During an active experiment the data storage, event, gas, and air cool states are controlled by the method parameters and/or method segments. However, you may manually change the state of these selections using the Control menu.
To display the status of these items, select Tools/User Preferences/Status Page and select the desired items to display on the status line. This will allow you to monitor the changes in these items.
Click the topic below for more information:
Use Event on the Control menu to activate (on) or deactivate (off) an external device connected to the instrument with the event cable (e.g., RCS set to Auto mode). This control overrides the Event segment in the method.
If the selected Event state is currently active, no action will be seen.
Data storage is typically controlled by the method segments; however, you can use Control/Store to manually turn data storage on and off during a running method. This control overrides the Data Storage segment in the method.
If the selected Data Storage option is currently active, no action will be seen.
Use Control/Gas/Gas #1/#2 to switch between the primary and secondary purge gases used during an experiment. A purge gas is a gas, often inert (e.g., nitrogen), that flows through the cell during the experiment. For example, oxidative stability experiments use nitrogen for the primary gas and oxygen for the secondary gas.
NOTE: Please make sure that you run your experiments with the same gas that you used to calibrate the system. For example, if you calibrate using nitrogen, make your runs with nitrogen.
Gas #1 refers to the primary purge gas connected to the instrument. Gas #1 is controlled by the Mass Flow Controller or plumbed directly to the Gas 1 connection.
Gas #2 refers to any secondary purge gas connected to the instrument. Gas #2 is controlled by the Mass Flow Controller or plumbed directly to the Gas 2 connection.
With Mass Flow Controllers a gas may also be switched during an experiment using the Gas Selection method segment.
If the selected Gas state is currently active, no action will be seen.
NOTE: If you are using laboratory purge gas, rather than bottle purge gas, you will need to install an external drier.
CAUTION: Use of corrosive gases will shorten the life of the instrument. NEVER use a corrosive gas in the instrument. The MFC is not designed to work with corrosive gases.
WARNING: Use of an explosive gas as a purge gas is dangerous and is not recommended for this instrument.
Click here for more information on the DSC purge gases. more
Use Air Cool on the Control menu to open (on) or close (off) the internal solenoid valve connected to the Air Cool connector on the exterior of the instrument.
It is not recommended that you turn off Air Cool while running an experiment with a Finned Air Cooling System.