
Global strain and strain frate were calculated as the mean
of the segmental strains, but with one modification: The
apicolateral and apicoinferolateral segments, as well as
the apicoseptal and apicoanteroseptal segments were
averaged to reduce the number of apical segments to 16,
before calculating GLS and GLSR as the mean of 16
segments. The rationale for this was that there is far
less myocardium in the apex, and using 18 segments would
have over representyed the apical myocardium. This,
however, is an assumption, but using 18 segment would in
practice just have been another assumption, sjowing that
no method is free of such assumptions.
The results were
compared
to
other methods in a subset of subjects, showing small
differences. The study consisted of 673 women with a
mean BP of 127/71 ,mean age of 47,3 years and BMI of 25.8
and 623 men, with mean BP of 133/77, mean age of 50.6 and
BMI of 26.5. Both sexes were normally distributed with an
SD of 13.6 and 13.7 years, respectively. 20% of both sexes
were current smokers. Basic echo findings are in
accordance with other studies, like the findings of
Schirmer et al (
156,
157),
so
the study population may be assumed to be representative.
Normal values for global
systolic left ventricular strain and strain rate by age
and gender from the HUNT study. From (153).
|
Female
|
Male
|
All
|
|
End systolic strain
(%)
|
Peak systolic strain
rate
|
End systolic strain |
Peak systolic strain
rate |
End systolic
strain |
Peak systolic
strain rate |
< 40 years
|
-17.9% (2.1)
|
-1.09s-1 (0.12)
|
-16.8% (2.0)
|
-1.06s-1 (0.13) |
17.4 (2.2)
|
1.08 (0.12)
|
40 - 60 years
|
-17.6% (2.1)
|
-1.06s-1 (0.13) |
-18.8% (2.2)
|
-1.01s-1 (0.12) |
16.7 (2.3)
|
1.03 (0.12)
|
> 60 years
|
-15.9% (2.4)
|
-0.97s-1 (0.14) |
-15.5% (2.4)
|
-0.97s-1 (0.14) |
15.6 (2.4)
|
0.98 (0.14)
|
Over all
|
-17.4% (2.3)
|
-1.05s-1 (0.13) |
-15.9% (2.3)
|
-1.01s-1 (0.13) |
16.7 (2.4)
|
1.03 (0.13)
|
Values are given as mean
( SD). The customary definition of normal values as mean
± 2SD, giving about 95% of the normal population,
results in wider normal limits than previously shown as
cut off values in small patient studies. The values were
normally distributed, and with no clinically significant
differences between levels or walls. Values decline with
age, as does the velocity.
In a
later study (
417)
in the same material, we calculated normalised MAPSE as
generic, or linear strain, as has been done later as well (
444):
Normal values for global
systolic left ventricular generic (linear) strain from the
HUNT study
|
Women
|
Men
|
All
|
Age
|
Strain (%)
|
Strain (%) |
Strain (%) |
< 40
|
-18.1 (2.0)
|
-16.5 (2.0)
|
|
40 - 60
|
-17.0 (2.2)
|
-15.4 (1.9)
|
|
> 60
|
-14.8 (2.1)
|
-14.9 (1.9)
|
|
All
|
-17.0 (2.5)
|
-15.5 (2.0)
|
-16.3 (2.4)
|

|

|
MAPSE
in the septal and lateral mitral points from the
four chamber view. The averaged value is the most
representative for the global.
|
Different
denominators for strain calculation. The straight
line is the most robust, using clear anatomical
landmarks, as well as having the same angle as the
M-mode.
|
The results were very close to the segmental method (-16.3 vs
-16.7%). This is an attractive method in the meaning that the
strain is derived by generic measurements, and thus are
applicable across methods like ulttrasound, MR (
445)
or CT.
However, the actual numbers rest on the assumption of using
the straight line distance from apex to base as a proxy for
wall length. Using the actual length of the curved wall would
give a larger denominator, and thus a lower numerical strain
value, using the midwall ventricular length would give a
smaller denominator, and thus a larger numerical strain value.
This is illustrated in the example below:
Illustration on how the choice of reference length
will affect the strain value. The curved
lines,representing the longest wall measurements, will
give the lowes GLS value, the straight lines will be in
between, while the mean ventricular length will be the
shortest, and thus give the highest strain value.
Thus, the normal values rests on the
choice of assumptions, again independent of method (
446).
Speckle tracking derived strain, however, is somewhat
different.

|

|
Speckle
tracking. The application tracks speckles
inside the ROI. As can be seen, the ROI both
shortens and thickens at the same time.
|
Resulting
strain curves and end systolic strain values.
|
Global strain is a mean value of the strain within the
ROI. Using an ROI with a certain thickness, the mid ROI
line can be considered a reasonable proxy for mean global
values. But this line is curved as is the ROI. When the
ROI thickens,
the
midwall line will move inwards, and, as it is curved,
shorten due to wall thickening alone. This effect
comes in addition to the shortening due to MAPSE. This is
illustrated below:
Illustration of the behavior of
midwall and endocardial lengths in relation to
deformation. A: End diastole with midwall line
in red, endocardial line in blue. B: The isolated
effect of wall (ROI) thickening Will move the
midwall and endocardial lines inwards, mostly the
endocardial line, being pushed by the whole
thickness, while the midwall line is pushed only by
the outer half. AS the line is curved, this inward
movement will shorten the line. C: the additional
effect of ventricular shortening (MAPSE)
Thus, in this case, it seems that the GLS incorporates
both ventricular shortening and wall thickening into the
total shortening measure. And more so if tracking is done
along the endocardial border. But as
wall
thickening is largely the effect of wall shortening due
to incompressibility of the myocardium, this may
mean that the wall shortening is over represented in the
total GLS product. This may be a reason for why speckle
tracking GLS (
423,
427,
447-449)
is higher than what we found with segmental strain (153),
and also what we and other have found with linear strain (
417,
444),
and in my opinioun means that ST strain is less in harmony
with
actual
geometry of strain.
As with all other strains, speckle tracking strains rest
on assumptions: Mid/mean vs endocardial, width of the ROI,
and especially relation between apical and basal width,
weighting and numbers of apical segments, as
the
curvature is biggest in the apex.
In addition, the specific algorithms, tracking, weghting
of acoustic markers in terms of echogeneity and stability,
weighing of the relative contribution of the MAPSE, spline
smoothing, are all part of the "black boxes" of the
vendors, and industrial secrets.
There is no universal definition, and thus no
"ground truth" at all for GLS. Methods cannot be
validated, and GLS has to be considered only within
the method used.
A thorough overview over sources of discrepancies is given
in (
450)
Back to section index